Why does OpenChoice Desktop "Get Screen" operation time-out and display "Data transfer ended unexpectedly", when “Select Instrument”, “Get Data”, “Get & Send Settings” operations are successful?

問題:

Why does OpenChoice Desktop "Get Screen" operation time-out and display "Data transfer ended unexpectedly", when “Select Instrument”, “Get Data”, “Get & Send Settings” operations are successful?

答案:

When connecting to a TPS2000/B, TDS1000, TDS2000, or TDS3000/B series scopes with a Centronics parallel printer port using direct RS232, or RS232 to USB adapter, you must set the scope’s Printer Port to “RS232” (TDS1000/2000, TPS2000/B; Utility, Options, Printer Setup menu), (TDS3000/B; Utility, System Hardcopy, Port menu) so OpenChoice Desktop “Get Screen” operation will not time out on the first try.


此常見問答集適用於: