Why do I get an error when creating a pattern with a block size smaller than 960 on the DTG5274 and DTG5334 or 240 on the DTG5078?

問題:

Why do I get an error when creating a pattern with a block size smaller than 960 on the DTG5274 and DTG5334 or 240 on the DTG5078?

答案:

In hardware sequencer mode this is the limitation, but using the software sequencer mode you can select a block size as low as 1. To activate the software sequencer mode use the selection in the Sequencer menu. You will notice when software sequencer is used some of the trigger and jump features are disabled.