問題:
I have the P6015A probe without option 1R+ Probe Readout. Now I have a scope that will apply the 1000X attenuation factor automatically if the probe had that feature. Can it be added?
答案:
Yes, the BNC connector on the probe compensation box can be unsoldered, unscrewed, and replaced with the Tektronix part number 131-5320-00. All probe coding logic is contained in the new connector.