問題:
I am using the 80SSPAR software and the Iconnect S-parameter Wizard software and Some of the waveforms are not positioned properly on screen.
答案:
The deskew process, before measuring the first DUT, had trouble. If using a test fixture to connect the DUT, perform the deskew with just the cables.