Can the optional lead set for the single ended P6960 be used on the P6980?

問題:

Can the optional lead set for the single ended P6960 be used on the P6980?

答案:

Yes, you can pass differential signals through the data channels of the leadset, but only single ended clocks/qualifiers since the minus side of these signals are grounded on the leadset board.  

There will be up to 0.5" of skew between the + and – sides of the differential signals, as the traces on the leadset board are not matched in length. 