Digital Multimeters Datasheet
Tektronix DMM4050 and DMM4040
Tektronix 不再銷售此產品規格表上的產品。
檢視 Tektronix Encore 以取得檢修測試設備。
檢查這些產品的支援和保固狀態。
Features & Benefits
Key Performance Specifications
- 6.5 Digit Resolution
- Basic VDC Accuracy of up to 0.0024% (1 yr.)
- 100 mV to 1000 V Voltage Range, with up to 100 nV Resolution
- 100 μA to 10 A Current Range, with up to 100 pA Resolution
- 10 Ω to 1 GΩ Range, with up to 10 μΩ Resolution
- CAT I 1000 V, CAT II 600 V
Available Functions and Features
- Volts, Ohms, and Amps Measurements
- True RMS (AC, AC + DC) Measurements
- Diode and Continuity Testing
- Frequency and Period Measurements
- Temperature and Capacitance Measurements (DMM4050)
- 2×4 Ohms 4-wire Measurement Technique
- TrendPlot™ Paperless Data Recorder Mode
- Measurement Statistics
- Histogram Mode
Connectivity
- Front and Rear 2×4 Measurement Inputs
- USB Host Port on Front Panel for Easy Storage of Measurement Data and Instrument Settings
- RS-232, LAN, and GPIB Ports on Rear Panel for Quick PC Connectivity
- Includes USB to RS-232 Interface Adapter Cable
- Includes National Instrument’s LabVIEW SignalExpress™ TE Limited Edition for Connecting Your Bench
3-year Warranty
Feature-Rich Tools for Precision Measurements
As the circuits in embedded system designs become more sophisticated with tighter tolerances, you must measure a multitude of different parameters with a high degree of accuracy to validate your design. The Tektronix DMM4050 and DMM4040 6.5 digit bench multimeters pack many different functions and analysis into one instrument, all with exceptional precision and performance.
Typical multimeter measurements – volts, ohms, and amps – are made with a basic VDC accuracy of up to 0.0024%, and resolution of 100 pA and 10 μΩ, ensuring you have the performance you need for today’s demanding designs. You can also use the DMM4050/4040 to measure frequency and period, and to perform continuity and diode tests. For additional flexibility, the DMM4050 offers temperature and capacitance measurements. This allows you to replace your temperature meter, capacitance meter, counter, continuity tester, and traditional DMM with one versatile instrument, saving bench space and cost.
Analyze Your Device with Graphical Display Modes
With the unique dual display of the DMM4050/4040, you can measure two different parameters of the same signal from one test connection. To reveal signal quality issues like drift, intermittent transients, and stability, you can view the data as a real-time trend plot or a histogram with the DMM4050/4040’s graphical display mode, or you can use measurement statistics to track how signal parameters are changing over time.
TrendPlot™ Paperless Recorder Mode
TrendPlot display.
Depending on your test case, your signal parameters may change from instant to instant. By taking multiple measurements over minutes, hours, or days, you can quantify those changes. With TrendPlot™, you can graphically plot the trend of a measured value over time, from short time spans to extended periods of time. TrendPlot can be used with measurements like DC voltage, DC current, frequency, resistance, and temperature. AC voltage and current can be plotted as RMS measurements.
Measurement Statistics
A sample of a Min/Max/Avg/SD Statistics Report.
With integrated statistics processing, you can calculate both the average and standard deviation of a measurement, as well as track the minimum and maximum measured values, with the push of a button. Statistics can be performed on DC voltage, AC voltage, AC voltage in dB, DC current, AC current, resistance, capacitance, frequency, period, and temperature measurements.
Histogram Plots
Histogram display.
To graphically see the average and standard deviation of a set of measurements, you can use the histogram function to see the distribution of measurement results.
Designed to Make Your Work Easier
The DMM4050/4040 multimeter is designed with the ease of use and familiar operation you expect from Tektronix.
Intuitive Operation
Dedicated front-panel buttons provide fast access to frequently used functions and parameters, reducing setup time. You no longer need to search through software menus to find the function you need.
Dual Display
With the unique dual display, you can measure two different parameters of the same signal from one test connection.
Easy Data Storage and PC Connectivity
A USB port on the front panel makes it easy to store measurement data and instrument setups to a memory stick. Or connect to your PC with the LAN, RS-232, or GPIB port on the back panel. A USB to RS-232 interface adapter cable is included standard with the DMM4050/4040 to make it easy to connect to your PC’s USB port.
Simple and Accurate 4-wire Measurements
Patented split terminal jacks for the 2×4 ohms function allow you to perform 4-wire measurements using only two leads instead of four. Special test lead accessories are available to enable you to establish the connection. You get excellent resolution and accuracy plus the convenience and ease of using a single pair of leads.
Connect Your Bench for Intelligent Debug
SignalExpress acquiring data from Tektronix DMM4050 and DPO3052.
SignalExpress used to export DMM4050 data into Excel.
Easily capture, save, and analyze measurement results from your multimeter with the special Tektronix Edition of National Instrument’s LabVIEW SignalExpress™ software. Every DMM4050 and DMM4040 ships with a free copy of the Limited Edition version of SignalExpress for basic instrument control, data logging, and analysis. The optional Professional Edition offers over 200 built-in functions that provide additional signal processing, advanced analysis, sweeping, limit testing, and user-defined step capabilities.
SignalExpress supports the range of Tektronix bench instruments*1, enabling you to connect your entire test bench. You can then access the feature-rich tools packed into each instrument from one intuitive software interface. This allows you to automate complex measurements requiring multiple instruments, log data for an extended period of time, time correlate data from multiple instruments, and easily capture and analyze your results, all from your PC. Only Tektronix offers a connected test bench of intelligent instruments to simplify and speed debug of your complex design.
Performance You Can Count On
In addition to industry-leading service and support, every DMM4050 and DMM4040 multimeter comes backed with a three-year standard warranty.
*1 NI LabVIEW SignalExpress supports the following Tektronix instruments: MSO/DPO4000/3000/2000 Series oscilloscopes, TDS3000C/2000B/1000B Series oscilloscopes, AFG3000 Series arbitrary/function generators, DMM4050/4040/4020 Series digital multimeters.
Characteristics
General Specifications
Voltage
Characteristic |
Description |
---|---|
100 V Setting |
90 V to 110 V |
120 V Setting |
108 V to 132 V |
220 V Setting |
198 V to 242 V |
240 V Setting |
216 V to 264 V |
Frequency |
47 Hz to 440 Hz. Automatically sensed at power-on |
Power Consumption |
28 VA peak (12 Watt average) |
Dimensions
Dimension |
mm |
in. |
---|---|---|
Height |
88 |
3.46 |
Width |
217 |
8.56 |
Depth |
297 |
11.7 |
Weight |
kg |
lb. |
Net |
3.6 |
8.0 |
Shipping |
5.0 |
11.0 |
Display
Vacuum Fluorescent Display, dot matrix |
Environment
Characteristic |
Description |
---|---|
Temperature |
|
Operating |
0 °C to 55 °C |
Storage |
–40 °C to 70 °C |
Warm Up |
1 hour to full uncertainty specifications |
Relative Humidity (noncondensing) |
|
Operating |
<90% (0 °C to 28 °C) <80% (28 °C to 40 °C) <50% (40 °C to 55 °C) |
Storage |
<95% (–40 °C to 70 °C) |
Altitude |
|
Operating |
2,000 meters |
Storage |
12,000 meters |
Vibration |
Complies with Mil-T-28800F Type III, Class 5 (Sine only) |
Safety |
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V / CAT II 600 V |
EMC |
Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This meter has shown susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz |
Triggering
Characteristic |
Description |
---|---|
Samples per Trigger |
1 to 50,000 |
Trigger Delay |
0 s to 3600 s; in 10 μs increments |
External Trigger Delay |
<1 ms |
External Trigger Jitter |
<500 μs |
Trigger Input |
TTL Levels |
Trigger Output |
5 V max (open collector) |
Memory
Model |
Description |
---|---|
DMM4050/4040 |
10,000 measurements, internal, and up to 2 Gigabyte capacity with USB memory module (available separately) through front-panel USB port |
Math Functions
Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics (min/max/average/standard deviation), and Limit Test |
Dual Function Compatibility Matrix
Primary Function |
Secondary Function |
||||||||
---|---|---|---|---|---|---|---|---|---|
VDC |
VAC |
IDC |
IAC |
FREQ |
OHMS |
DIODE |
Period |
RTD |
|
VDC |
|
X |
|
|
|
|
|
|
|
VAC |
X |
|
|
|
X |
|
|
|
|
IDC |
|
|
|
X |
|
|
|
|
|
IAC |
|
|
X |
|
X |
|
|
|
|
FREQ |
|
|
|
|
|
|
|
X |
|
OHMS |
|
|
|
|
|
|
|
|
X |
DIODE |
|
|
|
|
|
|
|
|
|
Continuity |
|
|
|
|
X |
|
|
|
|
Electrical
Characteristic |
Description |
---|---|
Input Protection |
1000 V all ranges |
Overrange |
20% on all ranges except 1000 VDC, 1000 VAC, Diode, and 10 A ranges |
Remote Interfaces
RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available to connect the meter to a PC USB port) IEEE 488.2. LAN and “Ethernet 10/100BASE-T with DHCP (for IP address) option”. |
Warranty
Three years |
Electrical Specifications
DC Voltage Specifications
Accuracy specifications are valid for 6½ digit resolution mode after at least a 1 hour warm-up with Auto Zero enabled.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326-1:2000-11.
Characteristic |
Description |
---|---|
Maximum Input |
1000 V on any range |
Common Mode Rejection |
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance) |
Normal Mode Rejection |
60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1% 100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1% |
Measurement Method |
Multiramp A/D |
A/D Linearity |
0.0002% of measurement + 0.0001% of range |
Input Bias Current |
<30 pA at 25 °C |
Auto-zero Off Operation |
Following instrument warm-up at calibration temperature ±1 °C and less than 10 minutes, add error: 0.0002% range additional error +5 μV |
Analog Filter |
When using the analog filter, specifications are relative to within one hour of using the ZERO function for that range and NPLC setting |
DC Ratio |
Accuracy is ±(Input accuracy + Reference accuracy), where Input accuracy = DC Voltage accuracy for the HI to LO Input (in ppm of the Input voltage), and Reference accuracy = DC Voltage accuracy for the HI to LO (Sense) Reference (in ppm of the Reference voltage) |
Settling Considerations |
Measurement settling times are affected by source impedance, cable dielectric characteristics, and input signal changes |
Input Characteristics
Range |
Resolution |
Resolution |
Input Impedance |
||
---|---|---|---|---|---|
4½ Digits |
5½ Digits |
6½ Digits |
|||
100 mV |
100.0000 mV |
10 μV |
1 μV |
100 nV |
10 MΩ or >10 GΩ 2 |
1 V |
1.000000 V |
100 μV |
10 μV |
1 μV |
10 MΩ or >10 GΩ 2 |
10 V |
10.00000 V |
1 mV |
100 μV |
10 μV |
10 MΩ or >10 GΩ 2 |
100 V |
100.0000 V |
10 mV |
1 mV |
100 μV |
10 MΩ ±1% |
1000 V |
1,000.000 V |
100 mV |
10 mV |
1 mV |
10 MΩ ±1% |
2 Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance.
DMM4050 Accuracy
Range |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|
Accuracy is given as ±(% measurement + % of range) |
||||
100 mV |
0.0025 + 0.003 |
0.0025 + 0.0035 |
0.0037 + 0.0035 |
0.0005 + 0.0005 |
1 V |
0.0018 + 0.0006 |
0.0018 + 0.0007 |
0.0025 + 0.0007 |
0.0005 + 0.0001 |
10 V |
0.0013 + 0.0004 |
0.0018 + 0.0005 |
0.0024 + 0.0005 |
0.0005 + 0.0001 |
100 V |
0.0018 + 0.0006 |
0.0027 + 0.0006 |
0.0038 + 0.0006 |
0.0005 + 0.0001 |
1000 V |
0.0018 + 0.0006 |
0.0031 + 0.001 |
0.0041 + 0.001 |
0.0005 + 0.0001 |
DMM4040 Accuracy
Range |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|
Accuracy is given as ±(% measurement + % of range) |
||||
100 mV |
0.003 + 0.003 |
0.004 + 0.0035 |
0.005 + 0.0035 |
0.0005 + 0.0005 |
1 V |
0.002 + 0.0006 |
0.003 + 0.0007 |
0.004 + 0.0007 |
0.0005 + 0.0001 |
10 V |
0.0015 + 0.0004 |
0.002 + 0.0005 |
0.0035 + 0.0005 |
0.0005 + 0.0001 |
100 V |
0.002 + 0.0006 |
0.0035 + 0.0006 |
0.0045 + 0.0006 |
0.0005 + 0.0001 |
1000 V |
0.002 + 0.0006 |
0.0035 + 0.0010 |
0.0045 + 0.0010 |
0.0005 + 0.0001 |
Additional Errors
Digits |
NPLC |
Additional NPLC Noise Error |
---|---|---|
6½ |
100 |
0% of range |
6½ |
10 |
0% of range |
5½ |
1 |
0.001% of range |
5½ |
0.2 |
0.0025% of range ±12 μV |
4½ |
0.02 |
0.017% of range ±17 μV |
AC Voltage Specifications
AC Voltage specifications are for AC sinewave signals >5% of range. For inputs from 1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for 50 kHz to 100 kHz, add 0.13% of range.
Characteristic |
Description |
---|---|
Maximum Input |
1000 VRMS or 1414 V peak or 8 × 107 Volts-Hertz product (whichever is less) for any range |
Measurement Method |
AC-coupled true RMS. Measures the AC component of input with up to 1000 VDC bias on any range |
AC Filter Bandwidth |
|
Slow |
3 Hz - 300 kHz |
Medium |
20 Hz - 300 kHz |
Fast |
200 Hz - 300 kHz |
Common Mode Rejection |
70 dB at 50 Hz or 60 Hz ±0.1% (1 kΩ unbalance) |
Maximum Crest Factor |
5:1 at Full Scale |
Additional Crest Factor Errors (<100 Hz) |
Crest factor 1-2, 0.05% of full scale Crest factor 2-3, 0.2% of full scale Crest factor 3-4, 0.4% of full scale Crest factor 4-5, 0.5% of full scale Only applies for non-sinusoid signals |
Input Characteristics
Range |
Resolution |
Resolution |
Input Impedance |
||
---|---|---|---|---|---|
4½ Digits |
5½ Digits |
6½ Digits |
|||
100 mV |
100.0000 mV |
10 μV |
1 μV |
100 nV |
1 MΩ ±2% shunted by <100 pf |
1 V |
1.000000 V |
100 μV |
10 μV |
1 μV |
|
10 V |
10.00000 V |
1 mV |
100 μV |
10 μV |
|
100 V |
100.0000 V |
10 mV |
1 mV |
100 μV |
|
1000 V |
1,000.000 V |
100 mV |
10 mV |
1 mV |
DMM4050/4040 Accuracy
Range |
Frequency |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|---|
Accuracy is given as ±(% measurement + % of range) |
|||||
100 mV |
3 - 5 Hz |
1.0 + 0.03 |
1.0 + 0.04 |
1.0 + 0.04 |
0.1 + 0.004 |
5 - 10 Hz |
0.35 + 0.03 |
0.35 + 0.04 |
0.35 + 0.04 |
0.035 + 0.004 |
|
10 Hz - 20 kHz |
0.04 + 0.03 |
0.05 + 0.04 |
0.06 + 0.04 |
0.005 + 0.004 |
|
20 - 50 kHz |
0.1 + 0.05 |
0.11 + 0.05 |
0.12 + 0.05 |
0.011 + 0.005 |
|
50 - 100 kHz |
0.55 + 0.08 |
0.6 + 0.08 |
0.6 + 0.08 |
0.06 + 0.008 |
|
100 - 300 kHz 3 |
4.0 + 0.50 |
4.0 + 0.50 |
4.0 + 0.50 |
0.20 + 0.02 |
|
1 V |
3 - 5 Hz |
1.0 + 0.02 |
1.0 + 0.03 |
1.0 + 0.03 |
0.1 + 0.003 |
5 - 10 Hz |
0.35 + 0.02 |
0.35 + 0.03 |
0.35 + 0.03 |
0.035 + 0.003 |
|
10 Hz - 20 kHz |
0.04 + 0.02 |
0.05 + 0.03 |
0.06 + 0.03 |
0.005 + 0.003 |
|
20 - 50 kHz |
0.1 + 0.04 |
0.11 + 0.05 |
0.12 + 0.05 |
0.011 + 0.005 |
|
50 - 100 kHz |
0.55 + 0.08 |
0.6 + 0.08 |
0.6 + 0.08 |
0.06 + 0.008 |
|
100 - 300 kHz 3 |
4.0 + 0.50 |
4.0 + 0.50 |
4.0 + 0.50 |
0.2 + 0.02 |
|
10 V |
3 - 5 Hz |
1.0 + 0.02 |
1.0 + 0.03 |
1.0 + 0.03 |
0.1 + 0.003 |
5 - 10 Hz |
0.35 + 0.02 |
0.35 + 0.03 |
0.35 + 0.03 |
0.035 + 0.003 |
|
10 Hz - 20 kHz |
0.04 + 0.02 |
0.05 + 0.03 |
0.06 + 0.03 |
0.005 + 0.003 |
|
20 - 50 kHz |
0.1 + 0.04 |
0.11 + 0.05 |
0.12 + 0.05 |
0.011 + 0.005 |
|
50 - 100 kHz |
0.55 + 0.08 |
0.6 + 0.08 |
0.6 + 0.08 |
0.06 + 0.008 |
|
100 - 300 kHz 3 |
4.0 + 0.50 |
4.0 + 0.50 |
4.0 + 0.50 |
0.2 + 0.02 |
|
100 V |
3 - 5 Hz |
1.0 + 0.02 |
1.0 + 0.03 |
1.0 + 0.03 |
0.1 + 0.003 |
5 - 10 Hz |
0.35 + 0.02 |
0.35 + 0.03 |
0.35 + 0.03 |
0.035 + 0.003 |
|
10 Hz - 20 kHz |
0.04 + 0.02 |
0.05 + 0.03 |
0.06 + 0.03 |
0.005 + 0.003 |
|
20 - 50 kHz |
0.1 + 0.04 |
0.11 + 0.05 |
0.12 + 0.05 |
0.011 + 0.005 |
|
50 - 100 kHz |
0.55 + 0.08 |
0.6 + 0.08 |
0.6 + 0.08 |
0.06 + 0.008 |
|
100 - 300 kHz 3 |
4.0 + 0.50 |
4.0 + 0.50 |
4.0 + 0.50 |
0.2 + 0.02 |
|
1000 V |
3 - 5 Hz |
1.0 + 0.015 |
1.0 + 0.0225 |
1.0 + 0.0225 |
0.1 + 0.00225 |
5 - 10 Hz |
0.35 + 0.015 |
0.35 + 0.0225 |
0.35 + 0.0225 |
0.035 + 0.00225 |
|
10 Hz - 20 kHz |
0.04 + 0.015 |
0.05 + 0.0225 |
0.06 + 0.0225 |
0.005 + 0.00225 |
|
20 - 50 kHz |
0.1 + 0.03 |
0.11 + 0.0375 |
0.12 + 0.0375 |
0.011 + 0.00375 |
|
50 - 100 kHz 4 |
0.55 + 0.06 |
0.6 + 0.06 |
0.6 + 0.06 |
0.06 + 0.006 |
|
100 - 300 kHz 3, 4 |
4.0 + 0.375 |
4.0 + 0.375 |
4.0 + 0.375 |
0.2 + 0.015 |
3 Typically 30% reading error at 1 MHz.
4 1000 V range is limited to 8 × 107 Volt-Hertz.
Additional Low Frequency Errors
Frequency |
AC Filter |
||
---|---|---|---|
3 Hz (Slow) |
20 Hz (Medium) |
200 Hz (Fast) |
|
Error is stated as % of reading. |
|||
10 - 20 Hz |
0 |
0.25 |
— |
20 - 40 Hz |
0 |
0.02 |
— |
40 - 100 Hz |
0 |
0.01 |
0.55 |
100 - 200 Hz |
0 |
0 |
0.2 |
200 Hz - 1 kHz |
0 |
0 |
0.02 |
>1 kHz |
0 |
0 |
0 |
Resistance
Specifications are for 4-wire resistance function, 2 × 4-wire resistance, or 2-wire resistance with zero. If zero is not used, add 0.2 Ω for 2-wire resistance plus lead resistance, and add 20 mΩ for 2 × 4-wire resistance function.
Characteristic |
Description |
---|---|
Measurement Method |
Current source referenced to LO input |
Max Lead Resistance (4-wire ohms) |
10% of range per lead for 10 Ω, 100 Ω, 1 kΩ ranges. 1 kΩ per lead on all other ranges |
Input Protection |
1000 V on all ranges |
Common Mode Rejection |
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance) |
Normal Mode Rejection |
60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1% 100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1% |
Analog Filter |
When using the analog filter, specifications are relative to within one hour of using the ZERO function for that range and NPLC setting |
Input Characteristics
Range |
Resolution |
Resolution |
Source Current |
||
---|---|---|---|---|---|
4½ Digits |
5½ Digits |
6½ Digits |
|||
10 Ω |
10.00000 Ω |
1 mΩ |
100 μΩ |
10 μΩ |
5 mA / 13 V |
100 Ω |
100.0000 Ω |
10 mΩ |
1 mΩ |
100 μΩ |
1 mA / 6 V |
1 kΩ |
1.000000 kΩ |
100 mΩ |
10 mΩ |
1 mΩ |
1 mA / 6 V |
10 kΩ |
10.00000 kΩ |
1 Ω |
100 mΩ |
10 mΩ |
100 μA / 6 V |
100 kΩ |
100.0000 kΩ |
10 Ω |
1 Ω |
100 mΩ |
10 μA / 13 V |
1 MΩ |
1.000000 MΩ |
100 Ω |
10 Ω |
1 Ω |
10 μA / 13 V |
10 MΩ |
10.00000 MΩ |
1 kΩ |
100 Ω |
10 Ω |
1 μA / 13 V |
100 MΩ |
100.0000 MΩ |
10 kΩ |
1 kΩ |
100 Ω |
1 μA || 10 MΩ / 10 V |
1.0 GΩ |
1.000000 GΩ |
100 kΩ |
10 kΩ |
1 kΩ |
1 μA || 10 MΩ / 10 V |
DMM4050/4040 Accuracy
Range |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|
Accuracy is given as ±(% measurement + % of range) |
||||
10 Ω |
0.003 + 0.01 |
0.008 + 0.03 |
0.01+ 0.03 |
0.0006 + 0.0005 |
100 Ω |
0.003 + 0.003 |
0.008 + 0.004 |
0.01 + 0.004 |
0.0006 + 0.0005 |
1 kΩ |
0.002 + 0.0005 |
0.008 + 0.001 |
0.01 + 0.001 |
0.0006 + 0.0001 |
10 kΩ |
0.002 + 0.0005 |
0.008 + 0.001 |
0.01 + 0.001 |
0.0006 + 0.0001 |
100 kΩ |
0.002 + 0.0005 |
0.008 + 0.001 |
0.01 + 0.001 |
0.0006 + 0.0001 |
1 MΩ |
0.002 + 0.001 |
0.008 + 0.001 |
0.01 + 0.001 |
0.001 + 0.0002 |
10 MΩ |
0.015 + 0.001 |
0.02 + 0.001 |
0.04 + 0.001 |
0.003 + 0.0004 |
100 MΩ |
0.3 + 0.01 |
0.8 + 0.01 |
0.8 + 0.01 |
0.15 + 0.0002 |
1 GΩ |
1.0 + 0.01 |
1.5 + 0.01 |
2.0 + 0.01 |
0.6 + 0.0002 |
Additional Ohms Errors
Digits |
NPLC |
Additional NPLC Noise Error |
---|---|---|
6½ |
100 |
0% of range |
6½ |
10 |
0% of range |
5½ |
1 |
0.001% of range |
5½ |
0.2 |
0.003% of range ±7 mΩ |
4½ |
0.02 |
0.017% of range ±15 mΩ |
DC Current
Characteristic |
Description |
---|---|
Input Protection |
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V fuses, limits of 400 mA continuous 550 mA for 2 minutes on, 1 minute off |
Common Mode Rejection |
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance) |
Normal Mode Rejection |
60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1% 100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1% |
Analog Filter |
When using the analog filter, specifications are relative to within one hour of using the ZERO function for that range and NPLC setting |
Input Characteristics
Range |
Resolution |
Resolution |
Shunt Resistance (Ohms) |
Burden Voltage |
||
---|---|---|---|---|---|---|
4½ Digits |
5½ Digits |
6½ Digits |
||||
100 μA |
100.0000 μA |
10 nA |
1 nA |
100 pA |
100 Ω |
<0.015 V |
1 mA |
1.000000 mA |
100 nA |
10 nA |
1 nA |
100 Ω |
<0.15 V |
10 mA |
10.00000 mA |
1 μA |
100 nA |
10 nA |
1 Ω |
<0.025 V |
100 mA |
100.0000 mA |
10 μA |
1 μA |
100 nA |
1 Ω |
<0.25 V |
400 mA 7 |
400.000 mA |
100 μA |
10 μA |
1 μA |
1 Ω |
<0.50 V |
1 A 6 |
1.000000 A |
100 μA |
10 μA |
1 μA |
0.01 Ω |
<0.05 V |
3 A 5 |
3.00000 A |
1 mA |
100 μA |
10 μA |
0.01 Ω |
<0.15 V |
10 A |
10.00000 A |
1 mA |
100 μA |
10 μA |
0.01 Ω |
<0.5 V |
5 Part of 10 A range.
6 Available on the front-panel terminal only.
7 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
DMM4050/4040 Accuracy
Range |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|
Accuracy is given as ±(% measurement + % of range) |
||||
100 μA |
0.01 + 0.02 |
0.04 + 0.025 |
0.05 + 0.025 |
0.002 + 0.003 |
1 mA |
0.007 + 0.005 |
0.030 + 0.005 |
0.05 + 0.005 |
0.002 + 0.0005 |
10 mA |
0.007 + 0.02 |
0.03 + 0.02 |
0.05 + 0.02 |
0.002 + 0.002 |
100 mA |
0.01 + 0.004 |
0.03 + 0.005 |
0.05 + 0.005 |
0.002 + 0.0005 |
400 mA 7 |
0.03 + 0.004 |
0.04 + 0.005 |
0.05 + 0.005 |
0.005 + 0.0005 |
1 A 6 |
0.03 + 0.02 |
0.04 + 0.02 |
0.05 + 0.02 |
0.005 + 0.001 |
3 A 5, 6 |
0.05 + 0.02 |
0.08 + 0.02 |
0.1 + 0.02 |
0.005 + 0.002 |
10 A 6 |
0.1 + 0.008 |
0.12 + 0.008 |
0.15 + 0.008 |
0.005 + 0.0008 |
5 Part of 10 A range.
6 Available on the front-panel terminal only.
7 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
Additional Current Errors
Digits |
NPLC |
Additional NPLC Noise Error for 1 mA, 100 mA, 400 mA, 3 A, and 10 A |
Additional NPLC Noise Error for 100 μA, 10 mA, 1 A |
---|---|---|---|
6½ |
100 |
0% of range |
0% of range |
6½ |
10 |
0% of range |
0% of range |
5½ |
1 |
0.001% of range |
0.01% of range |
5½ |
0.2 |
0.11% of range ±4 μA |
0.11% of range ±4 μA |
4½ |
0.02 |
0.04% of range ±4 μA |
0.28% of range ±4 μA |
AC Current
The following AC current specifications are for sinusoidal signals with amplitudes greater than 5% of range. For inputs from 1% to 5% of range, add an additional error of 0.1% of range.
Characteristic |
Description |
---|---|
Input Protection |
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V fuses, limits of 400 mA continuous 550 mA for 2 minutes on, 1 minute off |
Measurement Method |
AC-coupled true RMS, DC-coupled to the fuse and shunt (no blocking capacitor) |
AC Filter Bandwidth |
|
Slow |
3 Hz to 10 kHz |
Medium |
20 Hz to 10 kHz |
Fast |
200 Hz to 10 kHz |
Maximum Crest Factor |
5:1 at Full Scale |
Additional Crest Factor Errors (<100 Hz) |
Crest factor 1-2, 0.05% of full scale Crest factor 2-3, 0.2% of full scale Crest factor 3-4, 0.4% of full scale Crest factor 4-5, 0.5% of full scale Only applies to non-sinusoid signals |
Input Characteristics
Range |
Resolution |
Resolution |
Shunt Resistance (Ohms) |
Burden Voltage |
||
---|---|---|---|---|---|---|
4½ Digits |
5½ Digits |
6½ Digits |
||||
100 μA |
100.0000 μA |
10 nA |
1 nA |
100 pA |
100 Ω |
<0.015 V |
1 mA |
1.000000 mA |
100 nA |
10 nA |
1 nA |
100 Ω |
<0.15 V |
10 mA |
10.00000 mA |
1 μA |
100 nA |
10 nA |
1 Ω |
<0.025 V |
100 mA |
100.0000 mA |
10 μA |
1 μA |
100 nA |
1 Ω |
<0.25 V |
400 mA 9 |
400.000 mA |
100 μA |
10 μA |
1 μA |
1 Ω |
<0.50 V |
1 A 8 |
1.000000 A |
100 μA |
10 μA |
1 μA |
0.01 Ω |
<0.05 V |
3 A 5, 8 |
3.00000 A |
1 mA |
100 μA |
10 μA |
0.01 Ω |
<0.05 V |
10 A 8 |
10.00000 A |
1 mA |
100 μA |
10 μA |
0.01 Ω |
<0.5 V |
5 Part of 10 A range.
8 Available at front-panel connectors only.
9 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400 mA.
DMM4050/4040 Accuracy
Range |
Frequency |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|---|
Accuracy is given as ±(% measurement + % of range) |
|||||
100 μA |
3 - 5 Hz |
1.1 + 0.06 |
1.1 + 0.06 |
1.1 + 0.06 |
0.2 + 0.006 |
5 - 10 Hz |
0.35 + 0.06 |
0.35 + 0.06 |
0.35 + 0.06 |
0.1 + 0.006 |
|
10 Hz - 5 kHz |
0.15 + 0.06 |
0.15 + 0.06 |
0.15 + 0.06 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.35 + 0.7 |
0.35 + 0.7 |
0.35 + 0.7 |
0.03 + 0.006 |
|
1 mA |
3 - 5 Hz |
1.0 + 0.04 |
1.0 + 0.04 |
1.0 + 0.04 |
0.1 + 0.006 |
5 - 10 Hz |
0.3 + 0.04 |
0.3 + 0.04 |
0.3 + 0.04 |
0.035 + 0.006 |
|
10 Hz - 5 kHz |
0.1 + 0.04 |
0.1 + 0.04 |
0.1 + 0.04 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.2 + 0.25 |
0.2 + 0.25 |
0.2 + 0.25 |
0.03 + 0.006 |
|
10 mA |
3 - 5 Hz |
1.1 + 0.06 |
1.1 + 0.06 |
1.1 + 0.06 |
0.2 + 0.006 |
5 - 10 Hz |
0.35 + 0.06 |
0.35 + 0.06 |
0.35 + 0.06 |
0.1 + 0.006 |
|
10 Hz - 5 kHz |
0.15 + 0.06 |
0.15 + 0.06 |
0.15+ 0.06 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.35 + 0.7 |
0.35 + 0.7 |
0.35 + 0.7 |
0.03 + 0.006 |
|
100 mA |
3 - 5 Hz |
1.0 + 0.04 |
1.0 + 0.04 |
1.0 + 0.04 |
0.1 + 0.006 |
5 - 10 Hz |
0.3 + 0.04 |
0.3 + 0.04 |
0.3 + 0.04 |
0.035 + 0.006 |
|
10 Hz - 5 kHz |
0.1 + 0.04 |
0.1 + 0.04 |
0.1 + 0.04 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.2 + 0.25 |
0.2 + 0.25 |
0.2 + 0.25 |
0.03 + 0.006 |
|
400 mA 7 |
3 - 5 Hz |
1.0 + 0.1 |
1.0 + 0.1 |
1.0 + 0.1 |
0.1 + 0.006 |
5 - 10 Hz |
0.3 + 0.1 |
0.3 + 0.1 |
0.3 + 0.1 |
0.035 + 0.006 |
|
10 Hz - 5 kHz |
0.1 + 0.1 |
0.1 + 0.1 |
0.1 + 0.1 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.2 + 0.7 |
0.2 + 0.7 |
0.2 + 0.7 |
0.03 + 0.006 |
|
1 A 6 |
3 - 5 Hz |
1.0 + 0.04 |
1.0 + 0.04 |
1.0 + 0.04 |
0.1 + 0.006 |
5 - 10 Hz |
0.3 + 0.04 |
0.3 + 0.04 |
0.3 + 0.04 |
0.035 + 0.006 |
|
10 Hz - 5 kHz |
0.1 + 0.04 |
0.1 + 0.04 |
0.1 + 0.04 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.35 + 0.7 |
0.35 + 0.7 |
0.35 + 0.7 |
0.03 + 0.006 |
|
3 A 5, 6 |
3 - 5 Hz |
1.1 + 0.06 |
1.1 + 0.06 |
1.1 + 0.06 |
0.1 + 0.006 |
5 - 10 Hz |
0.35 + 0.06 |
0.35 + 0.06 |
0.35 + 0.06 |
0.035 + 0.006 |
|
10 Hz - 5 kHz |
0.15 + 0.06 |
0.15 + 0.06 |
0.15 + 0.06 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.35 + 0.7 |
0.35 + 0.7 |
0.35 + 0.7 |
0.03 + 0.006 |
|
10 A 6 |
3 - 5 Hz |
1.1 + 0.06 |
1.1 + 0.06 |
1.1 + 0.06 |
0.1 + 0.006 |
5 - 10 Hz |
0.35 + 0.06 |
0.35 + 0.06 |
0.35 + 0.06 |
0.035 + 0.006 |
|
10 Hz - 5 kHz |
0.15 + 0.06 |
0.15 + 0.06 |
0.15 + 0.06 |
0.015 + 0.006 |
|
5 - 10 kHz |
0.35 + 0.7 |
0.35 + 0.7 |
0.35 + 0.7 |
0.03 + 0.006 |
5 Part of 10 A range.
6 Available on the front-panel terminal only.
7 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
Additional Low-frequency Errors
Frequency |
AC Filter |
||
---|---|---|---|
3 Hz (Slow) |
20 Hz (Medium) |
200 Hz (Fast) |
|
Error is stated as % of reading. |
|||
10 - 20 Hz |
0 |
0.25 |
— |
20 - 40 Hz |
0 |
0.02 |
— |
40 - 100 Hz |
0 |
0.01 |
0.55 |
100 - 200 Hz |
0 |
0 |
0.2 |
200 Hz - 1 kHz |
0 |
0 |
0.02 |
>1 kHz |
0 |
0 |
0 |
Frequency
Characteristic |
Description |
---|---|
Gate Time |
Programmable to 1 s, 100 ms, and 10 ms |
Measurement Method |
Flexible counting technique. AC-coupled input using the AC voltage measurement function |
Settling Considerations |
When measuring frequency or period after a DC offset voltage change, errors may occur. For the most accurate measurement, wait up to 1 second for the input blocking capacitor to settle |
Measurement Considerations |
To minimize measurement errors, shield inputs from external noise when measuring low-voltage, low-frequency signals |
DMM4050/4040 Accuracy
Range |
Frequency |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|---|
Accuracy is given as ± % measurement |
|||||
100 mV to 1000 V 10, 11 |
3 - 5 Hz |
0.1 |
0.1 |
0.1 |
0.005 |
5 - 10 Hz |
0.05 |
0.05 |
0.05 |
0.005 |
|
10 - 40 Hz |
0.03 |
0.03 |
0.03 |
0.001 |
|
40 Hz - 300 kHz |
0.006 |
0.01 |
0.01 |
0.001 |
|
300 kHz - 1 MHz |
0.006 |
0.01 |
0.01 |
0.001 |
10 Limited to 8 × 107 Volt-Hertz
11 Input >100 mV. For 10 - 100 mV, multiply percent measurement error by 10.
Gate Time vs. Resolution
Gate Time |
Resolution |
---|---|
0.01 |
5½ |
0.1 |
6½ |
1.0 |
6½ |
Additional Low-frequency Errors
Frequency |
NPLC |
||
---|---|---|---|
6½ |
5½ |
4½ |
|
Error stated as percent of measurement for inputs >100 mV. For 10 - 100 mV, multiply percent by 10. |
|||
3 - 5 Hz |
0 |
0.12 |
0.12 |
5 - 10 Hz |
0 |
0.17 |
0.17 |
10 - 40 Hz |
0 |
0.2 |
0.2 |
40 - 100 Hz |
0 |
0.06 |
0.21 |
100 - 300 Hz |
0 |
0.03 |
0.21 |
300 Hz - 1 kHz |
0 |
0.01 |
0.07 |
>1 kHz |
0 |
0 |
0.02 |
Capacitance (DMM4050 Only)
Range |
Resolution |
1 Year Accuracy 12 (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|
Accuracy is stated as ±(% of measurement + % of range) |
|||
1 nF |
1 pF |
2% ±2.5% |
0.05 + 0.05 |
10 nF |
10 pF |
1% ±0.5% |
0.05 + 0.01 |
100 nF |
100 pF |
1% ±0.5% |
0.01 + 0.01 |
1 μF |
1 nF |
1% ±0.5% |
0.01 + 0.01 |
10 μF |
10 nF |
1% ±0.5% |
0.01 + 0.01 |
100 μF |
100 nF |
1% ±0.5% |
0.01 + 0.01 |
1 mF |
1 μF |
1% ±0.5% |
0.01 + 0.01 |
10 mF |
10 μF |
1% ±0.5% |
0.01 + 0.01 |
100 mF |
100 μF |
4% ±0.2% |
0.05 + 0.05 |
12 Stated accuracy is attained when Zero function is used.
Temperature (DMM4050 only)
Range |
Resolution |
Accuracy |
Temperature Coefficient/°C Outside 18 to 28 °C |
|
---|---|---|---|---|
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
|||
Test Current: 1 mA Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 Ω lead resistance. The accuracy listed in the table below are valid only when using the 4-wire RTD measurement function. Specifications do not include probe accuracy, which must be added. |
||||
–200 °C |
0.001 °C |
0.06 |
0.09 |
0.0025 |
–100 °C |
0.001 °C |
0.05 |
0.08 |
0.002 |
0 °C |
0.001 °C |
0.04 |
0.06 |
0.002 |
100 °C |
0.001 °C |
0.05 |
0.08 |
0.002 |
300 °C |
0.001 °C |
0.1 |
0.12 |
0.002 |
600 °C |
0.001 °C |
0.18 |
0.22 |
0.002 |
Additional Errors
Digits |
NPLC |
Additional NPLC Noise Error |
---|---|---|
6½ |
100 |
0 °C |
6½ |
10 |
0 °C |
5½ |
1 |
0.03 °C |
5½ |
0.2 |
0.12 °C |
4½ |
0.02 |
0.6 °C |
Continuity
Characteristic |
Description |
---|---|
Continuity Threshold |
Selectable between 1 Ω and 1000 Ω |
Test Current |
1 mA |
Response Time |
300 S/s with audible tone |
Range |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|
Accuracy is given as ±(% measurements + % of range) |
||||
1000.0 Ω |
0.002 + 0.01 |
0.008 + 0.02 |
0.01 + 0.02 |
0.001 + 0.002 |
Diode Test
Characteristic |
Description |
---|---|
Test Current |
100 μA or 1 mA |
Response Time |
300 S/s with audible tone |
Range |
24 Hour (23 °C ±1 °C) |
90 Days (23 °C ±5 °C) |
1 Year (23 °C ±5 °C) |
Temperature Coefficient/°C Outside 18 to 28 °C |
---|---|---|---|---|
Accuracy is given as ±(% measurements + % of range) |
||||
5.0000 V |
0.002 + 0.002 |
0.008 + 0.002 |
0.01 + 0.002 |
0.001 + 0.002 |
10.0000 V |
0.002 + 0.001 |
0.008 + 0.002 |
0.01 + 0.002 |
0.001 + 0.002 |
Measurement Rates (IEEE48816)
Function |
Digits |
Setting |
Integration Time 60 Hz (50 Hz) |
Measurements/Second13 |
|
---|---|---|---|---|---|
DMM4040 |
DMM4050 |
||||
DC Volts, DC Current, and Resistance |
6½ |
100 NPLC |
1.67 (2) s |
0.6 (0.5) |
0.6 (0.5) |
6½ |
10 NPLC |
167 (200) ms |
6 (5) |
6 (5) |
|
5½ |
1 NPLC |
16.7 (20) ms |
60 (50) |
60 (50) |
|
5½ |
0.2 NPLC |
3.3 ms |
270 |
270 |
|
4½ |
0.02 NPLC |
500 μs |
995 |
995 |
|
AC Voltage and AC Current14 |
6½ |
3 Hz |
|
0.47 |
0.47 |
6½ |
20 Hz |
|
1.64 |
1.64 |
|
6½ |
200 Hz15 |
|
4.5 |
4.5 |
|
Frequency and Period |
6½ |
1 s |
|
1 |
1 |
5½ |
100 ms |
|
9.8 |
9.8 |
|
4½ |
10 ms |
|
80 |
80 |
|
Capacitance |
6½ |
|
|
NA |
2 |
13 Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
14 Maximum measurement rates for 0.01% of AC step. When DC input varies, additional settling delay is required.
15 For remote operation or external trigger using default settling delay.
16 Note that the measurement rates for RS232 can vary depending on the baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurements. The LAN bus has a maximum measurement rate of 963 measurements.
Ordering Information
Models
Model |
Description |
---|---|
DMM4040 |
6.5 Digit Multimeter |
DMM4050 |
6.5 Digit Multimeter |
DMM4050/4040 Includes: Meter, TL710 Test Leads, Line Cord, Calibration Certificate, Calibration Data Report, Safety and Installation Instructions including Warranty Statement, CD-ROM with User Manual (English, French, Italian, German, Spanish, Simplified Chinese, Traditional Chinese, Korean, Russian, Japanese) and Connectivity Information, RS-232 to USB Adapter Cable, National Instruments LabVIEW SignalExpress™ Tektronix Edition Limited Edition Software.
Please specify power plug when ordering.
Instrument Options
Power Plug Options
Option |
Description |
---|---|
Opt. A0 |
North America |
Opt. A1 |
Universal Euro |
Opt. A2 |
United Kingdom |
Opt. A3 |
Australia |
Opt. A5 |
Switzerland |
Opt. A6 |
Japan |
Opt. A10 |
China |
Opt. A11 |
India |
Opt. A12 |
Brazil |
Opt. E1 |
Euro and UK power cords |
Service Options*17
Option |
Description |
---|---|
Opt. CA1 |
Single Calibration or Functional Verification |
Opt. R5 |
Repair Service 5 Years (including warranty) |
Opt. SILV100 |
Standard Warranty Extended to 5 Years |
*17 Test Leads and accessories are not covered by the DMM warranty and Service Offerings. Refer to the datasheet of each Test Lead and accessory model for its unique warranty and calibration terms.
Recommended Accessories and Software
Accessory |
Description |
---|---|
Calibration Manual |
077-0362-xx |
Programmer's Manual |
077-0363-xx |
TP750 |
100 Ω RTD Temperature Probe (DMM4050 only) |
196-3520-xx |
Premium Test Leads (replacement/spare for TL710) |
TL705 |
2×4 Wire Ohm 1000 V Precision Test Lead |
TL725 |
2×4 Wire Ohm SMD Test Tweezers |
ACD4000 |
Soft Transit Case |
HCTEK4321 |
Hard Carrying Case |
RMU2U |
Rackmount Shelf Kit for 1 or 2 Units |
013-0369-xx |
Calibration Fixture 4-terminal short |
196-3520-00 |
TL710 Premium Test Lead |
SIGEXPTE |
NI LabVIEW SignalExpress Tektronix Edition Software – Full Version |