Contact us

Live Chat with Tek representatives. Available 6:00 AM - 4:30 PM

Call

Call us at

Available 6:00 AM – 5:00 PM (PST) Business Days

Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

Why do I measure a different value between channel 1 and channel 2 on my 2182A?

Question :

Why do I measure a different value between channel 1 and channel 2 on my 2182A?

Answer :

When measuring the same voltage source, if you notice a difference in reported voltage between channel 1 and channel 2 of your nanovoltmeter, you might be seeing the effects of pumpout current. This effect is significant when measuring voltages <1V and the resistance between Channel 1 LO and Channel 2 LO is >100kΩ. When using using Channel 2 for measurements of <1 V, ensure both channel LOs are connected together. 

If you must use Channel 2 LO to measure <1V, and Channel 2 LO cannot be connected to Channel 1 LO (or the resistance must be >100kΩ) you can use Low Charge Injection Mode to reduce the pumpout current. However, this mode greatly increases measurement noise.

For more information, see the 2182A User's Manual section on pumpout current.


This FAQ Applies to:

Product Series: Nano vôn kế Keithley Model 2182A

Product:

FAQ ID 783015

View all FAQs »