質問:
What is the maximum number of samples that can be analyzed by TDSJIT3 Advanced/Essentials on the DPO7000 series scopes?
回答:
TDSJIT3 Advanced/Essentials V3.0.0.7 is limited to a testing a maximum of 64 M on a DPO7000 series scopes with longer waveform record length? You can select the longer records while running TDSJIT3, but only the first 64 M will be analyzed by the application.
FAQの該当製品:
製品シリーズなし
製品:
FAQ ID 64181