お問い合わせ
ダウンロード
マニュアル、データシート、ソフトウェアなどのダウンロード:
フィードバック
Features and Benefits
- 9 kHz - 6 GHz frequency range
- 12 dB gain (nominal)
- 10 dB DANL improvement
Applications
- Measurement of low-power signals
TPA-N-PRE Preamplifier
The TPA-N-PRE preamplifier accessory for the MDO4000 and MDO3000 Series Mixed Domain Oscilloscopes provides 12 dB nominal gain, further lowering the already low noise floor in the MDO4000 and MDO3000 Series. This allows you to pull very small signals out of the noise in low-amplitude signal applications. For example, in the following screenshot we are measuring a noise floor of -114 dBm (-154 dBm/Hz) without the TPA-N-PRE preamplifier. It appears that there is only noise present.
In the screenshot below, we’ve turned on the preamplifier. Notice that the noise floor has dropped to -124 dBm (-164 dBm/Hz). As a result of the noise floor dropping significantly, we’re now able to see a -116 dBm signal present at 900 MHz that was hidden in the noise without the preamplifier.
Specifications
General
Specification | Description |
---|---|
Frequency range | Preamp alone: 9 kHz to 6 GHz Preamp and MDO: Frequency range specified by the MDO's frequency range |
Preamp gain | Amplifying state: +12 dB (nominal) Bypass state: -1.5 dB (nominal) |
Displayed average noise level (DANL) with the preamp attached to the MDO RF input | With the preamp mode set to “Auto” and the reference level set to -40 dBm (MDO4000 Series) or -30 dBm (MDO3000 Series): MDO4000C with SA6 option: 9 kHz to 50 kHz: 50 kHz to 5 MHz: 5 MHz to 400 MHz: 400 MHz to 3 GHz: 3 GHz to 4 GHz: 4 GHz to 6 GHz: MDO4000C with SA3 option: 9 kHz to 50 kHz: 50 kHz to 5 MHz: 5 MHz to 400 MHz: 400 MHz to 3 GHz: MDO3000 with MDO3SA option: 9 kHz to 50 kHz: 50 kHz to 5 MHz: 5 MHz to 2 GHz: 2 GHz to 3 GHz: |
Level measurement uncertainty with the preamp attached to the MDO RF input | Preamp mode set to “Auto”. Reference Level 10 dBm to –40 dBm. Input level ranging from reference level to 30 dB below reference level. Specifications exclude mismatch error. |
Maximum operating input level with preamp attached to the MDO RF input | Average continuous power: +30 dBm (1 W) (MDO4000 Series), +20 dBm (100 mW) (MDO3000 Series) DC maximum before damage: ±20 V DC Maximum power before damage (CW): +30 dBm (1 W) Maximum power before damage (pulse): +45 dBm (32 W) ( |
Reference level range with preamp attached to MDO RF input | Amplifying state: -30 dBm to DANL (MDO4000 Series), -20 dBm to DANL (MDO3000 Series) Bypass state: +30 dBm to DANL (MDO4000 Series), +20 dBm to DANL (MDO3000 Series) |
Connector type | SMA – female (outside threads) |
Environmental
Specification | Description |
---|---|
Temperature | |
Operating | 0 °C to +50 °C |
Nonoperating | -20 °C to +60 °C |
Humidity | |
Operating | High: 40 °C to 50 °C (104 °F to 122 °F), 10% to 60% RH Low: 0 °C to 40 °C (32 °F to 104 °F) 10% to 90% RH |
Nonoperating | High: 40 °C to 60 °C (104 °F to 122 °F), 10% to 60% RH Low: 0 °C to 40 °C (32 °F to 104 °F) 5% to 90% RH |
Attitude | |
Operating | Up to 3,000 meters |
Nonoperating | Up to 12,000 meters |
Power Requirements
The TPA-N-PRE is powered directly from the RF input on MDO4000 and MDO3000 Series oscilloscopes.
Regulatory
Compliance labeling: WEEE (European Union).
Recommended Oscilloscopes
MDO4000 and MDO3000 Mixed Domain Oscilloscopes
Note:
For best probe support, download and install the latest version of the oscilloscope firmware from www.tektronix.com.
Warranty information
One year parts and labor.
Ordering Information
TPA-N-PRE
MDO4000 and MDO3000 Pre-Amplifier. Includes Instruction Manual (English) (071-3058-XX)
Service Options
- Opt. C3 –
- Calibration Service 3 years (initial certification, plus 2 calibrations).
- Opt. C5 –
- Calibration Service 5 years (initial certification, plus 4 calibrations).
- Opt. R3 –
- Repair Service. Repair warranty extended to cover 3 years.
- Opt. R5 –
- Repair Service. Repair warranty extended to cover 3 years.
Additional Service Products Available During Warranty (DW)
- TPA-N-PRE R3DW –
- Repair Service Coverage 3 Years (includes product warranty period). 3-year period starts at time of customer instrument purchase
- TPA-N-PRE R5DW –
- Repair Service Coverage 5 Years (includes product warranty period). 5-year period starts at time of customer instrument purchase