Question :
The data sheet for my oscilloscope says that the vertical input sensitivity is 1 mV/div to 10 V/div in 1 MΩ mode. But I turn the Vertical Scale control fully clockwise and I can only get down to 10 mV/div. What am I doing wrong?
Answer :
You probably aren't doing anything wrong. But you do have a 10X probe attached to that channel, right?
The 1 mV/div specs relate to the oscilloscope inputs. If the signal’s source impedance is relatively low (<< 1 MOhm), the signal’s frequency is relatively low (less than a few MHz), and the voltage is within the input range of the oscilloscope, you can connect the signal directly to the oscilloscope with a cable and the vertical scale does go down to 1 mV/div in 1 MOhm mode.
However, when you attach an attenuating probe, the mV/div scale factor always increases by that same factor. This is a standard feature of oscilloscopes that automatically sense the probe attenuation.
This FAQ Applies to:
Product Series: DPO7000 High Voltage Probe: Single-Ended MDO3000 Mixed Domain Oscilloscope MDO4000C Mixed Domain Oscilloscope MSO/DPO70000DX Mixed Signal/Digital Phosphor Oscilloscopes MSO3000 / DPO3000 MSO/DPO4000 Mixed Signal Oscilloscope (Discontinued) MSO/DPO5000B Mixed Signal Oscilloscope (Discontinued) Passive Probe