질문:
Do you have a SCPI Sequence for 100V @ 10A on Model 2461 SMU?
답변:
Here is a sequence of SCPI commands the configure the Model 2461 SMU to output a 100V pulse up to 10A.
- RST
SYSTem:CLEar
TRACe:MAKE "PulseBuffer", 50
SOURce:FUNCtion CURRent
SOURce:CURRent:RANGe:AUTO OFF
SOURce:CURRent:RANGe 1
SOURce:CURRent 0
SOURce:CURRent:READ:BACK OFF
SOURce:CURRent:DELay:AUTO OFF
SOURce:CURRent:VLIMit 105
SENSe:FUNCtion "VOLTage"
SENSe:VOLTage:RANGe:AUTO OFF
SENSe:VOLTage:RANGe 100
SENSe:VOLTage:NPLCycles 0.01
SENSe:VOLTage:RSENse ON
ROUTe:TERMinals FRONt
SENSe:VOLTage:AZERo OFF
SOURce:PULSe:TRain:CURRent 0, 10, .001, 21, ON, "PulseBuffer", 0, .019, 2, 100, ON
--bias level, pulse level, pulse width, count, measEnable, bufferName, delay, offtime, xBiasLimit, xPulseLimit, FailAbort
INITiate:IMMediate
TRACe:DATA? 1, 21, "PulseBuffer"
FAQ ID 773831