연락처

텍트로닉스 담당자와 실시간 상담 6:00am-4:30pm PST에 이용 가능

전화

전화 문의

9:00am-6:00PM KST에 이용 가능

다운로드

매뉴얼, 데이터 시트, 소프트웨어 등을 다운로드할 수 있습니다.

다운로드 유형
모델 또는 키워드

피드백

I want to trigger on the primary frequency (perhaps 60 Hz)+ this frequency is modulated with the motor frequency (perhaps 600 Hz) with an amplitude of perhaps a third of the primary frequency's amplitude. The scope will not adequately trigger on the motor current signal.

질문:

I want to trigger on the primary frequency (perhaps 60 Hz)+ this frequency is modulated with the motor frequency (perhaps 600 Hz) with an amplitude of perhaps a third of the primary frequency's amplitude. The scope will not adequately trigger on the motor current signal.

답변:

This signal creates Multiple zero crossings which makes the triggering unstable. Turn down the amplitude of the composite signal (ch1) so that the scope's built in hysteresis will swamp out the motor drive zero crossings. This channel will be used to trigger the scope, a second channel can then be used to view the composite waveform with greater sensitivity. The proceedure, Reduce amplitude on ch1 and then view the same wfm on ch2. (Channel 1 can be turned off but will still trigger channel 2.)

이 FAQ는 다음에 적용 됩니다:

제품 시리즈: TPS2000

제품:

FAQ ID 58436

모든 FAQ 보기 »