질문:
I really like the idea of a removable hard disk drive for my oscilloscope+ but I’m concerned about losing the instrument calibration information. If I replace the original HDD+ will the cal constants be lost+ or is there a file that I need to transfer to the new HDD+ or are the constants stored in NVRAM?
답변:
Factory calibration data, serial number, and option keys are stored in the NVRAM.
Signal Path Compensation (SPC) data is stored on the hard drive. To be sure that your oscilloscope performance is optimized for the current temperature and to remove long-term drift, you should re-run SPC (under Utilities -> Instrument Calibration…).
이 FAQ는 다음에 적용 됩니다:
제품 시리즈: MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO5000B 혼합 신호 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 DPO7000 DPO7000 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 MSO/DPO70000DX 혼합 신호/디지털 포스퍼 오실로스코프 DPO7000 DPO7000