I am trying to control my TDS400+ TDS500+ TDS600 and TDS700 series product using the RS232 port and it is not working. Why?
답변:
The RS232 port on the TDS400, TDS500, TDS600 and TDS700 series products is a hardcopy port only and is not capable of controlling the instrument. Control of the instrument can only be accomplished through the GPIB port.