SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Models 2634B, 2635B, and 2636B System SourceMeter™ instrument. Specifications are the standards against which the 2634B, 2635B, and 2636B are tested. Upon leaving the factory, the 2634B, 2635B, and 2636B meet these specifications. Supplemental and typical values are nonwarranted, apply at 23 °C, and are provided solely as useful information.
Source and measurement accuracies are specified at the 2634B, 2635B, and 2636B terminals under these conditions:
- 18 °C to 28 °C, < 70 percent relative humidity
- After a two-hour warm-up period
- Speed normal (1 NPLC)
- A/D autozero enabled
- Remote sense operation or properly zeroed local operation
- Calibration period: One year
DC POWER SPECIFICATIONS
Voltage | Current | |
Maximum output power and source/sink limits |
30.3 W maximum per channel
|
30.3 W maximum per channel
|
Refer to Pulse Characteristics for pulsing details, such as duty cycle and pulse width.
1 Full power source operation regardless of load or number of channels used to 23 °C ambient temperature. Above 23 °C or power sink operation, refer to “Operating Boundaries” in the Series 2600B System SourceMeter Reference Manual for additional power derating information.
VOLTAGE ACCURACY SPECIFICATIONS2, 3
Range |
Source | Measure | |||
Programming resolution | Accuracy ± (% reading + volts) |
Typical noise (peak to peak) 0.1 Hz to 10 Hz |
Display resolution | Accuracy4 ± (% reading + volts) |
|
200 mV | 5 µV | 0.02% + 375 µV | 20 µV | 100 nV | 0.015% + 225 µV |
2 V | 50 µV | 0.02% + 600 µV | 50 µV | 1 µV | 0.02% + 350 µV |
20 V | 500 µV | 0.02% + 5 mV | 300 µV | 10 µV | 0.015% + 5 mV |
200 V | 5 mV | 0.02% + 50 mV | 2 mV | 100 µV | 0.015% + 50 mV |
CURRENT ACCURACY SPECIFICATIONS2
Range |
Source | Measure | |||
Programming resolution |
Accuracy ± (% reading + amperes) |
Typical noise (peak to peak) 0.1 Hz to 10 Hz |
Display resolution |
Accuracy4 |
|
100 pA5 | N/A | N/A | N/A | 100 aA | 0.15% + 120 fA6, 7 |
1 nA | 20 fA | 0.15% + 2 pA | 800 fA | 1 fA | 0.15% + 240 fA6, 8 |
10 nA | 200 fA | 0.15% + 5 pA | 2 pA | 10 fA | 0.15% + 3 pA |
100 nA | 2 pA | 0.06% + 50 pA | 5 pA | 100 fA | 0.06% + 40 pA |
1 µA | 20 pA | 0.03% + 700 pA | 25 pA | 1 pA | 0.025% + 400 pA |
10 µA | 200 pA | 0.03% + 5 nA | 60 pA | 10 pA | 0.025% + 1.5 nA |
100 µA | 2 nA | 0.03% + 60 nA | 3 nA | 100 pA | 0.02% + 25 nA |
1 mA | 20 nA | 0.03% + 300 nA | 6 nA | 1 nA | 0.02% + 200 nA |
10 mA | 200 nA | 0.03% + 6 µA | 200 nA | 10 nA | 0.02% + 2.5 µA |
100 mA | 2 µA | 0.03% + 30 µA | 600 nA | 100 nA | 0.02% + 20 µA |
1 A | 20 µA | 0.05% + 1.8 mA | 70 µA | 1 µA | 0.03% + 1.5 mA |
1.5 A | 50 µA | 0.06% + 4 mA | 150 µA | 1 µA | 0.05% + 3.5 mA |
10 A9 | 200 µA | 0.5% + 40 mA | N/A | 10 µA | 0.4% + 25 mA |
2 For temperatures 0 °C to 18 °C and 28 °C to 50 °C, accuracy is degraded by ± (0.15 × accuracy specification)/°C. High-Capacitance Mode accuracy is applicable at 23 °C ± 5 °C.
3 Add 50 µV to source accuracy specifications per volt of HI lead drop.
4 Derate accuracy specification for NPLC setting < 1 by increasing the error term. Add appropriate typical percent of reading term for resistive loads using the table below.
NPLC setting | 200 mV range | 2 V and 200 V ranges | 100 nA range | 1 µA to 100 mA ranges | 1 A to 1.5 A ranges |
0.1 | 0.01% | 0.01% | 0.01% | 0.01% | 0.01% |
0.01 | 0.08% | 0.07% | 0.1% | 0.05% | 0.05% |
0.001 | 0.8% | 0.6% | 1% | 0.5% | 1.1% |
6 10-NLPC, 11-Point Median Filter, < 200 V range, measurements made within 1 hour after zeroing. 23 ºC ± 1 ºC.
7 Under default specification conditions: ± (0.15% + 750 fA).
8 Under default specification conditions: ± (0.15% + 1 pA).
SUPPLEMENTAL CHARACTERISTICS
The following specifications are supplemental characteristics that provide additional information about instrument functions and performance. These characteristics are nonwarranted specifications; they describe the typical performance of the 2634B, 2635B, and 2636B.
PULSE CHARACTERISTICS
Pulse region specifications |
|||
Region quadrant diagram | Region maximums | Maximum pulse width10 | Maximum duty cycle11 |
1 | 100 mA at 200 V | DC, no limit | 100% |
1 | 1.5 A at 20 V | DC, no limit | 100% |
2 | 1 A at 180 V | 8.5 ms | 1% |
312 | 1 A at 200 V | 2.2 ms | 1% |
4 | 10 A at 5 V | 1 ms | 2.2% |
10 Times measured from the start of pulse to the start off-time; see figure below
11 Thermally limited in sink mode (quadrants 2 and 4) and ambient temperatures above 23 °C. See power equations in the Series 2600B System SourceMeter Reference Manual for more information.Minimum programmable pulse width9 |
100 µs
|
|||
Source value | Load | Source settling time (% of range) | Minimum pulse width | |
5 V | 0.5 Ω | 1% | 300 µs | |
20 V | 200 Ω | 0.2% | 200 µs | |
180 V | 180 Ω | 0.2% | 5 ms | |
200 V (1.5 A Limit) | 200 Ω | 0.2% | 1.5 ms | |
100 mA | 200 Ω | 1% | 200 µs | |
1 A | 200 Ω | 1% | 500 µs | |
1 A | 180 Ω | 0.2% | 5 ms | |
10 A | 0.5 Ω | 0.5% | 300 µs | |
Pulse width programming resolution | 1 µs | |||
Pulse width programming accuracy | ± 5 µs |
|||
Pulse width jitter | 2 µs |
ADDITIONAL SOURCE CHARACTERISTICS
Noise 10 Hz to 20 MHz |
< 20 mV peak to peak, < 3 mV RMS
|
|
Transient response time | < 70 µs for the output to recover to within 0.1% for a 10% to 90% step change in load | |
Overshoot |
Voltage:
Current:
|
|
Range change overshoot |
Voltage:
Current:13
|
|
Guard offset voltage |
< 4 mV
|
|
Remote sense operating range14 |
Maximum voltage between HI and SENSE HI = 3 V Maximum voltage between LO and SENSE LO = 3 V |
|
Voltage output headroom |
200 V range
20 V range
|
|
Overtemperature protection | Internally sensed temperature overload puts the instrument in standby mode | |
Limit/compliance |
Bipolar limit (compliance) set with a single value Voltage:15
Current:16
|
|
Voltage source output settling time | Time required to reach within 0.1% of final value after source level command is processed on a fixed range. | |
Voltage range | Settling time | |
200 mV | < 50 µs | |
2 V | < 50 µs | |
20 V | < 110 µs | |
200 V | < 700 µs | |
Current source output settling time |
Time required to reach within 0.1% of final value after source level command is processed on a fixed range
|
|
Current range | Settling time | |
1.5 A and 1 A | < 120 µs (RLOAD > 6 Ω) | |
100 mA and 10 mA | < 80 µs | |
1 mA | < 100 µs | |
100 µA | < 150 µs | |
10 µA | < 500 µs | |
1 µA | < 2 ms | |
100 nA | < 20 ms | |
10 nA | < 40 ms | |
1 nA | < 150 ms |
ADDITIONAL MEASUREMENT CHARACTERISTICS
Contact check specifications17, 18 |
Speed | Maximum measurement time to memory for 60 Hz (50 Hz) |
Accuracy (1 year) 23 °C ± 5 °C ± (% reading + ohms) |
Fast | 1.1 ms (1.2 ms) | 5% + 10 Ω | |
Medium | 4.1 ms (5 ms) | 5% + 1 Ω | |
Slow | 36 ms (42 ms) | 5% + 0.3 Ω | |
Current measure settling time19 |
Time required to reach within 0.1% of final value after source level command is processed on a fixed range Values below for VOUT = 2 V |
||
Current range | Settling time |
||
1 mA | < 100 µs |
||
Input impedance | > 100 TΩ |
17 Includes measurement of SENSE HI to HI and SENSE LO to LO contact resistances.
18 Contact check is not available with the Model 2634B.
19 Compliance equal to 100 mA. Delay factor set to 1.
ADDITIONAL CHARACTERISTICS
Maximum load impedance | Normal mode 10 nF |
High‑capacitance
mode 50 µF |
Common mode voltage | 250 V dc |
|
Common mode isolation | > 1 GΩ < 4500 pF |
|
Sense high input impedance | > 100 TΩ |
|
Maximum sense lead resistance | 1 kΩ for rated accuracy |
|
Overrange | 101% of source range 102% of measure range |
HIGH-CAPACITANCE MODE20, 21, 22
Accuracy specifications | Accuracy specifications are applicable in both normal and high-capacitance modes. |
|
Voltage source output settling time |
Time required to reach within 0.1% of final value after source level command is processed on a fixed range. Current limit = 1 A |
|
Voltage range | Settling time with CLOAD = 4.7 µF | |
200 mV | < 600 µs | |
2 V | < 600 µs | |
20 V | < 1.5 ms | |
200 V | < 20 ms | |
Current measure settling time |
Time required to reach within 0.1% of final value after source level command is processed on a fixed range Values below for VOUT = 2 V unless noted |
|
Current range | Settling time | |
1.5 A and 1A | < 120 µs (RLOAD > 6 Ω) | |
100 mA and 10 mA | < 100 µs | |
1 mA | < 3 ms | |
100 µA | < 3 ms | |
10 µA | < 230 ms | |
1 µA | < 230 ms | |
Capacitor leakage performance using HIGH-C scripts23 |
200 ms at 50 nA
|
|
Mode change delay |
Current ranges of 100 µA and above:
Current ranges below 100 µA:
|
|
Voltmeter input impedance | 30 GΩ in parallel with 3300 pF |
|
Noise 10 Hz to 20 MHz |
< 30 mV peak to peak
|
|
Range change overshoot |
Voltage:
|
20 High-Capacitance Mode specifications are for dc measurements only.
21 100 nA range is not available in High-Capacitance Mode.
22 High-Capacitance Mode uses locked ranges. Autorange is disabled.
23 Part of KI Factory scripts. See the Series 2600B Reference Manual for details.
MEASUREMENT SPEED CHARACTERISTICS24, 25, 26
Maximum sweep operation rates (operations per second) for 60 Hz (50 Hz):
A/D converter speed | Trigger origin | Measure to memory (using user scripts) |
Measure to GPIB (using user scripts) |
Source measure to memory (using user scripts) |
Source measure to GPIB (using user scripts) |
Source measure to memory (using sweep API) |
Source measure to GPIB (using sweep API) |
0.001 NPLC | Internal | 20000 (20000) |
9800 (9800) |
7000 (7000) |
6200 (6200) |
12000 (12000) |
5900 (5900) |
0.001 NPLC | Digital I/O | 8100 (8100) |
7100 (7100) |
5500 (5500) |
5100 (5100) |
11200 (11200) |
5700 (5700) |
0.01 NPLC | Internal | 4900 (4000) |
3900 (3400) |
3400 (3000) |
3200 (2900) |
4200 (3700) |
4000 (3500) |
0.01 NPLC | Digital I/O | 3500 (3100) |
3400 |
3000 (2700) |
2900 (2600) |
4150 (3650) |
3800 (3400) |
0.1 NPLC | Internal | 580
(480) |
560
(470) |
550
(465) |
550
(460) |
560
(470) |
545
(460) |
0.1 NPLC | Digital I/O | 550
(460) |
550
(460) |
540
(450) |
540
(450) |
560
(470) |
545
(460) |
1.0 NPLC | Internal | 59 (49) |
59 (49) |
59 (49) |
59 (49) |
59 (49) |
59 (49) |
1.0 NPLC | Digital I/O | 59 (49) |
59 (49) |
59 (49) |
59 (49) |
59 (49) |
59 (49) |
Maximum single measurement rates (operations per second) for 60 Hz (50 Hz):
A/D converter speed | Trigger origin | Measure to GPIB | Source measure to GPIB | Source measure pass/fail to GPIB |
0.001 NPLC | Internal | 1900 (1800) | 1400 (1400) | 1400 (1400) |
0.01 NPLC | Internal | 1450 (1400) | 1200 (1200) | 1100 (1100) |
0.1 NPLC | Internal | 450 (390) | 425 (370) | 425 (375) |
1.0 NPLC | Internal | 58 (48) | 57 (48) | 57 (48) |
Maximum measurement range change rate | > 7000 per second for > 10 µA. When changing to or from a range ≥ 1 A, maximum rate is >> 2200/second. |
Maximum source range change rate | > 400 per second > 10 µA. When changing to or from a range ≥ 1 A, maximum rate is >> 190/second. |
Maximum source function change rate | > 1000 per second |
Command processing time |
< 1 ms
|
24 Tests performed with a Model 2636B using the following equipment: Computer hardware (Intel® Pentium® 4 2.4 GHz, 2 GB RAM, National Instruments™ PCI-GPIB); driver (NI-488.2 Version 2.2 PCI-GPIB); software (Microsoft® Windows® XP, Microsoft® Visual Studio® 2010, VISA™ version 4.1).
25 Exclude current measurement ranges less than 1 mA.
26 With default measurement delays and filters disabled.
TRIGGERING AND SYNCHRONIZATION CHARACTERISTICS
Triggering
Trigger in to trigger out | 0.5 μs |
Trigger in to source change27 | 10 μs |
Trigger timer accuracy | ±2 μs |
Source change after LXI trigger | 280 μs |
Synchronization
Multi-node synchronized source change27 | < 0.5 μs |
Single-node synchronized source change27 | < 0.5 μs |
27 Fixed source range with no polarity change.
SUPPLEMENTAL INFORMATION
Front-panel interface | Two-line vacuum fluorescent display (VFD) with keypad and navigation wheel |
Display |
|
Keypad operations |
|
Programming | Embedded Test Script Processor (TSP™) accessible from any host interface; responds to high-speed test scripts comprised of remote commands and statements (for example, branching, looping, and math); able to execute test scripts stored in memory without host intervention. |
Minimum user memory available | 16 MB (approximately 250,000 lines of TSP code) |
Reading buffers |
Nonvolatile memory uses dedicated storage areas reserved for measurement data. Reading buffers are arrays of measurement elements. Each element can store the following items:
Reading buffers can be filled using the front-panel STORE key, and retrieved using the RECALL key or host interface. |
Buffer size, with timestamp and source setting | > 60,000 samples |
Buffer size, without timestamp and source setting | > 140,000 samples |
TIMING
Timer | Free-running 47-bit counter with 1 MHz clock input. Reset each time instrument power is turned on. If the instrument is not turned off, the timer is automatically reset to zero (0) every four years. |
Timestamp | TIMER value is automatically saved when each measurement is triggered |
Resolution | 1 μs |
Timestamp accuracy | ±100 ppm |
GENERAL SPECIFICATIONS
IEEE-488 |
IEEE Std 488.1 compliant. Supports IEEE Std 488.2 common commands and status model topology |
RS-232 |
|
Ethernet | RJ-45 connector, LXI version 1.4 Core 2011, 10/100BaseT, Auto-MDIX |
LXI compliance | LXI version 1.4 Core 2011 |
Expansion interface28 |
|
USB Control (Rear) | USB 2.0 Device: USB-TMC488 protocol |
USB File System (Front) | USB 2.0 Host: Mass storage class device |
Power supply | 100 V ac to 240 V ac, 50 Hz or 60 Hz (autosensing), 240 VA maximum |
Cooling | Forced air; side intake and rear exhaust. One side must be unobstructed when rack mounted. |
Warranty | 1 year |
EMC | Conforms to European Union EMC Directive |
Safety |
NRTL listed to UL61010-1:2008 and CSA C22.2 No. 61010-1 Conforms to European Union Low Voltage Directive |
Environment |
For indoor use only Altitude: Maximum 2000 m (6562 ft) above sea level Operating: 0 °C to 50 °C, 70% relative humidity up to 35 °C. Derate 3% relative humidity/°C, 35 °C to 50 °C Storage: −25 °C to 65 °C |
Dimensions |
Rack mount: 89 mm high × 213 mm wide × 460 mm deep (3.5 in. × 8.4 in. × 17.5 in.) Bench configuration (with handle and feet): 104 mm high × 238 mm wide × 460 mm deep (4.1 in. × 9.4 in. × 17.5 in.) |
Weight |
2635B: 4.75 kg (10.4 lb) 2634B and 2636B: 5.50 kg (12.0 lb) |
28 TSP-Link is not available with the Model 2634B.
Digital I/O Interface
Digital I/O interface29 | |
Connector: 25-pin female D | |
Input/output pins: 14 open drain I/O bits | |
Absolute maximum input voltage: 5.25 V | |
Absolute minimum input voltage: −0.25 V | |
Maximum logic low input voltage: 0.7 V, +850 µA maximum | |
Minimum logic high input voltage: 2.1 V, +570 µA | |
Maximum source current (flowing out of digital I/O bit): +960 µA | |
Maximum sink current at maximum logic low voltage (0.7 ): −5.0 mA | |
Absolute maximum sink current (flowing into digital I/O pin): −11 mA | |
5 V power supply pins: Limited to 250 mA total, solid-state fuse protected | |
Safety interlock pin: A signal of > 3.4 V at 24 mA (at an absolute maximum of 6 V) must be externally applied to ensure 200 V operation. Connect the 5 V output and the interlock input of the 25-pin digital I/O connector on the back of the 2634B, 2635B, or 2636B to the switch in your fixture. The output will be disabled when the interlock signal is < 0.45 V. Absolute maximum input is −0.4 V to +6.0 V. |