연락처
텍트로닉스 담당자와 실시간 상담 6:00am-4:30pm PST에 이용 가능
전화 문의
9:00am-6:00PM KST에 이용 가능
다운로드
매뉴얼, 데이터 시트, 소프트웨어 등을 다운로드할 수 있습니다.
피드백
DC Coupled Current Probe
TCP202
이 데이터 시트의 제품은 텍트로닉스에서 더 이상 판매하지 않습니다.
리퍼브 테스트 장비에 대한 Tektronix Encore를 확인하십시오.
이러한 제품에 대한 지원 및 보증 상태를 살펴보십시오.
Features & Benefits
- Easy Current Measurements
- Probe Connects Directly to TDS Scope through TEKPROBE™ BNC Interface
- Split Core Construction for Easy Circuit Connection
- DC to 50 MHz Bandwidth
- 15 Amps DC + Peak AC Current
- 50 Amps Peak Pulse Current
- 500 × 10-6 Ampere-Sec Product
- Safety Certified
Applications
- Motor Drives
- Switching Power Supplies
- Disk Drive
- Electronic Ballasts
- Inverters
- Avionics
TCP202 DC Coupled Current Probe
When combined with a Tektronix TDS oscilloscope, the TCP202 enables users to make direct readout DC and AC current measurements, resulting in accurately displayed and recorded measurements.
The TCP202 is a general-purpose probe for displaying and measuring current in electronic circuits. It is ideal for power supply and motor drive design, and for device testing.
Recommended Oscilloscopes
All TDS3000C and TDS5000B Series oscilloscopes. Also operational with TDS500, TDS600, and TDS700 Series, recommended with use of calibration fixture (015-0672-50) for best accuracy.
For TCP202 operation with other TDS400, TDS500, TDS600, and TDS700 Series oscilloscopes, the scale factor will be correct but measurement will be indicated as V instead of A.
Increase the Accuracy of Your Current Measurements
The accuracy of the TCP202 AC/DC current probe is normally 3%. However, the accuracy may be improved to 1% when the probe is calibrated on certain TDS oscilloscope models using the optional Calibrator P/N 015-0672-50. This improved accuracy is available on the TDS500, TDS600, and TDS700 Series, however the optional calibrator is only functional with the TDS500, TDS600, or TDS700 Series oscilloscope calibrator pins. The calibration fixture is not required for TDS3000C and TDS5000B Series oscilloscopes.
Characteristics
Characteristic |
Description |
---|---|
Bandwidth |
DC to 50 MHz |
Max DC + Peak AC Current |
15 A |
Max Peak Pulse Current |
50 A |
Minimum Sensitivity |
10 mA/div |
DC Accuracy |
±1% (with probe calibrator) ±3% (without probe calibrator) |
Max Bare-wire Voltage |
300 V (CAT I) |
Typical Characteristics
Characteristic |
Description |
---|---|
Insertion Impedance |
0.002 Ω at 1 kHz 0.1 Ω at 5 MHz |
Propagation Delay |
17 ns |
Max Conductor Diameter |
3.8 mm (0.15 in.) |
Electrostatic Immunity |
(per IEC 801-2) >8 kV |
Ampere-Sec Product |
500 × 10-6 Amp × Sec |
Rise Time |
≤7.0 ns |
Physical Characteristics
Characteristic |
Description |
---|---|
Weight |
0.19 kg (0.44 lb.) |
Cable Length |
2.2 m (7.2 ft.) |
Power Requirements |
TCP202 is powered by the TDS oscilloscope through the TEKPROBE interface |
Safety
UL 3111-1, CSA1010.1, EN61010-1, IEC61010-1, EN61010-2-032, IEC61010-2-032 |
Ordering Information
TCP202
DC Coupled Current Probe.
Includes: Instruction manual (070-9542-xx), reference manual (070-9543-xx).
Recommended Accessories
Accessory |
Description |
---|---|
015-0672-50 |
Current Loop Calibrator for TDS500, TDS600, and TDS700 Series oscilloscopes |
Related Products
Product |
Description |
---|---|
1103 |
TEKPROBE® Power Supply (for scopes without TEKPROBE BNC interfaces) |
ADA400A |
Differential Preamplifier |
P5205 |
High-voltage Differential Probe |
P5210 |
High-voltage Differential Probe |
Service Options
Option |
Description |
---|---|
Opt. SILV400 |