Question :
What is the difference between “Trigger out” and “Sync out”?
Réponses :
Trigger and Sync Out are almost the same. They are only different in the burst mode, in inifinte-cycle mode.
Figure 1: Trigger out mode
Scope CH1 ->AFGCH1
Scope CH2 ->AFGTrigger out
Scope CH3 ->AFGTrigger in(with T connector)
In this case a Trigger out signal will happen every cycle. For frequencies higher than 5 MHz Trigger out will not happen every cycle but will be in sync with the output waveform.
Figure 2: Sync out mode
Scope CH1 ->AFGCH1
Scope CH2 ->AFGSync out
Scope CH3 ->AFGTrigger in(with T connector)
The Sync out is only one pulse. The Sync mode is used for creating a Master-Slave setup with two AFG products.
Cette FAQ concerne :
Série du produit : Générateur de fonctions arbitraires série AFG3000C Générateur de fonctions arbitraires série AFG3000C Générateur de fonctions arbitraires série AFG3000C Générateur de fonctions arbitraires série AFG3000C Générateur de fonctions arbitraires série AFG3000C Générateur de fonctions arbitraires série AFG3000C Générateur de fonctions arbitraires série AFG3000C
Produit :
Numéro de la FAQ 56091
Afficher toutes les FAQ »