Do the signal outputs of the AFG3000 have a short circuit protection?

Question :

Do the signal outputs of the AFG3000 have a short circuit protection?

Réponses :

There is no need for a dedicated short-circuit protection because the output circuits are immune from a short circuit of the output. However, the outputs can be protected against over-voltage from an external source with a fuse that is included as a standard accessory.

Numéro de la FAQ 56461