Contact us
Call us at
Available 6:00 AM – 5:00 PM (PST) Business Days
Download
Download Manuals, Datasheets, Software and more:
Feedback
AC/DC Current Probe
TCP0150 Datasheet
The TCP0150 is a high performance and easy-to-use 150 A AC/DC current probe designed for use and direct connection to the TekVPI™ probe interface.
Key performance specifications
- DC to 20 MHz bandwidth
- 150 ARMS maximum current capability
- 500 a peak pulse current capability
- High accuracy with typically less than 1% DC gain error
- Accurately measures current levels as low as 5 mA
Key features
- Easy-to-Use and accurate AC/DC current measurements
- Connects directly to oscilloscopes with the TekVPI™ probe interface
- Low noise and DC drift
- Provides automatic units scaling and readout on the oscilloscope's display
- Remote GPIB/USB probe control through the oscilloscope
- Split-core probe head construction allows easy connection to conductors
- Easy to degauss and autozero
- Setup controls and probe status and diagnostic indicators are provided on both the probe hardware and through an Easy-to-Access scope UI display menu
Applications
- Power supplies
- Semiconductor devices
- Power inverters/converters
- Electronic ballasts
- Industrial/consumer electronics
- Motor drives
- Transportation systems
This AC/DC current measurement probe provides 20 MHz of frequency bandwidth with selectable range control for 25 A and 150 A measurement ranges. It also provides low-current measurement capability and accuracy to current levels as low as 5 mA per division — important for achieving broad dynamic current range of high-level and low-level measurement signals.
Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
Warranted characteristics
- Bandwidth
- DC to ≥20 MHz (-3 dB). Derate with frequency. See Figure 1. (Calculated from Rise Time BW = .35/Rise Time)
- Rise time
- ≤17.5 ns
- Max RMS current
- 150 A
- Max peak pulse current
- 500 A
- Minimum sensitivity
- 5 mA (on oscilloscope's 1 mV/div setting)
- AC coupling (on oscilloscopes that support AC/DC coupling)
- Current ranges
- 25 A and 150 A
- Maxbare-wire voltage
- 600 VRMS CAT I and II
- 300 VRMS CAT III
Typical characteristics
- DC accuracy
- ±1% typical (3% warranted)
- Max amp-second product
- 25 A range
- 3,000 Aμs
- 150 A range
- 15,000 Aμs
- Insertion impedance
- 0.03 Ω at 1 MHz
- 0.075 Ω at 5 MHz
- 0.125 at 10 MHz
- 0.3 at 20 MHz
- Signal delay
- ≈ 21 ns
- Displayed RMS noise
- ≤500 μARMS (probe only)
Physical characteristics
- Cable length
- 2 m (79 in.)
- Probe head size
- Length
- 26.8 cm (10.55 in.)
- Width
- 4.1 cm (1.6 in.)
- Height
- 15.6 cm (6.13 in.)
- Max conductor diameter
- 21 mm x 25 mm (0.83 in. x 1.0 in.)
- Shipping weight
- 0.7865 kg (1.734 lb.)
- Power requirements
- TCP0150 is powered directly by oscilloscopes with the TekVPI interface
Environmental
- Temperature
- Operating
- 0 ºC to +50 ºC
- Nonoperating
- -40 ºC to +75 ºC
- Humidity
- Operating
- 5% to 95% RH (Relative humidity) at up to +30 ºC
- 5% to 85% RH above 30 ºC up to +50 ºC, noncondensing
- Nonoperating
- 5% to 95% RH (Relative humidity) at up to +30 ºC
- 5% to 85% RH above 30 ºC up to +75 ºC, noncondensing
- Altitude
- Operating
- Up to 2,000 m (6,560 feet)
- Nonoperating
- Up to 12,000 m (40,000 feet)
Recommended oscilloscopes
Oscilloscopes with the TekVPI probe interface.
For best probe support, download and install the latest version of the oscilloscope software from http://tek.com.
Performance graphs
Ordering information
- TCP0150
- AC/DC current probe with TekVPI interface. Includes protective cover (016-1924-xx), certificate of traceable calibration, and data standard.
Language options
- Opt. L0
- English manual
- Opt. L5
- Japanese manual
- Opt. L7
- Simplified Chinese manual
- Opt. L9
- Korean manual
Service options
- Opt. CA1
- Single Calibration or Functional Verification
- Opt. C3
- Calibration Service 3 Years
- Opt. C5
- Calibration Service 5 Years
- Opt. D3
- Calibration Data Report 3 Years (with Opt. C3)
- Opt. D5
- Calibration Data Report 5 Years (with Opt. C5)
- Opt. R3
- Repair Service 3 Years (including warranty)
- Opt. R5
- Repair Service 5 Years (including warranty)
- Opt. SILV600
- Standard warranty extended to 5 years
Recommended accessories
- 015-0601-50
- Current loop, 1 turn, 50 Ω with BNC connector used for performance verification
- 067-1701-xx
- TekVPI calibration fixture
- 067-1686-xx
- Cal/Deskew fixture for DPO7000/70000 series