Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

Analog Oscilloscope

2205

The products on this datasheet are no longer being sold by Tektronix.
View Tektronix Encore for reconditioned test equipment.
Check support and warranty status for these products.

Features & Benefits

  • 40-MHz Bandwidth Option 40
  • Peak-to-Peak Auto Triggering
  • 0.3-Div Trigger Sensitivity
  • Front-Panel Z-Axis Input
  • 10-ns/div Max. Sweep Rate

Characteristics

Vertical System (2 Identical Channels)

Bandwidth (-3 dB) and Rise Time - 20 MHz and 17.5 ns (5 to 35 °C). 15 MHz and 23.3 ns (0 to 40 °C).

2205 Option 40 Bandwidth (-3 dB) and Rise Time - 40 MHz and 8.75 ns (+5 to +35 °C). 35 MHz and 10 ns (0 to 40 °C).

Deflection Factor and Accuracy - 5 mV/div to 5 V/div ±3%, (15 to 35 °C); ±5%, (0 to 40 °C).

Vertical Operating Modes - CH 1, CH 2, CH 2 IN-VERT, ADD, ALT, CHOP.

CMRR - At least 10:1 at 10 MHz.

Input R and C - 1 Megohm, 25 pF.

Max Input Voltage - 400 V (DC + peak AC), 800 Vpk-pk.

Channel Isolation - 100:1 at 20 MHz.

Horizontal System

Sweep Speeds - 0.5 s/div to 0.1 µs/div (X10 MAG to 10 ns/div)

Accuracy - ±3%; Magnified ±4% (degrade by 1% outside 15 to 35 °C)

Horizontal Operating Modes - X1, X10, X-Y

Trigger System

Trigger Sensitivity -

Internal: 0.30 div at 5 MHz, 1.0 div at 30 MHz

External: 40 mV at 5 MHz, 150 mV at 30 MHz (150 mV at 45 MHz for option 40)

Trigger Operating Modes - Peak-Peak AUTO, NORM, TV FIELD, TV LINE, SGL SWP

Trigger Source - VERT MODE, CH 1, CH 2, LINE, EXT, EXT/10

Trigger Coupling - AC, DC

X-Y Operation

Deflection Factors - Same as vertical system

Bandwidth -

X-Axis: 2 MHz

Y-Axis: Same as vertical system

Phase Difference - ±3 degree from DC to 50 kHz

CRT System

Display - 8 cm × 10 cm, 1.8 kV nominal voltage (2205); 6 kV nominal voltage (Opt. 40).

Controls - INTENSITY, TRACE ROTATION, BEAM FIND, FOCUS.

Z-Axis - 5 V causes noticeable modulation. Usable to 5 MHz.

Power Requirements

Line Voltage Range -

Low: 95 V AC to 128 V AC

High: 185 V AC to 250 V AC

Line Frequency - 48 Hz to 440 Hz

Maximum Power Consumption - 40 W (60 VA)

Environmental Specifications

Instruments meet in part, the environmental requirements of MIL-T-28800D or C for Type III, Class 3, Style D or C equipment as described below.

Ambient Temperature -

Operating: 0 to +40 °C.

Nonoperating: -55 to +75 °C.

Humidity -

Percent: 95% five cycles, (120 hours) 90% +40 °C 4 hours.

Reference: MIL-T-28800D, paragraph 4.5.5.1.2.2.

Tektronix 062-2847-00, ll.

Altitude -

Operating: To 4,500 m (15,000 ft.), maximum operating temp decreases 1 °C per 300 m above 1500 m.

Nonoperating: To 15,000 m (50,000 ft.).

EMC - Meets Class B requirements per VDE 0871-B for radiated and conducted emissions and FCC requirements.

Vibration -

Operating: 15 minutes along each of the 3 major axes, 10 Hz to 55 Hz to 10 Hz in one-minute cycles. Hold for 10 minutes at 55 Hz. Displ. (in pk-pk): 0.015.

Shock -

Operating: 30 g, 1/2 sine, 11-ms duration, 3 shocks per axis along each major axis. Total of 18 shocks.

Other Characteristics

Safety - UL 1244 listed, Certified to CSA C22.2 No. 231-M89

Warranty - 1 year; 3 years optional

Physical Characteristics

Dimensions

mm

in.

Height

137

5.4

Width (with handle)

380

15.0

Depth (without front cover)

440

17.3

Weight

kg

lbs.

Net

6.7

14.8