Question :
Remote Operation: What Front Panel Settings are Required to use the VXIPnP drivers for SourceMeter?
Answer :
What Front Panel Settings are Required to use the VXI PnP drivers for the SourceMeter Instruments?
Selection and configuration of GPIB or RS-232 communication is important. Both are supported by the VXI PnP driver, however, the instrument must be configured from the front panel first.
Use the front panel buttons and select:
- Menu
- Communication
- GPIB or RS-232
select the address (default is 24) select SCPI protocol (not 488.1) BAUD: 9600 BITS: 8 Parity: NONE Terminator: LF Flow-CTRL: XON-XOFF
instrument at address 24 instrument on COM1 GPIB0::24::INSTR ASRL1::INSTR
FAQ ID 69611
View all FAQs »