Question :
MSO4000+ DPO4000+ MSO3000+ DPO3000: For Measurements/Statistics the Mean & Std Dev Samples tab+ what does the number which is entered for Samples represent?
Answer :
The number you enters for Samples represents the number of waveforms acquired. If you enter the number 5 the scope will use the first 5 waveforms acquired. When 6 waveforms are acquired the scope will now uses waveforms 2-6 and does not use waveform number 1 which was the first acquisition. When the scope reaches 7 acquisitions it no longer uses the first or second waveform acquisitions and so forth.
This FAQ Applies to:
Product Series: MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000 MSO3000 / DPO3000