Do you have a SCPI Sequence for 100V @ 10A on Model 2461 SMU?

Question :

Do you have a SCPI Sequence for 100V @ 10A on Model 2461 SMU?

Answer :

Here is a sequence of SCPI commands the configure the Model 2461 SMU to output a 100V pulse up to 10A.

 

  • RST

SYSTem:CLEar

TRACe:MAKE "PulseBuffer", 50

SOURce:FUNCtion CURRent

SOURce:CURRent:RANGe:AUTO OFF

SOURce:CURRent:RANGe 1

SOURce:CURRent 0

SOURce:CURRent:READ:BACK OFF

SOURce:CURRent:DELay:AUTO OFF

SOURce:CURRent:VLIMit 105

 

SENSe:FUNCtion "VOLTage"

SENSe:VOLTage:RANGe:AUTO OFF

SENSe:VOLTage:RANGe 100

SENSe:VOLTage:NPLCycles 0.01

SENSe:VOLTage:RSENse ON

ROUTe:TERMinals FRONt

SENSe:VOLTage:AZERo OFF

 

SOURce:PULSe:TRain:CURRent 0, 10, .001, 21, ON, "PulseBuffer", 0, .019, 2, 100, ON

          --bias level, pulse level, pulse width, count, measEnable, bufferName, delay, offtime, xBiasLimit, xPulseLimit, FailAbort

INITiate:IMMediate

TRACe:DATA? 1, 21, "PulseBuffer"