Key performance specifications

With the lowest input noise and up to 8 GHz analog bandwidth, the 6 Series MSO provides the best signal fidelity for analyzing and debugging today's embedded systems with GHz clock and bus speeds. The remarkably innovative pinch-swipe-zoom touchscreen user interface coupled with the industry's largest high definition display and 4 FlexChannel® inputs that let you measure one analog or eight digital signals per channel, the 6 Series MSO is ready for today's toughest challenges and tomorrow's too.

Input channels

  • 4 FlexChannel® inputs
  • Each FlexChannel provides:
    • One analog signal that can be displayed as a waveform view, a spectral view, or both simultaneously
    • Eight digital logic inputs with TLP058 logic probe

Bandwidth (all analog channels)

  • 1 GHz, 2.5 GHz, 4 GHz, 6 GHz, 8 GHz (upgradable)

Sample rate (all analog / digital channels)

  • Real-time: 25 GS/s
  • Interpolated: 2.5 TS/s

Record length (all analog / digital channels)

  • 62.5 Mpoints standard
  • 125, 250, 500 Mpoints, or 1 Gpoints (optional)

Waveform capture rate

  • >500,000 waveforms/s

Vertical resolution

  • 12-bit ADC
  • Up to 16-bits in High Res mode

Standard trigger types

  • Edge, Pulse Width, Runt, Timeout, Window, Logic, Setup & Hold, Rise/Fall Time, Parallel Bus, Sequence, Visual Trigger, Video (optional), RF vs. Time (optional)
  • Auxiliary Trigger ≤5 VRMS, 50Ω, 400 MHz (Edge Trigger only)

Standard analysis

  • Cursors: Waveform, V Bars, H Bars, V&H Bars
  • Measurements: 36
  • Spectrum View: Frequency-domain analysis with independent controls for frequency and time domains
  • FastFrame™: Segmented memory acquisition mode with maximum trigger rate >5,000,000 waveforms per second
  • Plots: Time Trend, Histogram, Spectrum and Phase Noise

  • Math: Basic waveform arithmetic, FFT, and advanced equation editor
  • Search: Search on any trigger criteria
  • Jitter: TIE and Phase Noise

Optional analysis

  • Advanced Jitter and Eye Diagram Analysis
  • User-defined filtering
  • Advanced Spectrum View
  • RF vs. Time traces, triggers, Spectrograms, and IQ capture
  • Digital Power Management
  • Mask/Limit Testing
  • LVDS Debug and Analysis
  • PAM3 Analysis
  • Advanced Power Measurements and Analysis
  • Advanced Vector Signal Analysis (SignalVu-PC)

Optional protocol trigger, decode, and analysis

I2C, SPI, eSPI, I3C, RS-232/422/485/UART, SPMI, SMBus, CAN, CAN FD, CAN XL, LIN, FlexRay, SENT, PSI5, CXPI, Automotive Ethernet,MIPI C-PHY,MIPI D-PHY, USB 2.0, eUSB2.0, USB3.0, PCIe Gen1/Gen2, Ethernet, EtherCAT, Audio, MIL-STD-1553, ARINC 429, Spacewire, 8B/10B,NRZ, Manchester, SVID, SDLC, 1-Wire, MDIO, and NFC

Optional serial compliance test

  • Ethernet, USB 2.0, Automotive Ethernet, Multi-gigabit Automotive Ethernet, Industrial Ethernet, MIPI D-PHY 1.2, MIPI D-PHY 2.1, MIPI C-PHY 2.0

Optional memory analysis

  • DDR3 debug, analysis, and compliance test

Arbitrary/Function Generator (optional and upgradable)

  • 50 MHz waveform generation
  • Waveform Types: Arbitrary, Sine, Square, Pulse, Ramp, Triangle, DC Level, Gaussian, Lorentz, Exponential Rise/Fall, Sin(x)/x, Random Noise, Haversine, Cardiac

Digital voltmeter (free with product registration)

  • 4-digit AC RMS, DC, and DC+AC RMS voltage measurements

Trigger frequency counter (free with product registration)

  • 8-digit

Display

  • 15.6 inch (396 mm) TFT color
  • High Definition (1,920 x 1,080) resolution
  • Capacitive (multi-touch) touchscreen

Connectivity

  • USB Host (7 ports), USB 3.0 Device (1 port), LAN (10/100/1000 Base-T Ethernet), DisplayPort, DVI-I, VGA

e*Scope®

  • Remotely view and control the oscilloscope over a network connection through a standard web browser

Warranty

  • 3 years standard

Dimensions

  • 12.2 in (309 mm) H x 17.9 in (454 mm) W x 8.0 in (204 mm) D
  • Weight: < 28.4 lbs. (12.88 kg)

FlexChannel® technology enables maximum flexibility and broader system visibility

The 6 Series MSO redefines what a Mixed Signal Oscilloscope (MSO) should be. FlexChannel technology enables each channel input to be used as a single analog channel, eight digital logic inputs (with the TLP058 logic probe), or simultaneous analog and spectrum views with independent acquisition controls for each domain. Imagine the flexibility and configurability this provides.

You can change the configuration at any time by simply adding or removing TLP058 logic probes, so you always have the right number of digital channels.

MSO6 EN_US_48W 61353 25
FlexChannel technology enables the ultimate in flexibility. Each input can be configured as a single analog or eight digital channels based on the type of probe you attach.

Previous-generation MSOs required tradeoffs, with digital channels having lower sample rates or shorter record lengths than analog channels. The 6 Series MSO offers a new level of integration of digital channels. Digital channels share the same high sample rate (up to 25 GS/s), and long record length (up to 250M points) as analog channels.

MSO6 EN_US_48W 61353 25
The TLP058 provides eight high performance digital inputs. Connect as many TLP058 probes as you like, enabling up to a maximum of 32 digital channels.
MSO6 EN_US_48W 61353 25
Channel 2 has a TLP058 Logic Probe connected to the eight inputs of a DAC. Notice the green and blue color coding, where ones are green and zeros are blue. Another TLP058 Logic Probe on Channel 3 is probing the SPI bus driving the DAC. The white edges indicate higher frequency information is available by either zooming in or moving to a faster sweep speed on the next acquisition.
MSO6 EN_US_48W 61353 25
Beyond just analog and digital, FlexChannel inputs include Spectrum View. This Tektronix-patented technology enables you to simultaneously view both analog and spectral views of all your analog signals, with independent controls in each domain. For the first time ever, oscilloscope-based frequency-domain analysis is as easy as using a spectrum analyzer while retaining the ability to correlate frequency-domain activity with other time-domain phenomena.

Unprecedented signal viewing capability

The stunning 15.6 inch (396 mm) display is the largest display in the industry. It is also the highest resolution display, with full HD resolution (1,920 x 1,080), enabling you to see many signals at once with ample room for critical readouts and analysis.

The viewing area is optimized to ensure that the maximum vertical space is available for waveforms. The Results Bar on the right can be hidden, enabling the waveform view to use the full width of the display.

MSO6 EN_US_48W 61353 25
Stacked display mode enables easy visibility of all waveforms while maintaining maximum ADC resolution on each input for the most accurate measurements.

The 6 Series MSO offers a revolutionary new Stacked display mode. Historically, scopes have overlaid all waveforms in the same graticule, forcing difficult tradeoffs:

  • To make each waveform visible, you vertically scale and position each waveform so that they don't overlap. Each waveform uses a small percentage of the available ADC range, leading to less accurate measurements.

  • For measurement accuracy, you vertically scale and position each waveform to cover the entire display. The waveforms overlap each other, making it hard to distinguish signal details on individual waveforms

The new Stacked display eliminates this tradeoff. It automatically adds and removes additional horizontal waveform 'slices' (additional graticules) as waveforms are created and removed. Each slice represents the full ADC range for the waveform. All waveforms are visually separated from each other while still using the full ADC range, enabling maximum visibility and accuracy. And it's all done automatically as waveforms are added or removed! Channels can easily be reordered in stacked display mode by dragging and dropping the channel and waveform badges in the Settings bar at the bottom of the display. Groups of channels can also be overlaid within a slice to simplify visual comparison of signals.

The massive display also provides plenty of viewing area not only for signals, but also for plots, measurement results tables, bus decode tables and more. You can easily resize and relocate the various views to suit your application.

MSO6 EN_US_48W 61353 25
Viewing three analog channels, eight digital channels, a decoded serial bus waveform, decoded serial packet results table, four measurements, a measurement histogram, measurements results table with statistics and a search on serial bus events - simultaneously!

Exceptionally easy-to-use user interface lets you focus on the task at hand

The Settings Bar - key parameters and waveform management

Waveform and scope operating parameters are displayed in a series of “badges” in the Settings Bar that runs along the bottom of the display. The Settings Bar provides Immediate access for the most common waveform management tasks. With a single tap, you can:

  • Turn on channels
  • Add math waveforms
  • Add reference waveforms
  • Add bus waveforms
  • Enable the optional integrated Arbitrary/Function generator (AFG)
  • Enable the optional integrated digital voltmeter (DVM)

The Results Bar - analysis and measurements

The Results Bar on the right side of the display includes immediate, one-tap access to the most common analytical tools such as cursors, measurements, searches, measurement and bus decode results tables, plots, and notes.

DVM, measurement and search results badges are displayed in the Results Bar without sacrificing any waveform viewing area. For additional waveform viewing area, the Results Bar can be dismissed and brought back at any time.

MSO6 EN_US_48W 61353 25
Configuration menus are accessed by simply double-tapping on the item of interest on the display. In this case, the Trigger badge was double-tapped to open the Trigger configuration menu.

Touch interaction finally done right

Oscilloscopes have included touch screens for years, but the touch interface has been an afterthought. The 6 Series MSO 15.6" display includes a capacitive touchscreen and provides the industry's first oscilloscope user interface truly designed for touch.

The touch interactions that you use with phones and tablets, and expect in a touch enabled device, are supported.

  • Drag waveforms left/right or up/down to adjust horizontal and vertical position or to pan a zoomed view
  • Pinch and expand to change scale or zoom in/out in either horizontal or vertical directions
  • Drag items off the edge of the screen to delete them
  • Swipe in from the right to reveal the Results Bar or down from the top to access the menus in the upper left corner of the display

Smooth, responsive front panel controls allow you to make adjustments with familiar knobs and buttons, and you can add a mouse or keyboard as a third interaction method.

MSO6 EN_US_48W 61353 25
Interact with the capacitive touch display in the same way you do on your phones and tablets.

Variable font size

Historically, oscilloscope user interfaces have been designed with fixed font sizes to optimize viewing of waveforms and readouts. This implementation is fine if all users have the same viewing preferences, but they don't. Users spend a significant amount of time staring at screens, and Tektronix recognizes this. The 6 Series MSO offers a user preference for variable font sizes; scaling down to 12 points or up to 20 points. As you adjust the font size, the user interface dynamically scales so you can easily choose the best size for your application.

MSO6 EN_US_48W 61353 25
Comparison showing how the user interface scales as font size changes.
MSO6 EN_US_48W 61353 25
Efficient and intuitive front panel provides critical controls while still leaving room for the massive 15.6" high definition display.

Attention to detail in the front-panel controls

Traditionally, the front face of a scope has been roughly 50% display and 50% controls. The 6 Series MSO display fills about 85% of the face of the instrument. To achieve this, it has a streamlined front panel that retains critical controls for simple intuitive operation, but with a reduced number of menu buttons for functions directly accessed via objects on the display.

Color-coded LED light rings indicate trigger source and vertical scale/position knob assignments. Large, dedicated Run/Stop and Single Sequence buttons are placed prominently in the upper right, and other functions like Force Trigger, Trigger Slope, Trigger Mode, Default Setup, Autoset and Quick-save functions are all available using dedicated front panel buttons.

Windows or not - you choose

The 6 Series MSOoffers you the choice of whether to include a Microsoft Windows™ operating system. Opening an access panel on the bottom of the instrument reveals a connection for a solid state drive (SSD). When the SSD is not present, the instrument boots as a dedicated scope with no ability to run or install other programs.

MSO6 EN_US_48W 61353 25

When the SSD is present, the instrument boots in an open Windows 10 configuration, so you can minimize the oscilloscope application and access a Windows desktop where you can install and run additional applications on the oscilloscope or you can connect additional monitors and extend your desktop.

Whether you run Windows or not, the oscilloscope operates in exactly the same way with the same look and feel and UI interaction.

Need higher channel density?

The 6 Series is also available as a low-profile digitizer - the LPD64. With four SMA input channels plus an auxiliary trigger input, in a 2U high package and 12-bit ADC's, the 6 Series Low Profile Digitizer sets a new standard for performance in applications where extreme channel density is required.

MSO6 EN_US_48W 61353 25

Experience the performance difference

With up to 8 GHz analog bandwidth, 25 GS/s sample rates, standard 62.5 Mpts record length and a 12-bit analog to digital converter (ADC), the 6 Series MSO has the performance you need to capture waveforms with the best possible signal fidelity and resolution for seeing small waveform details.

Digital Phosphor technology with FastAcq™ high-speed waveform capture

To debug a design problem, first you must know it exists. Digital phosphor technology with FastAcq provides you with fast insight into the real operation of your device. Its fast waveform capture rate - greater than 500,000 waveforms per second - gives you a high probability of seeing the infrequent problems common in digital systems: runt pulses, glitches, timing issues, and more. To further enhance the visibility of rarely occurring events, intensity grading indicates how often rare transients are occurring relative to normal signal characteristics.

MSO6 EN_US_48W 61353 25
FastAcq's high waveform capture rate enables you to discover infrequent problems common in digital design.

Industry leading vertical resolution and low noise

The 6 Series MSO provides the performance to capture the signals of interest while minimizing the effects of unwanted noise when you need to capture high-amplitude signals while seeing smaller signal details. At the heart of the instrument are 12-bit analog-to-digital converters (ADCs) that provide 16 times the vertical resolution of traditional 8-bit ADCs.

A new High Res mode applies a hardware-based unique Finite Impulse Response (FIR) filter based on the selected sample rate. The FIR filter maintains the maximum bandwidth possible for that sample rate while preventing aliasing and removing noise from the oscilloscope amplifiers and ADC above the usable bandwidth for the selected sample rate.

High Res mode always provides at least 12 bits of vertical resolution and extends all the way to 16 bits of vertical resolution at ≤ 625 MS/s sample rates and 200 MHz of bandwidth. The following table shows the number of bits of vertical resolution for each sample rate setting when in High Res.

Sample rate Number of bits of vertical resolution
25 GS/s 8
12.5 GS/s 12
6.25 GS/s 13
3.125 GS/s 14
1.25 GS/s 15
≤625 MS/s 16

New lower-noise front end amplifiers further improve your ability to resolve fine signal detail.

MSO6 EN_US_48W 61353 25
12-bit ADC, along with the new High Res mode, enable industry leading vertical resolution.

A new TEK061 front end amplifier sets a new standard for low-noise acquisition providing the best signal fidelity to capture small signals with high resolution.

MSO6 EN_US_48W 61353 25

A key attribute to being able to view fine signal details on small, high-speed signals is noise. The higher a measurement systems' intrinsic noise, the less true signal detail will be visible. This becomes more critical on an oscilloscope when the vertical settings are set to high sensitivity (like ≤ 10 mV/div) in order to view small signals that are prevalent in high-speed bus topologies. The 6 Series MSO has a new front-end ASIC, the TEK061, that enables breakthrough noise performance at the highest sensitivity settings. The table below shows a comparison of typical noise performance of the 6 Series MSO and prior generations of Tektronix oscilloscopes in this bandwidth range.

50 Ω, RMS voltage, typical
Bandwidth V/Div 6 Series MSO DPO7000C MSO/DPO70000C
1 GHz 1 mV 54.8 µV 90 µV 1N/A
10 mV 90.9 µV 279 µV N/A
100 mV 941 µV 2.7 mV N/A
4 GHz 1 mV 97.4 µV N/A N/A
10 mV 192 µV N/A 500 µV
100 mV 1.92 mV N/A 4.3 mV
8 GHz 1 mV 158 µV N/A N/A
10 mV 342 µV N/A 580 µV
100 mV 3.46 mV N/A 4.5 mV

Triggering

Discovering a device fault is only the first step. Next, you must capture the event of interest to identify root cause. The 6 Series MSO provides a complete set of advanced triggers, including:

  • Runt
  • Logic
  • Pulse width
  • Window
  • Timeout
  • Rise/Fall time
  • Setup and Hold violation
  • Serial packet
  • Parallel data
  • Sequence
  • Video
  • Visual Trigger
  • RF Frequency vs. Time
  • RF Magnitude vs. Time

With up to a 1 Gpoint record length, you can capture many events of interest, even thousands of serial packets in a single acquisition, providing high-resolution to zoom in on fine signal details and record reliable measurements.

MSO6 EN_US_48W 61353 25
The wide variety of trigger types and context-sensitive help in the trigger menu make it easier than ever to isolate the event of interest.

Visual trigger - Finding the signal of interest quickly

Finding the right cycle of a complex bus can require hours of collecting and sorting through thousands of acquisitions for an event of interest. Defining a trigger that isolates the desired event speeds up debug and analysis efforts.

Visual Trigger extends the 6 Series MSO's triggering capabilities by scanning through all waveform acquisitions and comparing them to on-screen areas (geometric shapes). An unlimited number of areas can be created using a mouse or touchscreen, and a variety of shapes (triangles, rectangles, hexagons, or trapezoids) can be used to specify the desired trigger behavior. Once shapes are created, they can be edited interactively to create custom shapes and ideal trigger conditions.

MSO6 EN_US_48W 61353 25
Visual Trigger areas isolate an event of interest, saving time by only capturing the events you want to see.

By triggering only on the most important signal events, Visual Trigger can save hours of capturing and manually searching through acquisitions. In seconds or minutes, you can find the critical events and complete your debug and analysis efforts. Visual Trigger even works across multiple channels, extending its usefulness to complex system troubleshooting and debug tasks.

MSO6 EN_US_48W 61353 25
Multiple channel triggering. Visual Trigger areas can be associated with events spanning multiple channels such as packets transmitted on two bus signals simultaneously.

Once multiple areas are defined, a Boolean logic equation can be used to set complex trigger conditions using on-screen editing features.

MSO6 EN_US_48W 61353 25
Boolean logic trigger qualification. Boolean logic using logical OR allows triggering on a specific anomaly in the signal.

TekVPI Probe Interface

The TekVPI® probe interface sets the standard for ease of use in probing. In addition to the secure, reliable connection that the interface provides, many TekVPI probes feature status indicators and controls, as well as a probe menu button right on the comp box itself. This button brings up a probe menu on the oscilloscope display with all relevant settings and controls for the probe. The TekVPI interface enables direct attachment of current probes without requiring a separate power supply. TekVPI probes can be controlled remotely through USB or LAN, enabling more versatile solutions in ATE environments. The 6 Series MSO provides up to 40 W of power to the front panel connectors, sufficient to power all connected TekVPI probes without the need for an additional probe power supply.

Convenient high-speed passive voltage probing

The TPP Series passive voltage probes offer all the benefits of general-purpose probes - high dynamic range, flexible connection options, and robust mechanical design - while providing the performance of active probes. Up to 1 GHz analog bandwidth enables you to see high frequency components in your signals, and extremely low 3.9 pF capacitive loading minimizes adverse effects on your circuits and is more forgiving of longer ground leads.

An optional, low-attenuation (2X) version of the TPP probe is available for measuring low voltages. Unlike other low-attenuation passive probes, the TPP0502 has high bandwidth (500 MHz) as well as low capacitive loading (12.7 pF).

MSO6 EN_US_48W 61353 25
The instrument comes with standard one TPP1000 (1 GHz, 2.5 GHz models) probe per channel.

TDP7700 Series TriMode Probes

The TDP7700 Series TriMode probes provide the highest probe fidelity available for real-time oscilloscopes. The TDP7700 is designed for use with the 6 Series MSO, with full AC calibration of the probe and tip's signal path based on unique S-parameter models. The probe communicates the S-parameters to the scope via the TekVPI probe interface and the 6 Series MSO includes them to achieve the very best signal fidelity possible from probe tip to acquisition memory. Connectivity innovations such as solder-down tips with the probe's input buffer mounted only a few millimeters from the end of the tip, the TDP7700 Series probes provide unmatched usability for connecting to today's most challenging electronic designs.

MSO6 EN_US_48W 61353 25
TDP7700 Series probe with a selection of available tips

With TriMode probing one probe setup makes differential, single ended, and common mode measurements accurately. This unique capability allows you to work more effectively and efficiently, switching between differential, single ended and common mode measurements without moving the probe's connection point.

IsoVu™ Isolated Measurement System

Whether designing an inverter, optimizing a power supply, testing communication links, measuring across a current shunt resistor, debugging EMI or ESD issues, or trying to eliminate ground loops in your test setup, common mode interference has caused engineers to design, debug, evaluate, and optimize "blind" until now.

The revolutionary Tektronix IsoVu technology uses optical communications and power-over-fiber for complete galvanic isolation. When combined with the 6 Series MSO equipped with the TekVPI interface, it is the first, and only, measurement system capable of accurately resolving high bandwidth, differential signals, in the presence of large common mode voltage with:

  • Complete galvanic isolation
  • Up to 1 GHz bandwidth
  • 1 Million to 1 (120 dB) common mode rejection at 100 MHz
  • 10,000 to 1 (80 dB) of common mode rejection at full bandwidth
  • Up to 2,500 V differential dynamic range
  • 60 kV common mode voltage range
MSO6 EN_US_48W 61353 25
The Tektronix TIVP Series IsoVu™ Measurement System offers a galvanically isolated measurement solution to accurately resolve high bandwidth, differential signals up to 2,500 Vpk in the presence of large common mode voltages, with the best-in-class common mode rejection performance across its bandwidth.

High-side gate voltage measurement with IsoVu

The following image shows a comparison of the high-side gate voltage for a standard differential probe versus an optically isolated probe. For both at turn-off and turn-on, high-frequency ringing can be seen on the gate after the device’s gate passes through the threshold region. Due to coupling between the gate and power loop, some ringing is expected. However, in the case of the differential probe, the ringing has a significantly higher amplitude than is measured by the optically isolated probe. This is likely due to the changing reference voltage inducing common mode currents within the probe and an artifact of a standard differential probe. While the waveform measured by the differential probe appears to pass the maximum gate voltage of the device, the more accurate measurement of the optically isolated probe makes it clear that the device is within specification. Application designers using standard differential probes for gate voltage measurements should use caution as it may not be possible to differentiate between the probing and measurement system artifact shown here and an actual violation of the device ratings. This measurement artifact may cause the designer to increase the gate resistance to slow down the switching transient and reduce the ringing. However, this would unnecessarily increase losses in the SiC device. For this reason, it is essential to have a measurement system that accurately reflects the actual dynamics of the device, in order to appropriately design the system and optimize performance.

MSO6 EN_US_48W 61353 25
Differential Probe (blue trace) vs. IsoVu Optically Isolated Probe (yellow trace)

Comprehensive analysis for fast insight

Basic waveform analysis

Verifying that your prototype's performance matches simulations and meets the project's design goals requires careful analysis, ranging from simple checks of rise times and pulse widths to sophisticated power loss analysis, characterization of system clocks, and investigation of noise sources.

The 6 Series MSO offers a comprehensive set of standard analysis tools including:

  • Waveform- and screen-based cursors
  • 36 automated measurements. Measurement results include all instances in the record, the ability to navigate from one occurrence to the next, and immediate viewing of the minimum or maximum result found in the record
  • Basic waveform math
  • Basic FFT analysis
  • Advanced waveform math including arbitrary equation editing with filters and variables
  • FastFrame™ Segmented Memory enables you to make efficient use of the oscilloscope’s acquisition memory by capturing many trigger events in a single record while eliminating the large time gaps between events of interest. View and measure the segments individually or as an overlay.

Standard amplitude and time measurements annotate the waveform display with visual bars and markers to indicate relative information. Measurement results tables provide comprehensive statistical views of measurement results with statistics across both the current acquisition and all acquisitions.

MSO6 EN_US_48W 61353 25
Using measurements to characterize burst width and Frequency.

Callouts

  1. Note: Write and position a text box on the screen.
  2. Arrow: Write and position a text box, then add an arrow to a specific location on the screen.
  3. Rectangle: Write text and outline a specific region on the screen indicated by a resizable box.
  4. Bookmark: Create a dynamic readout at a specific time relevant to a trigger point. This readout includes text, magnitude of the signal, signal units, as well as a line and target indicating the bookmark reference point.

Documenting test results and methods is critical when sharing data across a team, recreating a measurement at a later date, or delivering a customer report. With a few taps on the screen, you can create as many custom callouts as needed; enabling you to document the specific details of your test results. With each callout, you can customize the text, location, color, font size, and font.

MSO6 EN_US_48W 61353 25
Easy to use callouts (Note, Arrow, Rectangle, Bookmark) that are detailing the specifics of this test setup and corresponding results.

Navigation and search

Finding your event of interest in a long waveform record can be time consuming without the right search tools. With today's record lengths of many millions of data points, locating your event can mean scrolling through literally thousands of screens of signal activity.

The 6 Series MSO offers the industry's most comprehensive search and waveform navigation with its innovative Wave Inspector® controls. These controls speed panning and zooming through your record. With a unique force-feedback system, you can move from one end of your record to the other in just seconds. Or, use intuitive drag and pinch/expand gestures on the display itself to investigate areas of interest in a long record.

The Search feature allows you to automatically search through your long acquisition looking for user-defined events. All occurrences of the event are highlighted with search marks and are easily navigated to, using the Previous ( ← ) and Next ( → ) buttons found on the front panel or on the Search badge on the display. Search types include edge, pulse width, timeout, runt, window, logic, setup and hold, rise/fall time and parallel/serial bus packet content. You can define as many unique searches as you like.

You can also quickly jump to the minimum and maximum value of search results by using the Min and Max buttons on the Search badge.

MSO6 EN_US_48W 61353 25
Earlier, FastAcq revealed the presence of a runt pulse in a digital data stream prompting further investigation.

Mask and limit testing (optional)

Whether you are focused on signal integrity or setting up pass/fail conditions for production, mask testing is an efficient tool to characterize the behavior of certain signals in a system. Quickly create custom masks by drawing mask segments on the screen. Tailor a test to your specific requirements and set actions to take when a mask hit is registered, or when a complete test passes or fails.

Limit testing is an insightful way to monitor the long-term behavior of signals, helping you characterize a new design or confirm hardware performance during production line testing. Limit tests compare your live signal to an ideal, or golden version of the same signal with user-defined vertical and horizontal tolerances.

You can easily tailor a mask or limit test to your specific requirements by:

  • Defining test duration in number of waveforms
  • Setting a violation threshold that must be met before considering a test a failure
  • Counting violations/failures and reporting statistical information
  • Setting actions upon violations, test failure, and test complete

MSO6 EN_US_48W 61353 25
Custom, multiple segment mask capturing the presence of a signal glitch and runt pulse in a waveform.

User-defined filtering (optional)

In the broad sense, any system that processes a signal can be thought of as a filter. For example, an oscilloscope channel operates as a low pass filter where its 3 dB down point is referred to as its bandwidth. Given a waveform of any shape, a filter can be designed that can transform it into a defined shape within the context of some basic rules, assumptions, and limitations.

Digital filters have some significant advantages over analog filters. For example, the tolerance values of analog filter circuit components are high enough that high order filters are difficult or even impossible to implement. High order filters are easily implemented as digital filters. Digital filters can be implemented as Infinite Impulse Response (IIR) or Finite Impulse Response (FIR). The choice of IIR or FIR filters are based upon design requirements and application.

The 6 Series MSO has the ability to apply designated filters to math waveforms through a MATH arbitrary function. Option 6-UDFLT takes this functionality a level deeper, providing more than MATH arbitrary basic functions and adds flexibility to support standard filters and can be used for application centric filter designs.

MSO6 EN_US_48W 61353 25
Filters can be created through the Math dialog. Once a filter is edited, it can be easily applied, saved, and recalled for use or modification later.

Filter types supported on the 6 Series MSO include:

  • Low pass
  • High pass
  • Band pass
  • Band stop
  • All pass
  • Hilbert
  • Differentiator
  • Custom

Filter response types supported on the 6 Series MSO include:

  • Butterworth
  • Chebyshev I
  • Chebyshev II
  • Elliptical
  • Gaussian
  • Bessel-Thomson
The Filter Response control is available for all Filter Types except All-pass, Hilbert, or Differentiator.

MSO6 EN_US_48W 61353 25
Filter creation dialog showing selection for Filter Type, Filter Response, Cutoff Frequency, Filter Order, and a graphical representation of Magnitude/Phase, Impulse Response, and Step Response

Filter designs can be saved, recalled, and applied once any editing has been completed.

Protocol decode and analysis (optional)

During debugging, it can be invaluable to trace the flow of activity through a system by observing the traffic on one or more serial buses. It could take many minutes to manually decode a single serial packet, much less the thousands of packets that may be present in a long acquisition.

And if you know the event of interest that you are attempting to capture occurs when a particular command is sent across a serial bus, wouldn't it be nice if you could trigger on that event? Unfortunately, it's not as easy as simply specifying an edge or a pulse width trigger.

MSO6 EN_US_48W 61353 25
Triggering on a USB full-speed serial bus. A bus waveform provides time-correlated decoded packet content including Start, Sync, PID, Address, End Point, CRC, Data values, and Stop, while the bus decode table presents all packet content from the entire acquisition.

The 6 Series MSO offers a robust set of tools for working with the most common serial buses found in embedded design including I2C, SPI, eSPI, I3C, RS-232/422/485/UART, SPMI, SMBus, CAN, CAN FD, CAN XL, LIN, FlexRay, SENT, PSI5, CXPI, Automotive Ethernet, MIPI C-PHY,MIPI D-PHY, USB 1.0 (1.5 Mbps), USB 1.1 (12 Mbps), USB 2.0 (480 Mbps), eUSB2.0, USB3.0 (5 Gbps), Ethernet 10/100, EtherCAT, Audio (I2S/LJ/RJ/TDM), MIL-STD-1553, ARINC 429, Spacewire, 8B/10B, NRZ, Manchester, SVID, SDLC, 1-Wire, MDIO, and NFC.

Protocol search enables you to search through a long acquisition of serial packets and find the ones that contain the specific packet content you specify. Each occurrence is highlighted by a search mark. Rapid navigation between marks is as simple as pressing the Previous ( ← ) and Next ( → ) buttons on the front panel or in the Search badge that appears in the Results Bar.

The tools described for serial buses also work on parallel buses. Support for parallel buses is standard in the instrument. Parallel buses can be up to 32 bits wide and can include a combination of analog and digital channels.

  • Serial protocol triggering lets you trigger on specific packet content including start of packet, specific addresses, specific data content, unique identifiers, and errors.
  • Bus waveforms provide a higher-level, combined view of the individual signals (clock, data, chip enable, and so on) that make up your bus, making it easy to identify where packets begin and end, and identifying sub-packet components such as address, data, identifier, CRC, and so on.
  • The bus waveform is time aligned with all other displayed signals, making it easy to measure timing relationships across various parts of the system under test.
  • Bus decode tables provide a tabular view of all decoded packets in an acquisition much like you would see in a software listing. Packets are time stamped and listed consecutively with columns for each component (Address, Data, and so on).

NFC decode and analysis (optional)

Evaluating the performance margins of NFC designs is often difficult due to an inability to trace the protocol-level result down to the parametric signal level. This means marginal passes may result in failures later in the test flow, especially when designs are susceptible to interference and signal integrity issues caused by design trade-offs or nearby electronics, requiring time consuming debug across multiple instruments like a protocol analyzer and RF signal analyzer.

The NFC Protocol Decode and Search option on the 6 Series MSO offers users the ability to view the transaction of the NFC link and trace the result through every step of signal manipulation in the standard, from the protocol-level down to the fundamental signal level to gain insight into exactly how your NFC chip, tag, reader, or mobile device is performing.

NFC transactions can be long. The software option uniquely uses the data coming from the hardware DDC used for Spectrum View, which allows for sample rate compression, saving transfer time and memory, allowing for 100s of milliseconds or even seconds of signal data to be captured and analyzed.

Additionally, because I/O signals are not always available to probe and trigger on from the device under test, triggering on the RF envelope itself is also a challenge considering NFC’s small modulation index. With Spectrum View, you can trigger on the 13.56 MHz envelope using RF vs. Time traces and triggers, which is also unique amongst instruments.

This capability simplifies up-front design validation and also provides a powerful debugging tool in a single instrument when failures do occur.

MSO6 EN_US_48W 61353 25
NFC software option allows you to decode and search through their digital NFC bit stream for performing NFC analog/RF and digital pre-conformance, debug, and troubleshooting in a single instrument.

Spectrum View

It is often easier to debug an issue by viewing one or more signals in the frequency domain. Oscilloscopes have included math-based FFTs for decades in an attempt to address this need. However, FFTs are notoriously difficult to use for two primary reasons.

First, when performing frequency-domain analysis, you think about controls like Center Frequency, Span, and Resolution Bandwidth (RBW), as you would typically find on a spectrum analyzer. But then you use an FFT, where you are stuck with traditional scope controls like sample rate, record length and time/div and have to perform all the mental translations to try to get the view you’re looking for in the frequency-domain.

Second, FFTs are driven by the same acquisition system that’s delivering the analog time-domain view. When you optimize acquisition settings for the analog view, your frequency-domain view isn’t what you want. When you get the frequency-domain view you want, your analog view is not what you want. With math-based FFTs, it is virtually impossible to get optimized views in both domains.

Spectrum View changes all of this. Tektronix’ patented technology provides both a decimator for the time-domain and a digital downconverter (DDC) for the frequency-domain behind each FlexChannel. The two different acquisition paths let you simultaneously observe both time- and frequency-domain views of the input signal with independent acquisition settings for each domain. Other manufacturers offer various ‘spectral analysis’ packages that claim ease-of-use, but they all exhibit the limitations described above. Only Spectrum View provides both exceptional ease-of-use and the ability to achieve optimal views in both domains simultaneously.

Traditionally, performing RF measurements, such as RF Channel Power (CHP), Adjacent Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW), required a dedicated spectrum or signal analyzer or spectrum analyzer software. This additional hardware or software leads to more complexity and higher costs. Available standard with Spectrum View, integrated RF Measurements on each channel saves users time, bench space, and costs with the ability to validate RF transmitter CHP, ACPR, and OBW directly on the oscilloscope.

Additionally, the DDC significantly reduces the required sample rate to resolve a signal compared to a conventional FFT since it becomes a function of span rather than center frequency. This allows for reduced file sizes, improved frequency resolution, and faster spectrum update rates, leading to a more responsive and accurate solution capable of capturing 10's of seconds of spectrum data.

MSO6 EN_US_48W 61353 25
Intuitive spectrum analyzer controls like center frequency, span and resolution bandwidth (RBW), independent from time domain controls, provide easy setup for frequency domain analysis. A spectrum view is available for each FlexChannel analog input, enabling multi-channel mixed domain analysis.
MSO6 EN_US_48W 61353 25
Spectrum Time gates the range of time where the FFT is being calculated. Represented by a small graphical rectangle in the time domain view, it can be positioned to provide time correlation with the time domain waveform. Perfect for conducting Mixed Domain Analysis. Up to 11 automated peak markers provide frequency and magnitude values of each peak. The Reference marker is always the highest peak shown and is indicated in red.

Visualizing changes in the RF signal (optional)

RF time domain traces make it easy to understand what’s happening with a time-varying RF signal. There are three RF time domain traces that are derived from the underlying I and Q data of Spectrum View:

  • Magnitude – The instantaneous amplitude of the spectrum vs. time.
  • Frequency – The instantaneous frequency of the spectrum relative to the center frequency vs. time.
  • Phase – The instantaneous phase of the spectrum relative to the center frequency vs. time.

Each of these traces can be turned on and off independently, and all three can be displayed simultaneously.

The data is stored as in-phase and quadrature (I&Q) samples and precise synchronization is maintained between the time domain data and the I&Q data.

When RF vs. Time traces are activated, IQ data can be captured and exported to file for more advanced analysis within 3rd party applications.

With frequency on the x-axis, time on the y-axis, and power level indicated by variations in color, the Spectrogram display (included with option RFVT) offers enhanced insight into changes in signal amplitude and frequency content over time, allowing you to see where and when changes in spectral activity occur. This makes it ideal for displaying trends in spectral data such as when diagnosing complex spurious, frequency hopping, multi-channel, and dynamic signals.

Spectrogram benefits include:

  • Ability to view all spectrum activity in a given span and acquisition immediately, without having to specify FFT overlap or Spectrum Time
  • Quickly compare spectrum at different moments in time using time-correlated cursors and up to three overlaid spectrum traces
  • Pinch and zoom in on spectral activity of interest with display resolution and FFT overlap automatically optimized
  • Adjust center frequency, span, RBW, and amplitude color-scaling as needed to view all signals of interest
  • Simultaneously view trends in multi-channel or non-contiguous spectrum by activating spectrograms on each available oscilloscope channel and independently setting center frequency and amplitude scaling
MSO6 EN_US_48W 61353 25
MSO6 EN_US_48W 61353 25
The lower trace is the frequency vs. time trace derived from the input signal. Notice that the Spectrum Time is positioned during a transition from the lowest frequency to the middle frequency, so the energy is spread across a number of frequencies. With the frequency vs. time trace, you can easily see the different frequency hops, simplifying characterization of how the device switches between frequencies.

Triggering on changes in the RF signal (optional)

Whether you need to find the source of electromagnetic interference or understand the behavior of a VCO, hardware triggers for RF versus time make it easy to isolate, capture, and understand the RF signal behavior. Trigger on edges, pulse widths, and timeout behavior of RF magnitude vs. time and RF frequency vs. time.

Comprehensive vector signal analysis with SignalVu-PC (optional)

The Tektronix 6 Series MSO, combined with available analysis software, offers cost-effective mid-range performance as a 4 channel, 8 GHz bandwidth multichannel, multi-domain Vector Signal Analysis (VSA) solution

When analysis needs go beyond the basic spectrum, amplitude, frequency, and phase vs. time you can employ the SignalVu-PC vector signal analysis application. This enables in-depth transient RF signal analysis, detailed RF pulse characterization, and comprehensive analog and digital RF modulation analysis.

Tektronix’ mixed signal oscilloscope based approach to 5G test, with dedicated DDC’s on each channel and 5G New Radio (NR) SignalVu-PC VSA software, offers a novel approach to validate 5G NR designs that the traditional RF engineer may not have considered previously due to technical limitations offered by traditional FFT-based oscilloscopes, and offers benefits for analyzing both time, frequency, and modulation domains simultaneously across multiple channels.

  • The separate digital signal path for time and frequency domains and phase-matching between channels are critical for beamformer calibration.
  • You can also analyze digital and analog/RF data simultaneously, to validate latency, modulation accuracy, and perform power efficiency or system-level debugging.

5G NR transmitter measurements core supported features

The 5GNR option (5GNRNL-SVPC) supports 5G NR modulation analysis measurements according to Release 15 and Release 16 of 3GPP’s TS 38 specifications, including:

  • Analysis of uplink and downlink frame structures
  • For the downlink, supported test models for FDD and TDD
  • For the uplink, supported test models for FDD
  • Modulation Accuracy (including Error Vector Magnitude (EVM) and IQ error)
  • Channel Power (CHP)
  • Adjacent Channel Power (ACP)
  • Spectrum Emission Mask (SEM)
  • Occupied Bandwidth
  • Power vs Time (PVT)
  • Summary table with all scalar results for Modulation Accuracy, ACP, CHP, SEM, and OBW measurements
  • In-depth analysis and troubleshooting with coupled measurements across domains, use multiple markers to correlate results to find the root cause.
  • Automate measurements using SCPI commands and save/recall configuration parameters and measurement results in .TIQ or .CSV format
  • Configurable parameters of PDSCH or PUSCH for each component carrier

MSO6 EN_US_48W 61353 25
5GNR measurements from SignalVu-PC provide insight into 5GNR designs

To enable the SignalVu-PC application on your 6 Series MSO Oscilloscope, three options are required.

  1. To run the application on the instrument, the Windows SSD (6-WIN) needs to be installed in the oscilloscope.
  2. The Spectrum View RF versus time traces option (6-SV-RFVT) needs to be installed in the oscilloscope to enable I/Q data transfer.
  3. The Connect (CONxx-SVPC) license needs to be installed on the SignalVu-PC to enable base features of application, which includes 16+ RF measurements and displays.

The RF digital down converters and integrated measurement engines behind each channel have your complex mixed-signal and mixed-domain analysis needs covered in one instrument.

MSO6 EN_US_48W 61353 25
MSO68 showing SignalVu-PC multi-channel VSA software running on the instrument

Advanced pulse analysis (optional)

The per-channel DDC available on the 6 Series MSO offers the ability to analyze RF signals independently on all channels, including configuration of separate timing, triggering and measurements. This capability extends to evaluation of time, frequency and modulation domains simultaneously when using the SignalVu-PC VSA software.

The Advanced Pulse Analysis Option (SVPNL-SVPC) allows you to analyze multiple radar signals across measurement channels on a common timebase with independent or coordinated controls and measurements.

MSO6 EN_US_48W 61353 25
The advanced pulse analysis application in SignalVu-PC offers up to 31 automated pulse parameter measurements, statistics, and displays for analysis on up to 8 signal sources simultaneously, streamlining workflows and boosting efficiency.

Jitter analysis

The 6 Series MSO has seamlessly integrated the DPOJET Essentials jitter and eye pattern analysis software package, extending the oscilloscope's capabilities to take measurements over contiguous clock and data cycles in a single-shot real-time acquisition. This enables measurement of key jitter and timing analysis parameters such as Time Interval Error and Phase Noise to help characterize possible system timing issues.

Analysis tools, such as plots for time trends and histograms, quickly show how timing parameters change over time, and spectrum analysis quickly shows the precise frequency and amplitude of jitter and modulation sources.

Option 6-DJA adds additional jitter analysis capability to better characterize your device's performance. The 31 additional measurements provide comprehensive jitter and eye-diagram analysis and jitter decomposition algorithms, enabling the discovery of signal integrity issues and their related sources in today's high-speed serial, digital, and communication system designs. Option 6-DJA also provides eye diagram mask testing for automated pass/fail testing.

MSO6 EN_US_48W 61353 25
The unique Jitter Summary provides a comprehensive view of your device's performance in a matter of seconds.

Power analysis (optional)

The 6 Series MSO has also integrated the optional power analysis package into the oscilloscope automatic measurement system to enable quick and repeatable analysis of power quality, input capacitance, in-rush current, harmonics, switching loss, safe operating area (SOA), modulation, ripple, Magnetics measurements, efficiency, amplitude and timing measurements, slew rate (dv/dt and di/dt), Control Loop Response (Bode Plot), and Power Supply Rejection Ratio (PSRR).

Measurement automation optimizes the measurement quality and repeatability at the touch of a button, without the need for an external PC or complex software setup.

MSO6 EN_US_48W 61353 25
The Power Analysis measurements display a variety of waveforms and plots.

Digital power management (optional)

The Digital Power Management and analysis (DPM) software option provides automated power rail measurements for Power Integrity Analysis on the 6 Series MSO oscilloscopes. The solution enables both simultaneous analysis of multiple power rails (using power rail probes) and sequencing of measurements (using passive probes). The solution is designed with the user work flow in mind to help design engineers meet their time-to-market needs. It also generates an automated report that includes measurements, test results, and plot images.

Key measurements include ripple, ripple-on-ripple, power sequencing, jitter analysis, transient analysis, power integrity and signal integrity analysis.

The Power Supply Induced Jitter (PSIJ) measurement acts as a tool that gives insights and confidence to signal integrity engineers to model the effects of hardware changes, to test their effectiveness before actually making them. The measurement provides essential results such as eye height, eye width, PJ, and TIE before and after filtering.

MSO6 EN_US_48W 61353 25

Inverter Motor Drive Analysis (IMDA)(optional)

MSO6 EN_US_48W 61353 25
On the left is a Phasor diagram displaying the phase and magnitude of current and voltage measurements for all three phases of power. In the results badge on the right, are the results from the automated measurements of power quality.

During the design and validation of systems that utilize 3-Phase power, it can be difficult to correlate control systems and power electronics with the performance of the overall system.

This will give you deeper insights enabling you to debug the design, efficiency and reliability of:

  • 3-Phase power inverters, converters, power supplies, and automotive 3-Phase designs for DC-AC topology
  • Motors (brushless AC, brushless DC, induction, permanent magnet, universal, stepper, rotor)
  • Drives (AC, DC, variable frequency, servo)

The automated measurements that are included with 6-IMDA are:

  • Input analysis
    • Power quality with phasor diagram
    • Harmonics
    • Input voltage
    • Input current
    • Input power
  • Ripple analysis
    • Line Ripple
    • Switching Ripple
  • Output analysis
    • Phasor diagram
    • Efficiency
    • Mechanical Power
    • System Efficiency
  • Wiring configurations
    • 1 Volt/1 Current - 1P2W
    • 2 Volt/2 Current - 1P3W
    • 2 Volt/2 Current - 3P3W
    • 3 Volt/3 Current - 3P3W
    • 3 Volt/3 Current - 3P4W

Compliance test

A key focus area for embedded designers is testing various embedded and interface technologies for compliance. This ensures the device passes the logo certification at plugfests and achieves successful interoperability when working with other compliant devices.

The compliance test specifications for high speed serial standards like USB, Ethernet, Memory, Display and MIPI are developed by the respective consortiums, or governing bodies. Working closely with these consortiums, Tektronix has developed oscilloscope-based compliance applications that not only focus on providing pass/fail results but also provide deeper insight into any failures by providing relevant measurement tools such as jitter and timing analysis to debug failing designs.

These automated compliance applications are built on a framework that provides:

  • Complete test coverage per the specification.

  • Fast test times with optimized acquisitions and test sequencing based on customized settings.

  • Analysis based on previously-acquired signals, allowing the device under test (DUT) to be disconnected from the setup once all acquisitions are completed. This also allows analysis of waveforms acquired on a different oscilloscope or captured at a remote lab, facilitating a very collaborative test environment.

  • Signal validation during acquisition to ensure the right signals are being captured.

  • Additional parametric measurements for design debug.

  • Custom eye diagram mask testing for insight into design margin.

  • Detailed reports in multiple formats with setup information, results, margins, waveform screenshots and plot images.

MSO6 EN_US_48W 61353 25
TekExpress USB2 (Option 6-CMUSB2) DUT panel configures the DUT-specific settings
MSO6 EN_US_48W 61353 25
6 Series MSO running 6-CMUSB2 Compliance Measurements as per USB 2.0 Specification

Wide Bandgap Double Pulse Test (optional)

The Wide Bandgap Double Pulse Test application offers precise Wide Bandgap measurements that make device and the system validation easier. It has an ability to test SiC or GaN devices and also Si MOSFET and IGBTs. The application is compatible with all the Tektronix VPI probes and when used with the Tektronix IsoVu™ probes, it helps uncover all the hidden artifacts of SiC or GaN devices at the circuit level. The application offers automated measurements as per the JEDEC and IEC standards. It offers unique features such as per-cycle analysis with annotation, flexibility with custom reference level settings, configurable integration points, and power preset that can be set based on the DUT designs.

Following measurements are performed:

  • Low side switching parameters and High side diode reverse recovery measurements
  • Low side and High side switching parameters

MSO6 EN_US_48W 61353 25
The image shows diode reverse recovery measurements with reverse recovery current and voltage captured on the high-side.

Designed with your needs in mind

Connectivity

The 6 Series MSO contains a number of ports which you can use to connect the instrument to a network, directly to a PC, or to other test equipment.

  • Two USB 2.0 and one USB 3.0 host ports on the front and four more USB host ports (two 2.0, two 3.0) on the rear panel enable easy transfer of screen shots, instrument settings, and waveform data to a USB mass storage device. A USB mouse and keyboard can also be attached to USB host ports for instrument control and data entry.
  • The rear panel USB Device port is useful for controlling the oscilloscope remotely from a PC.
  • The standard 10/100/1000BASE-T Ethernet port on the rear of the instrument enables easy connection to networks and provides LXI Core 2011 compatibility.
  • The DVI-D, DisplayPort and VGA ports on the rear of the instrument let you duplicate the instrument display on an external monitor or projector.

MSO6 EN_US_48W 61353 25
The I/O you need to connect the 6 Series MSO to the rest of your design environment.

Upgrade Automated Test Equipment (ATE) systems quickly and smoothly

Anyone working closely with automated test systems knows that moving to a new model or platform can be painful. Modifying an existing codebase for a new product can be prohibitively expensive and complicated. Now there's a solution.

All 6 Series MSO’s include a Programmatic Interface (PI) Translator. When enabled, the PI Translator acts as an intermediate layer between your test application and the oscilloscope. It recognizes a subset of legacy commands from the popular DPO/MSO5000B and DPO7000C platforms and translates them on the fly into supported commands for the 6 Series MSO. The Translator interface is designed to be human-readable and easily extensible, which means that you can customize its behavior to minimize the amount of effort required when transitioning to your new oscilloscope.

Remote operation to improve collaboration

Want to collaborate with a design team on the other side of the world?

The included e*Scope® capability enables fast control of an oscilloscope running the Embedded Operating System over a network connection. This can be viewed from any PC or device through a standard web browser.

Simply enter the IP address or network name of the oscilloscope and a web page will be served to the browser. Control the oscilloscope remotely in the exact same way that you do in-person using the built-in touchscreen. Alternatively for oscilloscopes with the Microsoft Windows 10 Operating System, you can use Windows Remote Desktop™ to connect directly to the instrument and control it remotely.

The industry-standard TekVISA™ protocol interface is included for using and enhancing Windows applications for data analysis and documentation. IVI-COM instrument drivers are included to enable easy communication with the oscilloscope using LAN or USBTMC connections from an external PC.

MSO6 EN_US_48W 61353 25
e*Scope provides simple remote viewing and control using common web browsers.

PC-based analysis and remote connection to your oscilloscope

Get the analysis capability of an award-winning oscilloscope on your PC. Analyze waveforms anywhere, anytime. The basic license lets you view and analyze waveforms, perform many types of measurements and decode the most common serial buses - all while remotely accessing your oscilloscope. Advanced license options add capabilities such as multi-scope analysis, more serial bus decoding options, and power measurements.

MSO6 EN_US_48W 61353 25
TekScope PC analysis software runs on a Windows computer with the same award-winning user experience as the 4, 5, and 6 Series MSO's

Key features of the TekScope PC analysis software include:

  • Connect to the instrument remotely and enable low-latency waveform updates using high-speed TekHSI data transfers.
  • Share the data remotely with your colleagues so that they can perform analysis and make measurements as if they were sitting in front of the oscilloscope
  • Synchronize waveforms from the multiple oscilloscopes in real-time
  • Perform advanced analysis even if your oscilloscope is not equipped with TekScope PC analysis software
  • Recall Tektronix oscilloscope sessions and waveform files from the equipment made by Tektronix and other vendors.
  • Waveform file formats supported include .wfm, .isf, .csv, .h5, .tr0, .trc, and .bin

TekDrive collaborative test and measurement workspace

Using TekDrive, you can upload, store, organize, search, download, and share any file type from any connected device. TekDrive is natively integrated into the instrument for seamless sharing and recalling of files - no USB stick is required. Analyze and explore standard files like .wfm, .isf, .tss, and .csv, directly in a browser with smooth interactive waveform viewers. TekDrive is purpose built for integration, automation, and security.

MSO6 EN_US_48W 61353 25
TekDrive collaborative workspace - save files directly from your instrument and share across your team

Arbitrary/Function Generator (AFG)

The instrument contains an optional integrated arbitrary/function generator, perfect for simulating sensor signals within a design or adding noise to signals to perform margin testing. The integrated function generator provides output of predefined waveforms up to MHz for sine, square, pulse, ramp/triangle, DC, noise, sin(x)/x (Sinc), Gaussian, Lorentz, exponential rise/fall, Haversine and cardiac. The AFG can load waveform records up to 128 k points in size from an internal file location or a USB mass storage device.

The AFG feature is compatible with Tektronix' ArbExpress PC-based waveform creation and editing software, making creation of complex waveforms fast and easy.

Digital Voltmeter (DVM) and Trigger Frequency Counter

The instrument contains an integrated 4-digit digital voltmeter (DVM) and 8-digit trigger frequency counter. Any of the analog inputs can be a source for the voltmeter, using the same probes that are already attached for general oscilloscope usage. The trigger frequency counter provides a very precise readout of the frequency of the trigger event on which you’re triggering.

Both the DVM and trigger frequency counter are available for free and are activated when you register your product.

Enhanced security

The 6 Series MSO provides you with the option to protect company data through the Security menu. This includes the option to restrict access to the instrument by password-protecting remote network access, I/O ports, and firmware updates to ensure the security of the data. By default, the oscilloscope disables remote access on initial use and gives you the option to enable remote access with or without a password.

To clear user data, run TekSecure from the menu. Sanitize the oscilloscope by removing the SSD from the bottom of the instrument.

MSO6 EN_US_48W 61353 25

Help when you need it

Several helpful resources are included so you can get your questions answered rapidly without having to find a manual or go to a website:

  • Graphical images and explanatory text are used in numerous menus to provide quick feature overviews.
  • All menus include a question mark icon in the upper right that takes you directly to the portion of the integrated help system that applies to that menu.
  • A short user interface tutorial is included in the Help menu for new users to come up to speed on the instrument in a matter of a few minutes.

MSO6 EN_US_48W 61353 25
Integrated help answers your questions rapidly without having to find a manual or go to the internet.

Specifications

All specifications are guaranteed and apply to all models unless noted otherwise.

Model overview

Oscilloscope
MSO64
FlexChannel inputs 4
Maximum analog channels 4
Maximum digital channels (with optional logic probes) 32
Bandwidth (calculated rise time) 1 GHz (400 ps), 2.5 GHz (160 ps), 4 GHz (100 ps), 6 GHz (66.67 ps), 8 GHz (50 ps)
DC Gain Accuracy

50 Ω: ±2.0% 2 at >2 mV/div (±2.0% at 2 mV/div typical, ±4% at 1`mV/div typical)

50 Ω: ±1.0% 3 of full scale at >2 mV/div, (±1.0% of full scale at 2 mV/div typical, ±2% at 1 mV/div typical)

1 MΩ: ±2.0%2 at >2 mV/div (±2% at 2 mV/div, ±2.5% at 1 mV/div typical and 500`μV/div typical)

1 MΩ: ±1.0%3 of full scale at >2 mV/div, (±1.0% of full scale at 2 mV/div typical, ±1.25% at 1 mV/div and 500`μV/div, typical)

ADC Resolution 12 bits
Vertical Resolution

8 bits @ 25 GS/s; 8 GHz on all channels

12 bits @ 12.5 GS/s; 4 GHz on all channels

13 bits @ 6.25 GS/s (High Res); 2 GHz on all channels

14 bits @ 3.125 GS/s (High Res); 1 GHz on all channels

15 bits @ 1.25 GS/s (High Res); 500 MHz on all channels

16 bits @ ≤625 MS/s (High Res); 200 MHz on all channels

Sample Rate 25 GS/s on all analog / digital channels (40 ps resolution)
Record Length 62.5 Mpoints on all analog / digital channels, 125 Mpoints on all analog / digital channels optional, 250 Mpoints on all analog / digital channels optional, 500 Mpoints on all analog / digital channels optional, and 1 Gpoints on all analog / digital channels optional
Waveform Capture Rate

>500,000 wfms/s (Peak Detect, Envelope acquisition mode),

>30,000 wfms/s (all other acquisition modes)

Arbitrary/Function Generator (opt.) 13 predefined waveform types with up to 50 MHz output
DVM 4-digit DVM (free with product registration)
Trigger Frequency Counter 8-digit frequency counter (free with product registration)

Vertical system- analog channels

Input coupling
DC, AC
Input impedance 1 MΩ DC coupled
1 MΩ ±1%
Input capacitance 1 MΩ DC coupled, typical
14.5 pF ±1.5 pF
Input impedance 50 Ω, DC coupled
50 Ω ±3%
Input sensitivity range
1 MΩ
500 µV/div to 10 V/div in a 1-2-5 sequence
Note: 500 μV/div is a 2X digital zoom of 1 mV/div.
50 Ω
1 mV/div to 1 V/div in a 1-2-5 sequence
1 mV/div is a 2X digital zoom of 2 mV/div.
Maximum input voltage
2.3 VRMS, at < 100 mV/div, with peaks ≤±20 V (Pulse Width ≤ 1 us)
5.5 VRMS, at ≥ 100 mV/div, with peaks ≤±20 V (Pulse Width ≤ 200 us)
1 MΩ: 300 VRMS
Derate at 20 dB/decade between 4.5 MHz and 45 MHz; derate 14 dB/decade between 45 MHz and 450 MHz. Above 450 MHz, 5.5 VRMS
Effective bits (ENOB), typical
2 mV/div, High Res mode, 50 Ω, 10 MHz input with 90% full screen
Bandwidth ENOB
4 GHz 5.9
3 GHz 6.1
2.5 GHz 6.2
2 GHz 6.35
1 GHz 6.8
500 MHz 7.2
350 MHz 7.4
250 MHz 7.5
200 MHz 7.75
20 MHz 8.8
50 mV/div, High Res mode, 50 Ω, 10 MHz input with 90% full screen
Bandwidth ENOB
4 GHz 7.25
3 GHz 7.5
2.5 GHz 7.6
2 GHz 7.8
1 GHz 8.2
500 MHz 8.5
350 MHz 8.8
250 MHz 8.9
200 MHz 9
20 MHz 9.8
2 mV/div, Sample mode, 50 Ω, 10 MHz input with 90% full screen
Bandwidth ENOB
8 GHz 5.1
7 GHz 5.3
6 GHz 5.5
5 GHz 5.65
4 GHz 5.9
3 GHz 6.05
2.5 GHz 6.2
2 GHz 6.35
1 GHz 6.8
500 MHz 7.2
350 MHz 7.3
250 MHz 7.5
200 MHz 7.3
20 MHz 7.6
50 mV/div, Sample mode, 50 Ω, 10 MHz input with 90% full screen
Bandwidth ENOB
8 GHz 6.5
7 GHz 6.6
6 GHz 6.8
5 GHz 7
4 GHz 7.2
3 GHz 7.4
2.5 GHz 7.6
2 GHz 7.7
1 GHz 8.2
500 MHz 8.4
350 MHz 8.7
250 MHz 8.8
200 MHz 7.8
20 MHz 7.9
DC gain accuracy
50 Ohm

±2.0%4 (±2.0% at 2 mV/div, ±4% at 1 mV/div, typical)

±1.0%5 of full scale, (±1.0% of full scale at 2 mV/div, ± 2% at 1 mV/div, typical)

Position range
±5 divisions
Offset ranges, maximum

Input signal cannot exceed maximum input voltage for the 50 Ω input path.

Volts/div Setting Maximum offset range, 50 Ω Input
1 mV/div - 99 mV/div ±1 V
100 mV/div - 1 V/div ±10 V
Volts/div Setting Maximum offset range, 1 MΩ Input
500 µV/div - 63 mV/div ±1 V
64 mV/div - 999 mV/div

±10 V

1 V/div - 10 V/div ±100 V
Offset accuracy

±(0.005 X | offset - position | + DC balance ); Offset, position, and DC Balance in units of Volts

Bandwidth selections
8 GHz model, 50 Ohm
20 MHz, 200 MHz, 250 MHz, 350 MHz, 500 MHz, 1 GHz, 2 GHz, 2.5 GHz, 3 GHz, 4 GHz, 5 GHz, 6 GHz, 7 GHz, and 8 GHz
6 GHz model, 50 Ohm
20 MHz, 200 MHz, 250 MHz, 350 MHz, 500 MHz, 1 GHz, 2 GHz, 2.5 GHz, 3 GHz, 4 GHz, 5 GHz, and 6 GHz
4 GHz model, 50 Ohm
20 MHz, 200 MHz, 250 MHz, 350 MHz, 500 MHz, 1 GHz, 2 GHz, 2.5 GHz, 3 GHz, and 4 GHz
2.5 GHz model, 50 Ohm
20 MHz, 200 MHz, 250 MHz, 350 MHz, 500 MHz, 1 GHz, 2 GHz, and 2.5 GHz
1 GHz model, 50 Ohm
20 MHz, 200 MHz, 250 MHz, 350 MHz, 500 MHz, and 1 GHz
1M Ohm
20 MHz, 200 MHz, 250 MHz, 350 MHz, and Full (500 MHz)
Bandwidth filtering optimized for
Flatness or Step response
Random noise, RMS, typical
50 Ω, typical
50 GS/s, Sample Mode, RMS
V/div1 mV/div 2 mV/div 5 mV/div 10 mV/div 20 mV/div 50 mV/div 100 mV/div 1 V/div
10 GHz183 μV 188 μV 228 μV 346 μV 602 μV 1.39 mV 3.58 mV 27.4 mV
9 GHz167 μV 172 μV 208 μV 315 μV 549 μV 1.27 mV 3.22 mV 25 mV
8 GHz153 μV 156 μV 192 μV 287 μV 501 μV 1.15 mV 2.94 mV 23.1 mV
7 GHz139 μV 141 μV 175 μV 262 μV 457 μV 1.07 mV 2.68 mV 21.1 mV
6 GHz124 μV 127 μV 156 μV 234 μV 412 μV 949 μV 2.39 mV 19 mV
25 GS/s, HiRes Mode, RMS
V/div1 mV/div2 mV/div5 mV/div10 mV/div20 mV/div50 mV/div 100 mV/div1 V/div
5 GHz111 μV 112 μV 134 μV 197 μV 338 μV 772 μV 1.99 mV 15.4 mV
4 GHz97.4 μV 98.7 μV 117 μV 171 μV 291 μV 672 μV 1.73 mV 13.3 mV
3 GHz83.8 μV 85 μV 101 μV 144 μV 245 μV 559 μV 1.46 mV 11.2 mV
2.5 GHz75.6 μV 76.6 μV 90.7 μV 128 μV 219 μV 498 μV 1.3 mV 9.85 mV
2 GHz68.9 μV 69.9 μV 81.7 μV 116 μV 195 μV 444 μV 1.17 mV 8.78 mV
1 GHz51.1 μV 51.8 μV 59.9 μV 82.9 μV 138 μV 314 μV 829 μV 6.22 mV
500 MHz37.5 μV 38 μV 43.4 μV 60 μV 99.9 μV 230 μV 607 μV 4.61 mV
350 MHz31.9 μV 32.3 μV 36.9 μV 49.9 μV 82.1 μV 185 μV 499 μV 3.62 mV
250 MHz28.1 μV 28.5 μV 32.5 μV 44 μV 71.5 μV 161 μV 440 μV 3.19 mV
200 MHz24.2 μV 24.5 μV 28 μV 37.9 μV 62.3 μV 140 μV 383 μV 2.78 mV
20 MHz8.68 μV 8.8 μV 10.1 μV 13.8 μV 22.9 μV 52.8 μV 136 μV 1.04 mV
1 MΩ, High Res mode (RMS), typical
V/div1 mV/div 2 mV/div 5 mV/div 10 mV/div 20 mV/div 50 mV/div 100 mV/div 1 V/div
500 MHz186 μV 202 μV 210 μV 236 μV 288 μV 522 μV 1.25 mV 13.4 mV
350 MHz134 μV 138 μV 145 μV 163 μV 216 μV 391 μV 974 μV 10.6 mV
250 MHz108 μV 110 μV 114 μV 131 μV 182 μV 374 μV 838 μV 9.63 mV
200 MHz106 μV 108 μV 109 μV 117 μV 149 μV 274 μV 674 μV 8.01 mV
20 MHz73 μV 73.2 μV 78.1 μV 99.6 μV 158 μV 361 μV 801 μV 8.29 mV
Crosstalk (channel isolation), typical

≥70 dB up to 2 GHz

≥60 dB up to 5 GHz

≥45 dB up to 8 GHz

for any two channels set to 200 mV/div.

Vertical system - digital channels

Number of channels
8 digital inputs (D7-D0) per installed TLP058 (traded off for one analog channel)
Vertical resolution
1 bit
Maximum input toggle rate
500 MHz
Minimum detectable pulse width, typical

300 ps

Thresholds
One threshold per digital channel
Threshold range
±40 V
Threshold resolution
10 mV
Threshold accuracy

± [100 mV + 3% of threshold setting after calibration]

Input hysteresis, typical
100 mV at the probe tip
Input dynamic range, typical
30 Vpp for Fin ≤ 200 MHz, 10 Vpp for Fin > 200 MHz
Absolute maximum input voltage, typical
±42 V peak
Minimum voltage swing, typical
400 mV peak-to-peak
Input impedance, typical
100 kΩ
Probe loading, typical
2 pF

Front end and RF system (all measurements are typical)

Sensitivity/Noise density
-157 dBm/Hz (1 mV/div, -38 dBm, 1.0001 GHz CF, 500 kHz span, 3 kHz RBW)
DANL

-163 dBm/Hz 10 MHz to 6 GHz, 1 mV/div

-160 dBm/Hz >6 GHz to 10 GHz, 1 mV/div

Noise figure
17 dB (1 mV/div, -38 dBm, 1.001 GHz, 500 kHz span, 3 kHz RBW)
SNR/Dynamic range
112 dB (1 GHz input carrier, 0 dBm scope input range, 1 GHz CF, 100 MHz span, 1 kHz RBW, measured ±20 MHz from center)
Absolute amplitude accuracy
±1 dB (0 - 8 GHz) for max 10 GHz BW
Phase noise @ 1GHz

10 MHz offset: -140 dBc/Hz

1 MHz offset: -132 dBc/Hz

100 kHz offset: -118 dBc/Hz

10 kHz offset: -118 dBc/Hz

EVM (256 QAM)

0.5% @ 20 MSymbols/s

1.1% @ 800 MSymbols/s

1.5% @ 1.2 GSymbols/s

1.6% @ 2 GSymbols/s

SFDR

60 dB @ 3 GHz, 5 GHz span

70 dB @ 2.35 GHz, 1.5 GHz span

Return Loss (<100 mV/div )

12 dB <5GHz

8 dB 5 GHz to 10 GHz

Harmonic distortion

2nd Harmonic: -58 dBC with a 0 dBm, 1 GHz signal

3rd Harmonic: -55 dBC with a 0 dBm, 1 GHz signal

Two-tone third order intercept point (at 99 mV/div)

25 dBm 10 MHz to 6 GHz

20 dBm 6 GHz to 8 GHz

12 dBm 8 GHz to 10 GHz

Horizontal system

Time base range
40 ps/div to 1,000 s/div
Sample rate range

6.25 S/s to 25 GS/s (real time)

50 GS/s to 2.5 TS/s (interpolated)

Record length range

Applies to analog and digital channels. All acquisition modes are 250 M maximum record length, down to 1 k minimum record length, adjustable in 1 sample increments.

Standard: 62.5 Mpoints

Option 6-RL-1: 125 Mpoints

Option 6-RL-2: 250 Mpoints

Seconds/Division range
Model 1 K 10 K 100 K 1 M 10 M 62.5 M 125 M 250 M 500 M 1 G
MSO64 Standard 62.5 M 40 ps - 16 s 400 ps - 160 s 4 ns - 1000 s 2.5 μs - 1000 s N/A N/A N/A N/A
MSO64 Option 6-RL-1 125 M 40 ps - 16 s 400 ps - 160 s 4 ns - 1000 s 2.5 μs - 1000 s 5 μs - 1000 s N/A N/A N/A
MSO64 Option 6-RL-2 250 M 40 ps - 16 s 400 ps - 160 s 4 ps - 1000 s 2.5 μs - 1000 s 5 μs - 1000 s 10 μs - 1000 s N/A N/A
Option 6-RL-3 500 Mpts 40 ps - 16 s 400 ps - 160 s 4 ps - 1000 s 2.5 us - 1000 s 5 us - 1000 s 10 us - 1000 s 20 us - 1000 s N/A
Option 6-RL-4: 1 Gpts 40 ps - 16 s 400 ps - 160 s 4 ps - 1000 s 2.5 us - 1000 s 5 us - 1000 s 10 us - 1000 s 20 us - 1000 s 40 us - 1000 s
Aperture uncertainty (sample jitter)
Time duration Typical jitter
<1 μs 80 fs
<1 ms 130 fs
Timebase accuracy
±1.0 x10-7 over any ≥1 ms time interval
Description Specification
Factory Tolerance ±12 ppb; at calibration, 25 °C ambient, over any ≥1 ms interval
Temperature stability ±20 ppb across the full operating range of 0 °C to 50 °C, after a sufficient soak time at the temperature; tested at operating temperatures
Crystal aging ±300 ppb; frequency tolerance change at 25 °C over a period of 1 year
Delta-time measurement accuracy, nominal

MSO6 EN_US_48W 61353 25

(assume edge shape that results from Gaussian filter response)

The formula to calculate delta-time measurement accuracy (DTA) for a given instrument setting and input signal assumes insignificant signal content above Nyquist frequency, where:

SR 1 = Slew Rate (1st Edge) around 1st point in measurement

SR 2 = Slew Rate (2nd Edge) around 2nd point in measurement

N = input-referred guaranteed noise limit (VRMS)

TBA = time base accuracy or reference frequency error

t p = delta-time measurement duration (sec)

Maximum duration at highest sample rate
2.5 ms (std.) or 5 ms (opt. 6-RL-1, 125 Mpoints) or 10 ms (opt. 6-RL-2, 250 Mpoints)
Time base delay time range
-10 divisions to 5,000 s
Deskew range
-125 ns to +125 ns with a resolution of 40 ps (for Peak Detect and Envelope acquisition modes).
-125 ns to +125 ns with a resolution of 1 ps (for all other acquisition modes).
Delay between analog channels, full bandwidth, typical

≤ 10 ps for any two channels with input impedance set to 50 Ω, DC coupling with equal Volts/div or above 10 mV/div

Delay between analog and digital FlexChannels, typical
< 1 ns when using a TLP058 and a passive probe matching the bandwidth of the scope, with no bandwidth limits applied
Delay between any two digital FlexChannels, typical
320 ps
Delay between any two bits of a digital FlexChannel, typical
200 ps

Trigger system

Trigger modes
Auto, Normal, and Single
Trigger coupling

DC, HF Reject (attenuates > 50 kHz), LF Reject (attenuates < 50 kHz), noise reject (reduces sensitivity)

Trigger holdoff range
0 ns to 10 seconds
Trigger bandwidth (edge, pulse and logic), typical
Model Trigger type Trigger bandwidth
MSO64 8 GHz Edge 8 GHz
MSO64 8 GHz Pulse, Logic 4 GHz
MSO64 6 GHz Edge 6 GHz
MSO64 6 GHz Pulse, Logic 4 GHz
MSO64 4 GHz, 2.5 GHz, 1 GHz: Edge, Pulse, Logic Product Bandwidth
Edge-type trigger sensitivity, DC coupled, typical
Path Range Specification
1 MΩ path (all models) 0.5 mV/div to 0.99 mV/div 5 mV from DC to instrument bandwidth
≥ 1 mV/div The greater of 5 mV or 0.7 div from DC to lesser of 500 MHz or instrument BW, & 6 mV or 0.8 div from > 500 MHz to instrument bandwidth
50 Ω path 1 mV/div to 9.98 mV/div 3.0 div from DC to instrument bandwidth
≥ 10 mV/div < 1.0 division from DC to instrument bandwidth
Line 90 V to 264 V line voltage at 50 - 60 Hz line frequency

103.5 V to 126.5 V

AUX Trigger in 250 mVPP, DC to 400`MHz
Edge-type trigger sensitivity, not DC coupled, typical
Trigger Coupling Typical Sensitivity
NOISE REJ 2.5 times the DC Coupled limits
HF REJ 1.0 times the DC Coupled limits from DC to 50 kHz. Attenuates signals above 50 kHz.
LF REJ 1.5 times the DC Coupled limits for frequencies above 50 kHz. Attenuates signals below 50 kHz.
Trigger jitter, typical
≤ 1.5 psRMS for sample mode and edge-type trigger
≤ 2 psRMS for edge-type trigger and FastAcq mode
≤ 40 psRMS for non edge-type trigger modes
Trigger jitter, AUX input, typical

≤ 40 psRMS for sample mode and edge-type trigger

AUX In trigger skew between instruments, typical

±100 ps jitter on each instrument with 1.5 ns skew; ≤1.7 ns total between instruments. With manual deskewing of individual channels, total instrument skew can reach 200 ps between different instrument channels.

Skew improves for pulse input voltages ≥1 Vpp

Trigger level ranges
This specification applies to logic and pulse thresholds.
Source Range
Any Channel ±5 divs from center of screen
Aux In Trigger ±5 V
Line Fixed at about 50% of line voltage
Trigger types
Edge:
Positive, negative, or either slope on any channel. Coupling includes DC, AC, noise reject, HF reject, and LF reject
Pulse Width:

Trigger on width of positive or negative pulses. Event can be time- or logic-qualified

Timeout:
Trigger on an event which remains high, low, or either, for a specified time period. Event can be logic-qualified
Runt:
Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again. Event can be time- or logic-qualified
Window:
Trigger on an event that enters, exits, stays inside or stays outside of a window defined by two user-adjustable thresholds. Event can be time- or logic-qualified
Logic:
Trigger when logic pattern goes true, goes false, or occurs coincident with a clock edge. Pattern (AND, OR, NAND, NOR) specified for all input channels defined as high, low, or don't care. Logic pattern going true can be time-qualified
Setup & Hold:
Trigger on violations of both setup time and hold time between clock and data present on any input channels
Rise / Fall Time:
Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either. Event can be logic-qualified
Video (option 6-VID):
Trigger on all lines, odd, even, or all fields of NTSC, PAL, and SECAM video signals
Sequence:
Trigger on B event X time or N events after A trigger with a reset on C event. In general, A and B trigger events can be set to any trigger type with a few exceptions: logic qualification is not supported, if A event or B event is set to Setup & Hold, then the other must be set to Edge, and Ethernet and High Speed USB (480 Mbps) are not supported
Visual trigger
Qualifies standard triggers by scanning all waveform acquisitions and comparing them to on-screen areas (geometric shapes). An unlimited number of areas can be defined with In, Out, or Don't Care as the qualifier for each area. A boolean expression can be defined using any combination of visual trigger areas to further qualify the events that get stored into acquisition memory. Shapes include rectangle, triangle, trapezoid, hexagon and user-defined.
Parallel Bus:
Trigger on a parallel bus data value. Parallel bus can be from 1 to 32 bits (from the digital and analog channels) in size. Supports Binary and Hex radices
I2C Bus (option 6-SREMBD):
Trigger on Start, Repeated Start, Stop, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 10 Mb/s
I3C Bus (option 6-SRI3C)
Trigger on Start, Repeated Start, Stop, Address, Data, I3C SDR Direct, I3C SDR Broadcast, Missing ACK, T-Bit Error, Broadcast Address Error, Hot-Join, HDR Restart, HDR Exit on I3C buses up to 10 Mb/s
SPI Bus (option 6-SREMBD):
Trigger on Slave Select, Idle Time, or Data (1-16 words) on SPI buses up to 20 Mb/s
RS-232/422/485/UART Bus (option 6-SRCOMP):
Trigger on Start Bit, End of Packet, Data, and Parity Error up to 15 Mb/s
CAN Bus (option 6-SRAUTO):
Trigger on Start of Frame, Type of Frame (Data, Remote, Error, or Overload), Identifier, Data, Identifier and Data, End Of Frame, Missing Ack, and Bit Stuff Error on CAN buses up to 1 Mb/s
CAN FD Bus (option 6-SRAUTO):
Trigger on Start of Frame, Type of Frame (Data, Remote, Error, or Overload), Identifier (Standard or Extended), Data (1-8 bytes), Identifier and Data, End Of Frame, Error (Missing Ack, Bit Stuffing Error, FD Form Error, Any Error) on CAN FD buses up to 16 Mb/s
LIN Bus (option 6-SRAUTO):
Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, and Error on LIN buses up to 1 Mb/s
FlexRay Bus (option 6-SRAUTO):
Trigger on Start of Frame, Indicator Bits (Normal, Payload, Null, Sync, Startup), Frame ID, Cycle Count, Header Fields (Indicator Bits, Identifier, Payload Length, Header CRC, and Cycle Count), Identifier, Data, Identifier and Data, End Of Frame, and Errors on FlexRay buses up to 10 Mb/s
SENT Bus (option 6-SRAUTOSEN)
Trigger on Start of Packet, Fast Channel Status and Data, Slow Channel Message ID and Data, and CRC Errors
SPMI Bus (option 6-SRPM):
Trigger on Sequence Start Condition, Reset, Sleep, Shutdown, Wakeup, Authenticate, Master Read, Master Write, Register Read, Register Write, Extended Register Read, Extended Register Write, Extended Register Read Long, Extended Register Write Long, Device Descriptor Block Master Read, Device Descriptor Block Slave Read, Register 0 Write, Transfer Bus Ownership, and Parity Error
USB 2.0 LS/FS/HS Bus (option 6-SRUSB2):
Trigger on Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake Packet, Special Packet, Error on USB buses up to 480 Mb/s
Ethernet Bus (option 6-SRENET):
Trigger on Start of Frame, MAC Addresses, MAC Q-tag, MAC Length/Type, MAC Data, IP Header, TCP Header, TCP/IPV4 Data, End of Packet, and FCS (CRC) Error on 10BASE-T and 100BASE-TX buses
Audio (I2S, LJ, RJ, TDM) Bus (option 6-SRAUDIO):
Trigger on Word Select, Frame Sync, or Data. Maximum data rate for I2S/LJ/RJ is 12.5 Mb/s. Maximum data rate for TDM is 25 Mb/s
MIL-STD-1553 Bus (option 6-SRAERO):
Trigger on Sync, Command (Transmit/Receive Bit, Parity, Subaddress / Mode, Word Count / Mode Count, RT Address), Status (Parity, Message Error, Instrumentation, Service Request, Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus Control Acceptance, Terminal Flag), Data, Time (RT/IMG), and Error (Parity Error, Sync Error, Manchester Error, Non-contiguous Data) on MIL-STD-1553 buses
ARINC 429 Bus (option 6-SRAERO):
Trigger on Word Start, Label, Data, Label and Data, Word End, and Error (Any Error, Parity Error, Word Error, Gap Error) on ARINC 429 buses up to 1 Mb/s
RF Magnitude vs. Time and RF Frequency vs. Time (option 6-SV-RFVT):
Trigger on edge, pulse width and timeout events

Acquisition system

Sample
Acquires sampled values
Peak Detect
Captures glitches as narrow as 160 ps at all sweep speeds
Averaging
From 2 to 10,240 waveforms
Maximum averaging speed = 180 waveforms/s
Fast Hardware Averaging
An acquisition mode for acquiring a large number of averages in a short amount of time. Fast hardware averaging optimizes the acquisition path, reducing storage truncation error and smoothing out fine scale non-linearity imperfections via an optional offset dithering technique. This feature is available through programmatic interface commands.
From 2 to 1,000,000 waveforms
Maximum averaging speed = 32,000 waveforms/s
Envelope
Min-max envelope reflecting Peak Detect data over multiple acquisitions
High Res

Applies a unique Finite Impulse Response (FIR) filter for each sample rate that maintains the maximum bandwidth possible for that sample rate while preventing aliasing and removing noise from the oscilloscope amplifiers and ADC above the usable bandwidth for the selected sample rate.

High Res mode always provides at least 12 bits of vertical resolution and extends all the way to 16 bits of vertical resolution at ≤ 625 MS/s sample rates.

FastAcq®

FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events.

Maximum waveform capture rate:

  • >500,000 wfms/s (Peak Detect or Envelope Acquisition mode)
  • >30,000 wfms/s (All other acquisition modes)

Roll mode
Scrolls sequential waveform points across the display in a right-to-left rolling motion, at timebase speeds of 40 ms/div and slower, when in Auto trigger mode.
History mode
Makes use of the maximum record length, allowing you to capture many triggered acquisitions, stop when you see something of interest, and quickly review all stored triggered acquisitions.

The number of available acquisitions stored in history is (Maximum record length) / (Current record length setting).

FastFrame™

Acquisition memory divided into segments.

Maximum trigger rate >5,000,000 waveforms per second

Minimum frame size = 50 points

Maximum Number of Frames: For frame size ≥ 1,000 points, maximum number of frames = record length / frame size.

For 50 point frames, maximum number of frames = 691,000

Waveform measurements

Cursor types
Waveform, V Bars, H Bars, V&H Bars, and Polar (XY/XYZ plots only)
DC voltage measurement accuracy, Average acquisition mode
Measurement Type DC Accuracy (In Volts)
Average of ≥ 16 waveforms ±((DC Gain Accuracy) * |reading - (offset - position)| + Offset Accuracy + 0.05 * V/div setting)
Delta volts between any two averages of ≥ 16 waveforms acquired with the same oscilloscope setup and ambient conditions ±(DC Gain Accuracy * |reading| + 0.1 div)
Automatic measurements
36, of which an unlimited number can be displayed as either individual measurement badges or collectively in a measurement results table
Amplitude measurements
Amplitude, Maximum, Minimum, Peak-to-Peak, Positive Overshoot, Negative Overshoot, Mean, RMS, AC RMS, Top, Base, and Area
Timing measurements
Period, Frequency, Unit Interval, Data Rate, Positive Pulse Width, Negative Pulse Width, Skew, Delay, Rise Time, Fall Time, Phase, Rising Slew Rate, Falling Slew Rate, Burst Width, Positive Duty Cycle, Negative Duty Cycle, Time Outside Level, Setup Time, Hold Time, Duration N-Periods, High Time, Low Time, Time to Minimum, and Time to Maximum
Jitter measurements (standard)
TIE and Phase Noise
Measurement statistics
Mean, Standard Deviation, Maximum, Minimum, and Population. Statistics are available on both the current acquisition and all acquisitions
Reference levels
User-definable reference levels for automatic measurements can be specified in either percent or units. Reference levels can be set to global for all measurements, per source channel or signal, or unique for each measurement
Gating
Screen, Cursors, Logic, Search, or Time. Specifies the region of an acquisition in which to take measurements. Gating can be set to Global (affects all measurements set to Global) or Local (all measurements can have a unique Time gate setting; only one Local gate is available for Screen, Cursors, Logic, and Search actions).
Measurement plots
Histogram, Time Trend, Spectrum, Eye Diagram (TIE measurement only), Phase Noise (Phase Noise measurement only)
Measurement limits
Pass/fail testing for user-definable limits on measurement values. Act on event for measurement value failures include Save Screen Capture, Save Waveform, System Request (SRQ), and Stop Acquisitions
Jitter analysis (option 6-DJA) adds the following:
Measurements

Jitter Summary, TJ@BER, RJ- δδ, DJ- δδ, PJ, RJ, DJ, DDJ, DCD, SRJ, J2, J9, NPJ, F/2, F/4, F/8, Eye Height, Eye Height@BER, Eye Width, Eye Width@BER, Eye High, Eye Low, Q-Factor, Bit High, Bit Low, Bit Amplitude, DC Common Mode, AC Common Mode (Pk-Pk), Differential Crossover, T/nT Ratio, SSC Freq Dev, SSC Modulation Rate

Measurement plots
Eye Diagram and Jitter Bathtub
Fast eye rendering: Shows the Unit Intervals (UIs) that define the boundaries of the eye along with a user specified number of surrounding UIs for added visual context
Complete eye rendering: Shows all valid Unit Intervals (UIs)
Measurement limits
Pass/fail testing for user-definable limits on measurement values. Act on event for measurement value failures include Save Screen Capture, Save Waveform, System Request (SRQ), and Stop Acquisitions
Eye diagram mask testing

Automated mask pass/fail testing with mask autofit

Power analysis (option 6-PWR) adds the following:
Measurements

Input Analysis (Frequency, VRMS, IRMS, voltage and current Crest Factors, True Power, Apparent Power, Reactive Power, Power Factor, Phase Angle, Harmonics, Inrush Current, Input Capacitance )

Amplitude Analysis (Cycle Amplitude, Cycle Top, Cycle Base, Cycle Maximum, Cycle Minimum, Cycle Peak-to-Peak)

Timing Analysis (Period, Frequency, Negative Duty Cycle, Positive Duty Cycle, Negative Pulse Width, Positive Pulse Width)

Switching Analysis (Switching Loss, dv/dt, di/dt, Safe Operating Area, RDSon)

Output Analysis (Line Ripple, Switching Ripple, Efficiency, Turn-on Time, Turn-off Time)

Magnetic Analysis (Inductance, I vs. Intg(V), Magnetic Loss, Magnetic Property)

Frequency Response Analysis (Control Loop Response Bode Plot, Power Supply Rejection Ratio, Impedance)

Measurement Plots
Harmonics Bar Graph, Switching Loss Trajectory Plot, and Safe Operating Area
Measurement limits
Pass/fail testing for user-definable limits on measurement values. Act on event for measurement value failures include Save Screen Capture, Save Waveform, System Request (SRQ), and Stop Acquisitions
Digital power management (option 6-DPM) adds the following:
Measurements
Ripple Analysis (Ripple)
Transient Analysis (Overshoot, Undershoot, Turn On Overshoot, DC Rail Voltage)
Power Sequence Analysis (Turn-on, Turn-off)
Jitter Analysis (TIE, PJ, RJ, DJ, Eye Height, Eye Width, Eye High, Eye Low)
PI/SI Analysis (PSIJ)
DDR3/LPDDR3 memory debug and analysis option (6-DBDDR3) adds the following:
Measurements

Amplitude Measurements (AOS, AUS, Vix(ac), AOS Per tCK, AUS Per tCK, AOS Per UI, AUS Per UI)

Time Measurements (tRPRE, tWPRE, tPST, Hold Diff, Setup Diff, tCH(avg), tCK(avg), tCL(avg), tCH(abs), tCL(abs), tJIT(duty), tJIT(per), tJIT(cc), tERR(n), tERR(m-n), tDQSCK, tCMD-CMD, tCKSRE, tCKSRX)

LVDS debug and analysis option (option 6-DBLVDS) adds the following:
Data Lane Measurements

Generic Test (Unit Interval, Rise Time, Fall Time, Data Width, Data Intra Skew (PN), Data Inter Skew (Lane-to-Lane), Data Peak-to-Peak)

Jitter Test (AC Timing, Clock Data Setup Time, Clock Data Hold Time, Eye Diagram (TIE), TJ@BER, DJ Delta, RJ Delta, DDJ, De-Emphasis Level)

Clock Lane Measurements

Generic Test (Frequency, Period, Duty Cycle, Rise Time, Fall Time, Clock Intra Skew (PN), Clock Peak-to-Peak)

Jitter Test (TIE, DJ, RJ)

SSC On (Mod Rate, Frequency Deviation Mean)

Waveform math

Number of math waveforms
Unlimited
Arithmetic
Add, subtract, multiply, and divide waveforms and scalars
Algebraic expressions
Define extensive algebraic expressions including waveforms, scalars, user-adjustable variables, and results of parametric measurements. Perform math on math using complex equations. For example (Integral (CH1 - Mean(CH1)) X 1.414 X VAR1)
Math functions
Invert, Integrate, Differentiate, Square Root, Exponential, Log 10, Log e, Abs, Ceiling, Floor, Min, Max, Degrees, Radians, Sin, Cos, Tan, ASin, ACos, and ATan
Relational
Boolean result of comparison >, <, ≥, ≤, =, and ≠
Logic
AND, OR, NAND, NOR, XOR, and EQV
Filtering function (standard)
Loading of user-definable filters. Users specify a file containing the coefficients of the filter.
Filtering function (option 6-UDFLT)
Filter types
Low pass, High pass, Band pass, Band stop, All pass, Hilbert, Differentiator, and Custom
Filter response types
Butterworth, Chebyshev I, Chebyshev II, Elliptical, Gaussian, and Bessel-Thomson
FFT functions
Spectral Magnitude and Phase, and Real and Imaginary Spectra
FFT vertical units

Magnitude: Linear and Log (dBm)

Phase: Degrees, Radians, and Group Delay

FFT window functions
Hanning, Rectangular, Hamming, Blackman-Harris, Flattop2, Gaussian, Kaiser-Bessel, and TekExp

Spectrum View

Center Frequency
Limited by instrument analog bandwidth
Span
74.5 Hz – 1.25 GHz (Standard)

74.5 Hz - 2 GHz

Coarse adjustment in a 1-2-5 sequence

RF Measurements
Channel Power (CHP), Adjacent Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW) measurements on Spectrum View trace data and display
RF vs. Time Traces
Magnitude vs. time, Frequency vs. time, Phase vs. time (with option 6-SV-RFVT)
RF vs. Time Trigger
Edge, pulse width, and timeout trigger on RF Magnitude vs. Time and RF Frequency vs. Time (with option 6-SV-RFVT)
Spectrograms
RF Frequency vs. Time vs. Amplitude display with frequency on x-axis, time on y-axis, and power level indicated by variations in color (with option 6-SV-RFVT)
Resolution Bandwidth (RBW)

93 μHz to 62.5 MHz

93 μHz to 100 MHz (with option 6-SV-BW-1)

IQ capture
The data is stored as in-phase and quadrature (I&Q) samples and precise synchronization is maintained between the time domain data and the I&Q data.
When RF vs. Time traces are activated (with option 6-SV-RFVT), IQ data can be captured and exported to file for more analysis within 3rd party applications.
The max acquisition time varies with span and sample rate. At 25 GS/s and 2 GHz span, the max acquisition time is 0.086 seconds. For 1 GHz span, the max acquisition time is 0.172 seconds. For 40 MHz span, the max acquisition time is 2.749 seconds. For 1 MHz span, the max acquisition time is 87.961 seconds.
Window types and factors
Window type Factor
Blackman-Harris 1.90
Flat-Top 2 3.77
Hamming 1.30
Hanning 1.44
Kaiser-Bessel 2.23
Rectangular 0.89
Spectrum Time
FFT Window Factor / RBW
Reference level
Reference level is automatically set by the analog channel Volts/div setting
Setting range: -42 dBm to +44 dBm
Vertical Position
-100 divs to +100 divs
Vertical units
dBm, dBµW, dBmV, dBµV, dBmA, dBµA
Vertical scaling
Linear, Log
Horizontal scaling
Linear, Log
Multi-channel spectrum analysis
Each FlexChannel input can be configured with Spectrum View, RF vs. Time traces (with option RFVT), and Spectrogram (with option RFVT).
Multiple RF measurements can be performed simultaneously across channels.
Spectrum Time and Center Frequency settings can be unlocked and moved independently from each other across channels. All Spectrum View channels must share the same Span, Resolution Bandwidth and Window Type.

Search

Number of searches
Unlimited
Search types
Search through long records to find all occurrences of user specified criteria including edges, pulse widths, timeouts, runt pulses, window violations, logic patterns, setup & hold violations, rise/fall times, and bus protocol events. Search results can be viewed in the Waveform View or in the Results table.

Save

Save
Save files directly to the oscilloscope or USB media, to a remote network drive, or to your TekDrive collaboration workspace.
Waveform type
Tektronix Waveform Data (.wfm), Comma Separated Values (.csv), MATLAB (.mat)
Waveform gating
Cursors, Screen, Resample (save every nth sample)
Screen capture type
Portable Network Graphic (*.png), 24-bit Bitmap (*.bmp), JPEG (*.jpg)
Setup type
Tektronix Setup (.set)
Report type
Adobe Portable Documents (.pdf), Single File web Pages (.mht)
Session type
Tektronix Session Setup (.tss)

Display

Display type
15.6 in. (395 mm) liquid-crystal TFT color display
Display resolution
1,920 horizontal × 1,080 vertical pixels (High Definition)
Display modes

Overlay: traditional oscilloscope display where traces overlay each other

Stacked: display mode where each waveform is placed in its own slice and can take advantage of the full ADC range while still being visually separated from other waveforms. Groups of channels can also be overlaid within a slice to simplify visual comparison of signals.

Zoom
Horizontal and vertical zooming is supported in all waveform and plot views.
Interpolation
Sin(x)/x and Linear
Waveform styles
Vectors, dots, variable persistence, and infinite persistence
Graticules
Movable and fixed graticules, selectable between Grid, Time, Full, and None
Color palettes

Normal and inverted for screen captures

Individual waveform colors are user-selectable

Fonts
Font size is user selectable from 12 to 20 (default is 15)
Format
YT, XY, and XYZ
Local Language User Interface
English, Japanese, Simplified Chinese, Traditional Chinese, French, German, Italian, Spanish, Portuguese, Russian, Korean
Local Language Help
English, Japanese, Simplified Chinese

Arbitrary-Function Generator (optional)

Modes of operation
Off, Continuous, Burst
Function types
Arbitrary, sine, square, pulse, ramp, triangle, DC level, Gaussian, Lorentz, exponential rise/fall, sin(x)/x, random noise, Haversine, Cardiac
Amplitude range
Values are peak-to-peak voltages
Waveform 50 Ω 1 MΩ
Arbitrary 10 mV to 2.5 V 20 mV to 5 V
Sine 10 mV to 2.5 V 20 mV to 5 V
Square 10 mV to 2.5 V 20 mV to 5 V
Pulse 10 mV to 2.5 V 20 mV to 5 V
Ramp 10 mV to 2.5 V 20 mV to 5 V
Triangle 10 mV to 2.5 V 20 mV to 5 V
Gaussian 10 mV to 1.25 V 20 mV to 2.5 V
Lorentz 10 mV to 1.2 V 20 mV to 2.4 V
Exponential Rise 10 mV to 1.25 V 20 mV to 2.5 V
Exponential Fall 10 mV to 1.25 V 20 mV to 2.5 V
Sine(x)/x 10 mV to 1.5 V 20 mV to 3.0 V
Random Noise 10 mV to 2.5 V 20 mV to 5 V
Haversine 10 mV to 1.25 V 20 mV to 2.5 V
Cardiac 10 mV to 2.5 V 20 mV to 5 V
Sine waveform
Frequency range
0.1 Hz to 50 MHz
Frequency setting resolution
0.1 Hz
Frequency accuracy

130 ppm (frequency ≤ 10 kHz), 50 ppm (frequency > 10 kHz)

This is for Sine, Ramp, Square and Pulse waveforms only.

Amplitude range
20 mVpp to 5 Vpp into Hi-Z; 10 mVpp to 2.5 Vpp into 50 Ω
Amplitude flatness, typical

±0.5 dB (relative to 1kHz level) at 30 MHz

±1.0 dB (relative to 1kHz level) at 50 MHz

Total harmonic distortion, typical

1% for amplitude ≥ 200 mVpp into 50 Ω load

2.5% for amplitude > 50 mV AND < 200 mVpp into 50 Ω load

Spurious free dynamic range, typical

40 dB (Vpp ≥ 0.1 V); 30 dB (Vpp ≥ 0.02 V), 50 Ω load

Square and pulse waveform
Frequency range
0.1 Hz to MHz
Frequency setting resolution
0.1 Hz
Frequency accuracy
130 ppm (frequency ≤ 10 kHz), 50 ppm (frequency > 10 kHz)
Amplitude range
20 mVpp to 5 Vpp into Hi-Z; 10 mVpp to 2.5 Vpp into 50 Ω
Duty cycle range
10% - 90% or 10 ns minimum pulse, whichever is larger
Minimum pulse time applies to both on and off time, so maximum duty cycle will reduce at higher frequencies to maintain 10 ns off time
Duty cycle resolution
0.1%
Minimum pulse width, typical
10 ns. This is the minimum time for either on or off duration.
Rise/Fall time, typical
5 ns, 10% - 90%
Pulse width resolution
100 ps
Overshoot, typical
< 6% for signal steps greater than 100 mVpp
This applies to overshoot of the positive-going transition (+overshoot) and of the negative-going (-overshoot) transition
Asymmetry, typical
±1% ±5 ns, at 50% duty cycle
Jitter, typical
< 60 ps TIERMS, ≥ 100 mVpp amplitude, 40%-60% duty cycle

Square and pulse waveforms, 5 GHz measurement BW.

Ramp and triangle waveform
Frequency range
0.1 Hz to
Frequency setting resolution
0.1 Hz
Frequency accuracy
130 ppm (frequency ≤ 10 kHz), 50 ppm (frequency > 10 kHz)
Amplitude range
20 mVpp to 5 Vpp into Hi-Z; 10 mVpp to 2.5 Vpp into 50 Ω
Variable symmetry
0% - 100%
Symmetry resolution
0.1%
DC level range

±2.5 V into Hi-Z

±1.25 V into 50 Ω

Random noise amplitude range

20 mVpp to 5 Vpp into Hi-Z

10 mVpp to 2.5 Vpp into 50 Ω

Sin(x)/x
Maximum frequency
MHz
Gaussian pulse, Haversine, and Lorentz pulse
Maximum frequency
MHz
Lorentz pulse
Frequency range
0.1 Hz to MHz
Amplitude range
20 mVpp to 2.4 Vpp into Hi-Z

10 mVpp to 1.2 Vpp into 50 Ω

Cardiac
Frequency range
0.1 Hz to
Amplitude range
20 mVpp to 5 Vpp into Hi-Z

10 mVpp to 2.5 Vpp into 50 Ω

Arbitrary
Memory depth
1 to 128 k
Amplitude range
20 mVpp to 5 Vpp into Hi-Z

10 mVpp to 2.5 Vpp into 50 Ω

Repetition rate
0.1 Hz to MHz
Sample rate
250 MS/s
Signal amplitude accuracy
±[ (1.5% of peak-to-peak amplitude setting) + (1.5% of absolute DC offset setting) + 1 mV ] (frequency = 1 kHz)
Signal amplitude resolution

1 mV (Hi-Z)

500 μV (50 Ω)

DC offset range

±2.5 V into Hi-Z

±1.25 V into 50 Ω

DC offset resolution

1 mV (Hi-Z)

500 μV (50 Ω)

DC offset accuracy

±[ (1.5% of absolute offset voltage setting) + 1 mV ]

Add 3 mV of uncertainty per 10 °C change from 25 °C ambient

Digital volt meter (DVM)

Measurement types

DC, ACRMS+DC, ACRMS, Trigger frequency count

Voltage resolution
4 digits
Voltage accuracy
DC:

±((1.5% * |reading - offset - position|) + (0.5% * |(offset - position)|) + (0.1 * Volts/div))

De-rated at 0.100%/°C of |reading - offset - position| above 30 °C

Signal ± 5 divisions from screen center

AC:

± 3% (40 Hz to 1 kHz) with no harmonic content outside 40 Hz to 1 kHz range

AC, typical: ± 2% (20 Hz to 10 kHz)

For AC measurements, the input channel vertical settings must allow the VPP input signal to cover between 4 and 10 divisions and must be fully visible on the screen

Trigger frequency counter

Resolution
8-digits
Accuracy

±(1 count + time base accuracy * input frequency)

The signal must be at least 8 mVpp or 2 div, whichever is greater.

Maximum input frequency
10 Hz to maximum bandwidth of the analog channel
The signal must be at least 8 mVpp or 2 div, whichever is greater.

Processor system

Host processor
Intel i5-4400E, 2.7 GHz, 64-bit, dual core processor, 8 GB system RAM
Operating system

Default instrument: Closed Embedded OS

Instrument with option 6-WIN installed: Microsoft Windows 10

Standard SSD with Embedded OS
≥ 250 GB removable solid state drive
Solid State Drive (SSD) with Microsoft Windows 10 OS (option 6-WIN)
≥ 512 GB SSD. Form factor is a 2.5-inch SSD with a SATA-3 interface. This drive is customer installable and includes the Microsoft Windows 10 Enterprise IoT 2016 LTSB (64-bit) operating system

Input-Output ports

DisplayPort connector
A 20-pin DisplayPort connector; connect to show the oscilloscope display on an external monitor or projector
DVI connector

A 29-pin DVI-I connector; connect to show the oscilloscope display on an external monitor or projector

VGA

DB-15 female connector; connect to show the oscilloscope display on an external monitor or projector

Probe compensator signal, typical
Connection:
Connectors are located on the lower front right of the instrument
Amplitude:
0 to 2.5 V
Frequency:
1 kHz
Source impedance:
1 kΩ
External reference input

The time-base system can phase lock to an external 10 MHz reference signal .

There are two ranges for the reference clock.

The instrument can accept a high-accuracy reference clock of 10 MHz ±2 ppm or a lower-accuracy reference clock of 10 MHz ±1 kppm.

USB interface (Host, Device ports)
Front panel USB Host ports: Two USB 2.0 Hi-Speed ports, one USB 3.0 SuperSpeed port
Rear panel USB Host ports: Two USB 2.0 Hi-Speed ports, two USB 3.0 SuperSpeed ports
Rear panel USB Device port: One USB 3.0 SuperSpeed Device port providing USBTMC support
Ethernet interface
10/100/1000 Mb/s
Auxiliary output

Rear-panel BNC connector. Output can be configured to provide a positive or negative pulse out when the oscilloscope triggers, the internal oscilloscope reference clock out, or an AFG sync pulse

Characteristic Limits
Vout (HI) ≥ 2.5 V open circuit; ≥ 1.0 V into a 50 Ω load to ground
Vout (LO) ≤ 0.7 V into a load of ≤ 4 mA; ≤0.25 V into a 50 Ω load to ground
Kensington-style lock
Rear-panel security slot connects to standard Kensington-style lock
LXI

Class: LXI Core 2011

Version: 1.4

Power source

Power
Power consumption

400 Watts maximum

Source voltage
100 - 240 V ±10% at 50 Hz to 60 Hz
115 V ±10% at 400 Hz

Physical characteristics

Dimensions

Height: 12.2 in (309 mm), feet folded in, handle to back

Height: 14.6 in (371 mm) feet folded in, handle up

Width: 17.9 in (454 mm) from handle hub to handle hub

Depth: 8.0 in (205 mm) from back of feet to front of knobs, handle up

Depth: 11.7 in (297.2 mm) feet folded in, handle to the back

Weight
< 29.8 lbs (13.52 kg)
Cooling
The clearance requirement for adequate cooling is 2.0 in (50.8 mm) on the right side of the instrument (when viewed from the front) and on the rear of the instrument
Rackmount configuration
7U (with optional RM5 Rackmount Kit)

Environmental specifications

Temperature
Operating
+0 °C to +50 °C (32 °F to 122 °F)
Non-operating
-20 °C to +60 °C (-4 °F to 140 °F)
Humidity
Operating
5% to 90% relative humidity (% RH) at up to +40 °C
5% to 55% RH above +40 °C up to +50 °C, noncondensing
Non-operating
5% to 90% relative humidity (% RH) at up to +60 °C, noncondensing
Altitude
Operating
Up to 3,000 meters (9,843 feet)
Non-operating
Up to 12,000 meters (39,370 feet)
Temperature
Operating
+0 °C to +50 °C (32 °F to 122 °F)
Non-operating

-20 °C to +60 °C (-4 °F to 140 °F)

Humidity
Operating

5% to 90% relative humidity (% RH) at up to +40 °C

5% to 55% RH above +40 °C up to +50 °C, noncondensing

Non-operating

5% to 90% relative humidity (% RH) at up to +60 °C, noncondensing

Altitude
Operating
Up to 3,000 meters (9,843 feet)
Non-operating
Up to 12,000 meters (39,370 feet)

EMC, Environmental, and Safety

Safety certification
US NRTL Listed - UL61010-1 and UL61010-2-030
Canadian Certification - CAN/CSA-C22.2 No. 61010.1 and CAN/CSA-C22.2 No 61010.2.030
EU Compliance - Low Voltage Directive 2014-35-EU and EN61010-1.
International Compliance - IEC 61010-1 and IEC61010-2-030
Regulatory

CE marked for the European Union and UL approved for the USA and Canada

RoHS compliant

Software

IVI driver

Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI, Microsoft .NET, and MATLAB. Compatible with Python, C/C++/C# and many other languages through VISA.

e*Scope®

Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or network name of the oscilloscope and a web page will be served to the browser. Transfer and save settings, waveforms, measurements, and screen images or make live control changes to settings on the oscilloscope directly from the web browser. Optionally configure e*Scope authentication to password protect access to control and view the oscilloscope.

TekDrive
Upload, store, organize, search, download, and share any file type from any connected device. TekDrive is natively integrated into the instrument for seamless sharing and recalling of files - no USB stick is required. Analyze and explore standard files like .wfm, .isf, .tss, and .csv, directly in a browser. Visit http://www.tek.com/software/tekdrive to learn more.
SignalVu-PC
Advanced vector signal analysis software that can run directly on your 6 Series MSO or on a separate Windows PC. Requires option 6-SV-RFVT installed on the 6 Series MSO. Requires Connect license (CONxx-SVPC) installed on SignalVu-PC, xx is NL for Node Locked license or FL for Floating License.
LXI Web interface

Connect to the oscilloscope through a standard Web browser by simply entering the oscilloscope's IP address or network name in the address bar of the browser. The Web interface enables viewing of instrument status and configuration, status and modification of network settings, and instrument control through the e*Scope web-based remote control.

Programming Examples

Programming with the 4/5/6 Series platforms has never been easier. With a programmers manual and a GitHub site you have many commands and examples to help you get started remotely automating your instrument. See https://github.com/tektronix/programmatic-control-examples.

Ordering information

Use the following steps to select the appropriate instrument and options for your measurement needs.

Step 1

Start by selecting the MSO64 model.
Model Number of FlexChannels
MSO644
Each model includes
Four TPP1000 1 GHz probes
Installation and safety manual
Embedded Help
Front cover with integrated accessory pouch
Power cord
Calibration certificate documenting traceability to National Metrology Institute(s) and ISO9001/ISO17025 quality system registration
Three -year warranty covering all parts and labor on the instrument.
One-year warranty covering all parts and labor on included probes

Step 2

Configure your oscilloscope by selecting the analog channel bandwidth you need
Choose the bandwidth you need today by choosing one of these bandwidth options. You can upgrade it later by purchasing an upgrade option.
Bandwidth Option Bandwidth
6-BW-1000 1 GHz
6-BW-2500 2.5 GHz
6-BW-4000 4 GHz
6-BW-6000 6 GHz
6-BW-8000 8 GHz

Note: For instruments of 4, 6, or 8 GHz bandwidth, consider a BNC-to-SMA adapter to optimize a high bandwidth connection to the oscilloscope. Tektronix part number 103-0503-xx.

Step 3

Add instrument functionality by adding an option bundle

Three classes of option bundles are offered (Starter, Pro, Ultimate), providing a range of options depending on your budget and application needs. For detailed information on the current contents of each bundle, please visit our website and view the software bundle brochure at https://www.tek.com/document/brochure/software-bundles-for-the-4-5-and-6-series-mso-oscilloscopes.

  1. Starter bundle offers the most common serial bus decoding, protocol analysis, power measurements and analysis, and hardware enhancing options bundled together.
  2. Pro bundles are application-specific (Serial trigger and decode, Power Integrity, Signal Integrity, Automotive, Automated Compliance Test, Military Government Aerospace), and include all options from the Starter bundle.
  3. Ultimate bundle includes all options from the Starter bundle in addition to the all options from all Pro bundles.

1 Year licensePerpetual licenseBundle description
6-STARTER-1Y 6-STARTER-PER Includes I2C, SPI, RS-232/422/UART serial trigger and analysis, AFG (Arbitrary/Function Generator), and power measurements and analysis
6-PRO-SERIAL-1Y 6-PRO-SERIAL-PER Includes 6-STARTER plus 250 MS/ch record length, and additional select serial analysis options
6-PRO-POWER-1Y 6-PRO-POWER-PER Includes 6-STARTER plus 250 MS/ch record length, and select power analysis options
6-PRO-SIGNAL-1Y 66-PRO-SIGNAL-PER Includes 6-STARTER plus 250 MS/ch record length, advanced Jitter, and select analysis options
6-PRO-COMPL-1Y 66-PRO-COMPL-PER Includes 6-STARTER plus 250 MS/ch record length, advanced Jitter, and select automated compliance test options
6-PRO-AUTO-1Y 66-PRO-AUTO-PER Includes 6-STARTER plus 250 MS/ch record length, advanced Jitter and select automotive analysis options
6-PRO-MILGOV-1Y 6-PRO-MILGOV-PER Includes 6-STARTER plus 250 MS/ch record length, advanced Jitter, mask test, and select serial analysis options
6-ULTIMATE-1Y6-ULTIMATE-PER Includes 6-STARTER, all 6-PRO bundle options plus 1 GS/ch record length, RF vs. Time traces, triggers, Spectrograms, and IQ capture, extended Spectrum View capture bandwidth, and video trigger options

Each purchased bundle has two duration options:

  • A 1-year subscription includes all features and free upgrades for the purchased bundle for one year; after which time the features are disabled. Additional 1-year subscription can be purchased for the selected bundle.
  • A perpetual subscription enables all features for the purchased bundle permanently. A perpetual subscription includes 1-year of free upgrades to the bundle feature set. After the year, the feature set is frozen to those enabled by the last update made.

    Perpetual bundles can continue to receive upgrades following the 1 year activation period with the purchase of a maintenance license. Maintenance license information can be found in the maintenance license table below and must be purchased for an existing Starter, Pro, or Ultimate bundle.

    Maintenance licenseDescription
    6-STARTER-MNT-1YIncludes Perpetual Starter Bundle updates for 1 Year on 6 Series MSO
    6-PRO-MNT-1YIncludes Perpetual Pro Bundle updates for 1 Year on 6 Series MSO
    6-ULTIMATE-MNT-1YIncludes Perpetual Ultimate Bundle updates for 1 Year on 6 Series MSO

Step 4

Add instrument functionality

Instrument functionality can be ordered with the instrument or later as an upgrade kit.

Instrument option Built-in functionality
6-RL-1 Extend record length from 62.5 Mpoints/channel to 125 Mpoints/channel
6-RL-2 Extend record length from 62.5 Mpts/channel to 250 Mpts/channel
6-RL-3

Extend record length from 62.5 Mpoints/channel to 500 Mpoints/channel

6-RL-4

Extend record length from 62.5 Mpoints/channel to 1 Gpoints/channel

6-WIN 7Add removable SSD with Microsoft Windows 10 operating system license
6-AFGAdd Arbitrary/Function Generator
6-SEC 89Add enhanced security for instrument declassification and password-protected enabling and disabling of all USBports and firmware upgrade.

Step 5

Add optional protocol triggering, decode, and search capabilities
Choose the protocol support you need today by choosing from these analysis options. You can upgrade later by purchasing an upgrade kit.
Instrument option Protocols supported
6-SRAERO Aerospace (MIL-STD-1553, ARINC 429)
6-SRAUDIO Audio (I2S, LJ, RJ, TDM)
6-SRAUTO Automotive (CAN, CAN FD, CAN XL, LIN, FlexRay, and CAN symbolic decoding)
6-SRAUTOEN1 100BASE-T1 Automotive Ethernet serial analysis
6-SRAUTOSEN Automotive sensor (SENT)
6-SRCOMP Computer (RS-232/422/485/UART)
6-SRCPHY MIPI C-PHY (DSI-2, CSI-2 decode and search only)
6-SRCXPICXPI (decode and search only)
6-SRDPHY MIPI D-PHY V1.2 (DSI-1, CSI-2 decode and search only)
6-SREMBD Embedded (I2C, SPI)
6-SRENETEthernet (10BASE-T, 100BASE-TX)
6-SRESPIeSPI (decode and search only)
6-SR8B10B 8B/10B (decode and search only)
6-SRI3C MIPI I3C
6-SRMANCHManchester (decode and search only)
6-SRMDIO MDIO (decode and search only)
6-SRNRZNRZ (decode and search only)
6-SRONEWIRE One wire (1-Wire decode and search only)
6-SRPMPower Management (SPMI)
6-SRPSI5 PSI5 (decode and search only)
6-SRSDLC​Synchronous Data Link Control Protocol Decode & Search
6-SRSPACEWIRESpacewire (decode and search only)
6-SRSVID SVID
6-SRUSB2 USB (USB2.0 LS, FS, HS)
6-SREUSB2 eUSB2.0 (decode and search only)

Differential serial bus? Be sure to check Add analog probes and adapters for differential probes.

Add third party serial bus decode and analysis capabilities
Third-party applications are available that provide serial bus decode and analysis capabilities to use on the 6 Series MSO. Use of the third-party software applications require a Windows 10 SSD (option 6-WIN).
Please see additional information at https://prodigytechno.com/oscilloscope-based-protocol-decode-software/

Step 6

Add optional serial bus compliance testing

Choose the serial compliance testing packages you need today by choosing from these options. You can upgrade later by purchasing an upgrade kit. All options in the table below require option 6-WIN (SSD with Microsoft Windows 10 operating system).

Instrument option Serial buses supported
6-CMAUTOEN Automotive Ethernet (100Base-T1, 1000Base-T1) automated compliance test solution.

≥2 GHz bandwidth required for 1000BASE-T1

6-CMAUTOEN10 Automotive Ethernet (10BASE-T1S Short Reach) automated compliance test solution.
6-CMAUTOEN10G Automotive Ethernet (MultiGBase-T1) automated compliance test solution. Requires option 6-DJA.
6-AUTOEN-BND Automotive Ethernet Compliance, Signal Separation, PAM3 Analysis, 100Base-T1 Decode software (requires options 6-DJA and 6-WIN)
6-AUTOEN-SS Automotive Ethernet Signal Separation
6-CMINDUEN10 Industrial Ethernet (10Base-T1L Long Reach) automated compliance test solution
6-CMCPHY20 MIPI C-PHY 2.0 Tx Test automated conformance test solution (requires option 6-DJA)
6-CMDPHY MIPI D-PHY 1.2 automated compliance test solution
6-CMDPHY21MIPI D-PHY 2.1 Tx Test automated conformance test solution (requires option 6-DJA)
6-CMENETEthernet automated compliance test solution (10BASE-T/100BASE-T/1000BASE-T).

≥1 GHz bandwidth required for 1000BASE-T

6-CMENETMLMultilane Ethernet (10Base-T, 100Base-T, 1000Base-T) automated compliance test solution
6-CMNBASET 2.5 and 5 GBASE-T Ethernet automated compliance test solution.

2.5 GHz is recommended

6-CMXGBT 10 GBASE-T Ethernet automated compliance test solution.

≥4 GHz is recommended

6-CMUSB2USB2.0 automated compliance test solution.

Requires TDSUSBF USB test fixture

≥2 GHz bandwidth required for high-speed USB

Step 6

Add optional serial bus compliance testing

Choose the serial compliance testing packages you need today by choosing from these options. You can upgrade later by purchasing an upgrade kit. All options in the table below require option 6-WIN (SSD with Microsoft Windows 10 operating system).

Instrument option Serial buses supported
6-CMAUTOEN Automotive Ethernet (100Base-T1, 1000Base-T1) automated compliance test solution.

≥2 GHz bandwidth required for 1000BASE-T1

6-CMAUTOEN10 Automotive Ethernet (10BASE-T1S Short Reach) automated compliance test solution.
6-CMAUTOEN10G Automotive Ethernet (MultiGBase-T1) automated compliance test solution.
6-AUTOEN-BND Automotive Ethernet Compliance, Signal Separation, PAM3 Analysis, 100Base-T1 Decode software (requires option 6-DJA and 6-WIN)
6-AUTOEN-SS Automotive Ethernet Signal Separation
6-CMCPHY20MIPI C-PHY 2.0 Tx Test automated conformance test solution (requires option 6-DJA)
6-CMDPHY MIPI D-PHY 1.2 automated compliance test solution
6-CMDPHY21MIPI D-PHY 2.1 Tx Test automated compliance test solution (requires option 6-DJA)
6-CMENETEthernet automated compliance test solution (10BASE-T/100BASE-T/1000BASE-T).

≥1 GHz bandwidth required for 1000BASE-T

6-CMENETMLMultilane Ethernet (10Base-T, 100Base-T, 1000Base-T) automated compliance test solution

6-CMNBASET

2.5 and 5 GBASE-T Ethernet automated compliance test solution.

2.5 GHz is recommended

6-CMXGBT

10 GBASE-T Ethernet automated compliance test solution.

≥4 GHz is recommended

6-CMUSB2USB2.0 automated compliance test solution.

Requires TDSUSBF USB test fixture

≥2 GHz bandwidth required for high-speed USB

Step 7

Add optional memory analysis
Instrument option Advanced analysis
6-DBDDR3 DDR3 and LPDDR3 Debug and Analysis
6-CMDDR3 DDR3 and LPDDR3 automated compliance test solution using TekExpress Automation Platform.

Requires options 6-DBDDR3, 6-DJA and 6-WIN (SSD with Microsoft Windows 10 operating system).

≥4 GHz required, 8 GHz recommended for testing of all DDR3 speeds.

Step 8

Add optional analysis capabilities
Instrument option Advanced analysis
6-DBLVDS TekExpress automated LVDS test solution (requires options 6-DJA and 6-WIN)
6-DJA Advanced Jitter and Eye Analysis
6-DPM Digital Power Management
6-MTM Mask and Limit testing
6-PAM3 PAM3 analysis (requires options 6-DJA and 6-WIN)
6-PS2 10Power solution bundle (6-PWR, THDP0200, TCP0030A, 067-1686-XX deskew fixture)
6-PWR 11Power Measurement and Analysis
6-SV-BW-1 Increase Spectrum View Capture Bandwidth to 2 GHz
6-SV-RFVTSpectrum View RF vs. Time traces, triggers, Spectrograms, and IQ capture
6-TDRTime Domain Reflectometry
6-UDFLTUser Defined Filter Creation Tool
6-VID NTSC, PAL, and SECAM video triggering
6-WBG-DPTWide Bandgap SiC/GaN Double Pulse Test Measurements and Analysis
Add vector signal analysis
SignalVu-PC is a stand-alone application that can be run on a 6 Series MSO or on a separate Windows PC to provide advanced vector signal analysis. In order to run SignalVu-PC on your 6 Series MSO, three options are required.
  1. To run the application on the instrument, the Windows SSD (6-WIN) needs to be installed in the oscilloscope.
  2. The Spectrum View RF versus time traces option (6-SV-RFVT) needs to be installed in the oscilloscope to enable I/Q data transfer.
  3. The Connect (CONxx-SVPC) license needs to be installed in SignalVu-PC to enable base features of the application, which includes 16+ RF measurements and displays.

Step 9

Add digital probes
Each FlexChannel input can be configured as eight digital channels simply by connecting a TLP058 logic probe to a FlexChannel input..
For this instrument Order To add
MSO64 1 to 4 TLP058 Probes 8 to 32 digital channels

Step 10

Add analog probes and adapters
Add additional recommended probes and adapters
Recommended Probe / Adapter Description
TAP1500 1.5 GHz TekVPI® active single-ended voltage probe, ±8 V input voltage
TAP2500 2.5 GHz TekVPI® active single-ended voltage probe, ±4 V input voltage
TAP3500 3.5 GHz TekVPI® active single-ended voltage probe, ±4 V input voltage
TAP4000 4 GHz TekVPI® active single-ended voltage probe, ±4 V input voltage
TCP0020 20 A AC/DC TekVPI® current probe, 50 MHz BW
TCP0030A 30 A AC/DC TekVPI current probe, 120 MHz BW
TCP0150 150 A AC/DC TekVPI® current probe, 20 MHz BW
TCPA300 100 MHz Current Probe, Amplifier (Requires Probe); Recommend using TPA-BNC adapter to provide autoscaling.
TCP312A DC-100 MHz, AC/DC Current Probe; 30 Amp DC
TRCP0300 30 MHz AC current probe, 250 mA to 300 A
TRCP0600 30 MHz AC current probe, 500 mA to 600 A
TRCP3000 16 MHz AC current probe, 500 mA to 3000 A
TDP0500 500 MHz TekVPI® differential voltage probe, ±42 V differential input voltage
TDP1000 1 GHz TekVPI® differential voltage probe, ±42 V differential input voltage
TDP1500 1.5 GHz TekVPI® differential voltage probe, ±8.5 V differential input voltage
TDP3500 3.5 GHz TekVPI® differential voltage probe, ±2 V differential input voltage
TDP4000 4 GHz TekVPI® differential voltage probe, ±2 V differential input voltage
TDP7704 4 GHz TriMode™ voltage probe
TDP7706 6 GHz TriMode™ voltage probe
TDP7708 8 GHz TriMode™ voltage probe
TDP771010 GHz TriMode™ voltage probe
THDP0100 ±6 kV, 100 MHz TekVPI® high-voltage differential probe
THDP0200 ±1.5 kV, 200 MHz TekVPI® high-voltage differential probe
TMDP0200 ±750 V, 200 MHz TekVPI® high-voltage differential probe
TPR1000 1 GHz, Single-Ended TekVPI® Power-Rail Probe; includes one TPR4KIT accessory kit
TPR4000 4 GHz, Single-Ended TekVPI® Power-Rail Probe; includes one TPR4KIT accessory kit
TIVP02Isolated Probe; 200 MHz, ±5 V to ±2500 V depending on tip; 2 meter cable
TIVP02LIsolated Probe; 200 MHz, ±5 V to ±2500 V depending on tip; 10 meter cable
TIVP05Isolated Probe; 500 MHz, ±5 V to ±2500 V depending on tip; 2 meter cable
TIVP05LIsolated Probe; 500 MHz, ±5 V to ±2500 V depending on tip; 10 meter cable
TIVP1Isolated Probe; 1 GHz, ±5 V to ±2500 V depending on tip; 2 meter cable
TIVP1LIsolated Probe; 1 GHz, ±5 V to ±2500 V depending on tip; 10 meter cable
TPP0502 500 MHz, 2X TekVPI® passive voltage probe, 12.7 pF input capacitance
TPP0850 2.5 kV, 800 MHz, 50X TekVPI® passive high-voltage probe
P6015A 20 kV, 75 MHz high-voltage passive probe
TPA-BNC TekVPI® to TekProbe™ BNC adapter (recommended for connecting your existing TekProbe probes to this instrument)
103-0503-xx BNC-to-SMA adapter; rated to 12 GHz
TEK-DPG TekVPI deskew pulse generator signal source
067-1686-xx Power measurement deskew and calibration fixture
Looking for other probes? Check out the probe selector tool at www.tek.com/probes.

Step 11

Add accessories
Add traveling or mounting accessories
Optional Accessory Description
HC5 Hard carrying case
RM5 Rackmount kit
GPIB to Ethernet adapter Order model 4865B (GPIB to Ethernet to Instrument Interface) directly from ICS Electronics

www.icselect.com/gpib_instrument_intfc.html

Step 12

Select power cord option
Power Cord Option Description
A0 North America power plug (115 V, 60 Hz)
A1 Universal Euro power plug (220 V, 50 Hz)
A2 United Kingdom power plug (240 V, 50 Hz)
A3 Australia power plug (240 V, 50 Hz)
A5 Switzerland power plug (220 V, 50 Hz)
A6 Japan power plug (100 V, 50/60 Hz)
A10 China power plug (50 Hz)
A11 India power plug (50 Hz)
A12 Brazil power plug (60 Hz)
A99 No power cord

Step 13

Add extended service and calibration options
Service Option Description
T3 Three-year Total Protection Plan, includes repair or replacement coverage from wear and tear, accidental damage, ESD or EOS.
R3 Standard warranty extended to 3 years. Covers parts, labor and 2-day shipping within country. Guarantees faster repair time than without coverage. All repairs include calibration and updates. Hassle free - a single call starts the process.
C3 Calibration service for 3 years. Includes traceable calibration or functional verification where applicable, for recommended calibrations. Coverage includes the initial calibration plus 2 years of calibration coverage.
T5 Five year Total Protection Plan, includes repair or replacement coverage from wear and tear, accidental damage, ESD or EOS.
R5 Standard warranty extended to 5 years. Covers parts, labor and 2-day shipping within country. Guarantees faster repair time than without coverage. All repairs include calibration and updates. Hassle free - a single call starts the process.
C5 Calibration service for 5 years. Includes traceable calibration or functional verification where applicable, for recommended calibrations. Coverage includes the initial calibration plus 4 years of calibration coverage.

Feature upgrades after purchase

Add feature upgrades in the future
The 6 Series products offer many ways to easily add functionality after the initial purchase. Node-locked licenses permanently enable optional features on a single product. Floating licenses allow license-enabled options to be easily moved between compatible instruments.
Upgrade feature Node-locked license upgrade Floating license upgrade Description
Add instrument functions SUP6-AFG SUP6-AFG-FL Add arbitrary function generator
SUP6-RL-1 SUP6-RL-1-FL Extend record length from 62.5 Mpts to 125 Mpts/channel
SUP6-RL-2 SUP6-RL-2-FL Extend record length from 62.5 Mpts to 250 Mpts/channel
SUP6-RL-3 SUP6-RL-3-FL Extend record length from 62.5 Mpts to 500 Mpts/channel
SUP6-RL-4 SUP6-RL-4-FL Extend record length from 62.5 Mpts to 1 Gpts/channel
SUP6-RL-1T2 SUP6-RL-1T2-FL Extend record length from 125 Mpts to 250 Mpts/channel
SUP6-RL-1T3 SUP6-RL-1T3-FL Extend record length from 125 Mpts to 500 Mpts/channel
SUP6-RL-1T4 SUP6-RL-1T4-FL Extend record length from 125 Mpts to 1 Gpts/channel
SUP6-RL-2T3 SUP6-RL-2T3-FL Extend record length from 250 Mpts to 500 Mpts/channel
SUP6-RL-2T4 SUP6-RL-2T4-FL Extend record length from 250 Mpts to 1 Gpts/channel
SUP6-RL-3T4 SUP6-RL-3T4-FL Extend record length from 500 Mpts to 1 Gpts/channel
Add protocol analysis SUP6-RFNFCSUP6-RFNFC-FLISO/IEC 15693, 14443A, 14443B, and FeliCa (decode and search only)
SUP6-SRAERO SUP6-SRAERO-FL Aerospace serial triggering and analysis (MIL-STD-1553, ARINC 429)
SUP6-SRAUDIO SUP6-SRAUDIO-FL Audio serial triggering and analysis (I2S, LJ, RJ, TDM)
SUP6-SRAUTO SUP6-SRAUTO-FL Automotive serial triggering and analysis (CAN, CAN FD, CAN XL, LIN, FlexRay, and CAN symbolic decoding)
SUP6-SRAUTOEN1 SUP6-SRAUTOEN1-FL 100Base-T1 Automotive Ethernet serial analysis
SUP6-SRAUTOSEN SUP6-SRAUTOSEN-FL Automotive sensor serial triggering and analysis (SENT)
SUP6-SRCOMP SUP6-SRCOMP-FL Computer serial triggering and analysis (RS-232/422/485/UART)
SUP6-SRCPHY SUP6-SRCPHY-FL MIPI C-PHY serial analysis (DSI-2, CSI-2)
SUP6-SRCXPI SUP6-SRCXPI-FLCXPI serial decoding and analysis
SUP6-SRDPHY SUP6-SRDPHY-FL MIPI D-PHY serial analysis (DSI-1, CSI-2)
SUP6-SREMBD SUP6-SREMBD-FL Embedded serial triggering and analysis (I2C, SPI)
SUP6-SRENET SUP6-SRENET-FL Ethernet serial triggering and analysis (10Base-T, 100Base-TX)
SUP6-SRESPI SUP6-SRESPI-FLeSPI serial decoding and analysis
SUP6-SRETHERCATSUP6-SRETHERCAT-FLEtherCAT serial decoding and analysis
SUP6-CMXGBTSUP6-CMXGBT-FL10 GBASE-T Ethernet automated compliance test solution ≥4 GHz is recommended
SUP6-SREUSB2 SUP6-SREUSB2-FL. Embedded USB2 (eUSB2) serial decoding and analysis
SUP6-SRI3C SUP6-SRI3C-FL MIPI I3C serial triggering and analysis
SUP6-SRMANCH SUP6-SRMANCH-FL Manchester serial analysis
SUP6-SRMDIO SUP6-SRMDIO-FL Management Data Input/Output (MDIO) serial decoding and analysis
SUP6-SR8B10B SUP6-SR8B10B-FL 8b/10b serial decoding and analysis
SUP6-SRNRZ SUP6-SRNRZ-FL NRZ serial decoding and analysis
SUP6-SRONEWIRE SUP6-SRONEWIRE-FL One wire (1-Wire) serial decoding and analysis
SUP6-SRPCIE321 SUP6-SRPCIE321-FL PCIe Gen1, Gen2 decoding and analysis
SUP6-SRPM SUP6-SRPM-FL Power Management serial triggering and analysis (SPMI)
SUP6-SRPSI5 SUP6-SRPSI5-FL PSI5 serial decoding and analysis
SUP6-SRSDLCSUP6-SRSDLC-FLSynchronous Data Link Control (SDLC) serial decoding and analysis
SUP6-SRSMBUS SUP6-SRSMBUS-FLSMBus serial decoding and analysis
SUP6-SRSPACEWIRE SUP6-SRSPACEWIRE-FL Spacewire serial analysis
SUP6-SRSVID SUP6-SRSVID-FL Serial Voltage Identification (SVID) serial triggering and analysis
SUP6-SRUSB2 SUP6-SRUSB2-FL USB 2.0 serial bus triggering and analysis (1.5 Mbps, 12 Mbps, 480 Mbps)
SUP6-SRUSB3 SUP6-SRUSB3-FLUSB 3.0 decoding and analysis

Add serial compliance

All serial compliance products require option 6-WIN (SSD with Microsoft Windows 10 operating system)

SUP6-CMAUTOEN SUP6-CMAUTOEN-FL Automotive Ethernet automated compliance test solution (100BASE-T1 and 1000BASE-T1)
SUP6-CMAUTOEN10 SUP6-CMAUTOEN10-FL Automotive Ethernet (10BASE-T1S Short Reach) automated compliance test solution
SUP6-CMAUTOEN10G SUP6-CMAUTOEN10G-FLAutomotive Ethernet (MultiGigBase-T1) automated compliance test solution (requires option 6-DJA)
SUP6-AUTOEN-BND Automotive Ethernet compliance, signal separation, PAM3 analysis, 100Base-T1 serial analysis (requires option 6-DJA)
SUP6-AUTOEN-SS SUP6-AUTOEN-SS-FL Automotive Ethernet signal separation
SUP6-CMINDUEN10 SUP6-CMINDUEN10-FL Industrial Ethernet (10Base-T1L Long Reach) automated compliance test solution
SUP6-CMCPHY20 SUP6-CMCPHY20-FL MIPI C-PHY 2.0 Tx automated conformance test solution (requires option 6-DJA)
SUP6-CMDPHY SUP6-CMDPHY-FL MIPI D-PHY 1.2 automated compliance test solution
SUP6-CMDPHY21SUP6-CMDPHY21-FLMIPI D-PHY 2.1 Tx automated conformance test solution (requires option 6-DJA)
SUP6-CMENET SUP6-CMENET-FL Ethernet automated compliance test solution (10BASE-T, 100BASE-T, and 1000BASE-T)
SUP6-CMENETMLSUP6-CMENETML-FLMultilane Ethernet automated compliance test solution (10Base-T, 100Base-T, 1000Base-T) (requires option 6-CMENET)
SUP6-CMNBASET SUP6-CMNBASET-FL 2.5 and 5 GBASE-T Ethernet automated compliance test (2.5 GHz is recommended)
SUP6-CMUSB2 SUP6-CMUSB2-FL USB 2.0 automated compliance test solution
Add advanced analysis SUP6-DBLVDS SUP6-DBLVDS-FL LVDS debug and analysis (requires options 6-DJA and 6-WIN)
SUP6-DJA SUP6-DJA-FL Advanced jitter and eye analysis
SUP6-DPM SUP6-DPM-FL Digital power management
SUP6-MTM SUP6-MTM-FL Mask and Limit testing
SUP6-PAM3 SUP6-PAM3-FL PAM3 analysis (requires options 6-DJA and 6-WIN)
SUP6-PS2 N/A Power solution bundle (6-PWR, THDP0200, TCP0030A, and 067-1686-XX deskew fixture)
SUP6-PWR SUP6-PWR-FL Advanced power measurements and analysis
SUP6-SV-BW-1 SUP6-SV-BW-1-FL Increase Spectrum View capture bandwidth to 2 GHz
SUP6-SV-RFVT SUP6-SV-RFVT-FL Spectrum View RF vs. Time traces, triggers, Spectrograms, and IQ capture
SUP6-UDFLT SUP6-UDFLT-FLUser defined filter creation tool
SUP6-VID SUP6-VID-FL NTSC, PAL, and SECAM video triggering
SUP6-WBG-DPTSUP6-WBG-DPT-FLWide Bandgap SiC/GaN Double Pulse Test Measurements and Analysis
SUP6B-IMDA SUP6B-IMDA-FL Inverter Motor Drive analysis
SUP6B-IMDA-DQ0 SUP6B-IMDA-DQ0-FL DQ0 feature for Inverter Motor Drive analysis (requires option 6-IMDA)
SUP6B-IMDA-MECHSUP6B-IMDA-MECH-FLMechanical measurements for Inverter Motor Drive analysis (requires option 6-IMDA)
SUP6-TDRSUP6-TDR-FLTime Domain Reflectometry
Add memory analysis SUP6-DBDDR3 SUP6-DBDDR3-FL DDR3 and LPDDR3 debug and analysis
SUP6-CMDDR3 SUP6-CMDDR3-FL DDR3 and LPDDR3 automated compliance test solution using TekExpress Automation Platform.

Requires options 6-DBDDR3, 6-DJA and SSD with Microsoft WIndows 10 operating system.

≥4 GHz required, 8 GHz recommended for testing of all DDR3 speeds.

Add digital voltmeter N/A N/A Add digital voltmeter/trigger frequency counter

(Free with product registration at www.tek.com/register6mso)

Upgrade feature Node-locked license upgrade Floating license upgrade Description
Add instrument functions SUP6-AFG SUP6-AFG-FL Add arbitrary function generator
SUP6-RL-1 SUP6-RL-1-FL Extend record length from 62.5 Mpts to 125 Mpts/channel
SUP6-RL-2 SUP6-RL-2-FL Extend record length from 62.5 Mpts to 250 Mpts/channel
SUP6-RL-3 SUP6-RL-3-FL Extend record length from 62.5 Mpts to 500 Mpts/channel
SUP6-RL-4 SUP6-RL-4-FL Extend record length from 62.5 Mpts to 1 Gpts/channel
SUP6-RL-1T2 SUP6-RL-1T2-FL Extend record length from 125 Mpts to 250 Mpts/channel
SUP6-RL-1T3 SUP6-RL-1T3-FL Extend record length from 125 Mpts to 500 Mpts/channel
SUP6-RL-1T4 SUP6-RL-1T4-FL Extend record length from 125 Mpts to 1 Gpts/channel
SUP6-RL-2T3 SUP6-RL-2T3-FL Extend record length from 250 Mpts to 500 Mpts/channel
SUP6-RL-2T4 SUP6-RL-2T4-FL Extend record length from 250 Mpts to 1 Gpts/channel
SUP6-RL-3T4 SUP6-RL-3T4-FL Extend record length from 500 Mpts to 1 Gpts/channel
Add protocol analysis SUP6-RFNFCSUP6-RFNFC-FLISO/IEC 15693 and ISO/IEC14443A (decode and search only)
SUP6-SRAERO SUP6-SRAERO-FL Aerospace serial triggering and analysis (MIL-STD-1553, ARINC 429)
SUP6-SRAUDIO SUP6-SRAUDIO-FL Audio serial triggering and analysis (I2S, LJ, RJ, TDM)
SUP6-SRAUTO SUP6-SRAUTO-FL Automotive serial triggering and analysis (CAN, CAN FD, LIN, FlexRay, and CAN symbolic decoding)
SUP6-SRAUTOEN1 SUP6-SRAUTOEN1-FL 100Base-T1 Automotive Ethernet serial analysis
SUP6-SRAUTOSEN SUP6-SRAUTOSEN-FL Automotive sensor serial triggering and analysis (SENT)
SUP6-SRCOMP SUP6-SRCOMP-FL Computer serial triggering and analysis (RS-232/422/485/UART)
SUP6-SRCPHY SUP6-SRCPHY-FL MIPI C-PHY serial analysis (DSI-2, CSI-2)
SUP6-SRCXPI SUP6-SRCXPI-FLCXPI serial decoding and analysis
SUP6-SRDPHYSUP6-SRDPHY-FLMIPI D-PHY V1.2 (DSI-1, CSI-2) serial decode and analysis
SUP6-SREMBD SUP6-SREMBD-FL Embedded serial triggering and analysis (I2C, SPI)
SUP6-SRENET SUP6-SRENET-FL Ethernet serial triggering and analysis (10Base-T, 100Base-TX)
SUP6-SRESPI SUP6-SRESPI-FLeSPI serial decoding and analysis
SUP6-SRETHERCATSUP6-SRETHERCAT-FLEtherCAT serial decoding and analysis
SUP6-SRI3C SUP6-SRI3C-FL MIPI I3C serial triggering and analysis
SUP6-SRMANCHSUP6-SRMANCH-FLManchester serial decoding and analysis
SUP6-SRMDIO SUP6-SRMDIO-FL Management Data Input/Output (MDIO) serial decoding and analysis
SUP6-SR8B10B SUP6-SR8B10B-FL 8b/10b serial decoding and analysis
SUP6-SRNRZ SUP6-SRNRZ-FL NRZ serial decoding and analysis
SUP6-SRONEWIRE SUP6-SRONEWIRE-FLOne wire (1-Wire) serial decoding and analysis
SUP6-SRPCIE321 SUP6-SRPCIE321-FL PCIe Gen1, Gen2 decoding and analysis
SUP6-SRPM SUP6-SRPM-FL Power Management serial triggering and analysis (SPMI)
SUP6-SRPSI5 SUP6-SRPSI5-FL PSI5 serial decoding and analysis
SUP6-SRSDLCSUP6-SRSDLCSynchronous Data Link Controller serial decoding and analysis
SUP6-SRSMBUS SUP6-SRSMBUS-FLSMBus serial decoding and analysis
SUP6-SRSPACEWIRE SUP6-SRSPACEWIRE-FL Spacewire serial analysis
SUP6-SRSVID SUP6-SRSVID-FL Serial Voltage Identification (SVID) serial triggering and analysis
SUP6-SRUSB2 SUP6-SRUSB2-FL USB 2.0 serial bus triggering and analysis (LS, FS, HS)
SUP6-SREUSB2 SUP6-SREUSB2-FL Embedded USB2 (eUSB2) serial decoding and analysis
SUP6-SRUSB3 SUP6-SRUSB3-FLUSB 3.0 decoding and analysis
SUP6-CMXGBTSUP6-CMXGBT-FL10 GBASE-T Ethernet automated compliance test solution. ≥4 GHz is recommended

Add serial compliance

All serial compliance products require option 6-WIN (SSD with Microsoft Windows 10 operating system)

SUP6-CMAUTOEN SUP6-CMAUTOEN-FL Automotive Ethernet automated compliance test solution (100BASE-T1 and 1000BASE-T1)
SUP6-CMAUTOEN10 SUP6-CMAUTOEN10-FL Automotive Ethernet (10BASE-T1S Short Reach) automated compliance test solution
SUP6-CMAUTOEN10G SUP6-CMAUTOEN10G-FL Automotive Ethernet (MultiGigBase-T1) automated compliance test solution (requires option 6-DJA)
SUP6-AUTOEN-BND Automotive Ethernet compliance, signal separation, PAM3 analysis, 100Base-T1 serial analysis (requires options 6-DJA and 6-WIN)
SUP6-AUTOEN-SS SUP6-AUTOEN-SS-FL Automotive Ethernet signal separation
SUP6-CMINDUEN10 SUP6-CMINDUEN10-FL Industrial Ethernet (10Base-T1L Long Reach) automated compliance test solution
SUP6-CMCPHY20 SUP6-CMCPHY20-FL MIPI C-PHY 2.0 Tx automated conformance test solution (requires option 6-DJA)
SUP6-CMDPHY SUP6-CMDPHY-FL MIPI D-PHY 1.2 automated compliance test solution
SUP6-CMDPHY21SUP6-CMDPHY21-FLMIPI D-PHY 2.1 Tx automated conformance test solution (requires option 6-DJA)
SUP6-CMENET SUP6-CMENET-FL Ethernet automated compliance test solution (10BASE-T, 100BASE-T, and 1000BASE-T)
SUP6-CMENETMLSUP6-CMENETML-FLMultilane Ethernet automated compliance test solution (10Base-T, 100Base-T, 1000Base-T)
SUP6-CMNBASET SUP6-CMNBASET-FL 2.5 and 5 GBASE-T Ethernet automated compliance test (2.5 GHz is recommended)
SUP6-CMUSB2 SUP6-CMUSB2-FL USB 2.0 automated compliance test solution
Add advanced analysis SUP6-DBLVDS SUP6-DBLVDS-FL LVDS debug and analysis (requires options 6-DJA and 6-WIN)
SUP6-DJA SUP6-DJA-FL Advanced jitter and eye analysis
SUP6-DPM SUP6-DPM-FL Digital power management
SUP6-MTM SUP6-MTM-FL Mask and Limit testing
SUP6-PAM3 SUP6-PAM3-FL PAM3 analysis (requires options 6-DJA and 6-WIN)
SUP6-PS2 N/A Power solution bundle (6-PWR, THDP0200, TCP0030A, and 067-1686-XX deskew fixture)
SUP6-PWR SUP6-PWR-FL Advanced power measurements and analysis
SUP6-SV-BW-1 SUP6-SV-BW-1-FL Increase Spectrum View capture bandwidth to 2 GHz
SUP6-SV-RFVT SUP6-SV-RFVT-FL Spectrum View RF vs. Time traces, triggers, Spectrograms, and IQ capture
SUP6-UDFLT SUP6-UDFLT-FLUser defined filter creation tool
SUP6-VID SUP6-VID-FL NTSC, PAL, and SECAM video triggering
SUP6-WBG-DPT SUP6-WBG-DPT-FLWide Bandgap SiC/GaN Double Pulse Test Measurements and Analysis
Add memory analysis SUP6-DBDDR3 SUP6-DBDDR3-FL DDR3 and LPDDR3 debug and analysis
SUP6-CMDDR3 SUP6-CMDDR3-FL

DDR3 and LPDDR3 automated compliance test solution using TekExpress Automation Platform.

Requires options 6-DBDDR3, 6-DJA and SSD with Microsoft WIndows 10 operating system.

≥4 GHz required, 8 GHz recommended for testing of all DDR3 speeds.

Add digital voltmeter N/AN/A Add digital voltmeter/trigger frequency counter

(Free with product registration at www.tek.com/register6mso)

Upgrade feature Upgrade Description
Add Windows operating system SUP6-WIN Add removable SSD with Windows 10 operating system

Bandwidth upgrades after purchase

Add bandwidth upgrades in the future
The analog bandwidth of 6 Series products can be upgraded after initial purchase. Bandwidth upgrades are purchased based on the current bandwidth and the desired bandwidth. All bandwidth upgrades can be performed in the field by installing a software license and a new front panel label.
Model to be upgraded Bandwidth before upgrade Bandwidth after upgrade Order this bandwidth upgrade

MSO64

1 GHz 2.5 GHz SUP6-BW10T254
1 GHz 4 GHz SUP6-BW10T404
1 GHz 6 GHz SUP6-BW10T604
1 GHz 8 GHz SUP6-BW10T804
2.5 GHz 4 GHz SUP6-BW25T404
2.5 GHz 6 GHz SUP6-BW25T604
2.5 GHz 8 GHz SUP6-BW25T804
4 GHz 6 GHz SUP6-BW40T604
4 GHz 8 GHz SUP6-BW40T804