Contact us
Live Chat with Tek representatives. Available 6:00 AM - 4:30 PM
Call us at
Available 6:00 AM – 5:00 PM (PST) Business Days
Download
Download Manuals, Datasheets, Software and more:
Feedback
Analog Oscilloscope
2205
The products on this datasheet are no longer being sold by Tektronix.
View Tektronix Encore for reconditioned test equipment.
Check support and warranty status for these products.
Features & Benefits
- 40-MHz Bandwidth Option 40
- Peak-to-Peak Auto Triggering
- 0.3-Div Trigger Sensitivity
- Front-Panel Z-Axis Input
- 10-ns/div Max. Sweep Rate
Characteristics
Vertical System (2 Identical Channels)
Bandwidth (-3 dB) and Rise Time - 20 MHz and 17.5 ns (5 to 35 °C). 15 MHz and 23.3 ns (0 to 40 °C).
2205 Option 40 Bandwidth (-3 dB) and Rise Time - 40 MHz and 8.75 ns (+5 to +35 °C). 35 MHz and 10 ns (0 to 40 °C).
Deflection Factor and Accuracy - 5 mV/div to 5 V/div ±3%, (15 to 35 °C); ±5%, (0 to 40 °C).
Vertical Operating Modes - CH 1, CH 2, CH 2 IN-VERT, ADD, ALT, CHOP.
CMRR - At least 10:1 at 10 MHz.
Input R and C - 1 Megohm, 25 pF.
Max Input Voltage - 400 V (DC + peak AC), 800 Vpk-pk.
Channel Isolation - 100:1 at 20 MHz.
Horizontal System
Sweep Speeds - 0.5 s/div to 0.1 µs/div (X10 MAG to 10 ns/div)
Accuracy - ±3%; Magnified ±4% (degrade by 1% outside 15 to 35 °C)
Horizontal Operating Modes - X1, X10, X-Y
Trigger System
Trigger Sensitivity -
Internal: 0.30 div at 5 MHz, 1.0 div at 30 MHz
External: 40 mV at 5 MHz, 150 mV at 30 MHz (150 mV at 45 MHz for option 40)
Trigger Operating Modes - Peak-Peak AUTO, NORM, TV FIELD, TV LINE, SGL SWP
Trigger Source - VERT MODE, CH 1, CH 2, LINE, EXT, EXT/10
Trigger Coupling - AC, DC
X-Y Operation
Deflection Factors - Same as vertical system
Bandwidth -
X-Axis: 2 MHz
Y-Axis: Same as vertical system
Phase Difference - ±3 degree from DC to 50 kHz
CRT System
Display - 8 cm × 10 cm, 1.8 kV nominal voltage (2205); 6 kV nominal voltage (Opt. 40).
Controls - INTENSITY, TRACE ROTATION, BEAM FIND, FOCUS.
Z-Axis - 5 V causes noticeable modulation. Usable to 5 MHz.
Power Requirements
Line Voltage Range -
Low: 95 V AC to 128 V AC
High: 185 V AC to 250 V AC
Line Frequency - 48 Hz to 440 Hz
Maximum Power Consumption - 40 W (60 VA)
Environmental Specifications
Instruments meet in part, the environmental requirements of MIL-T-28800D or C for Type III, Class 3, Style D or C equipment as described below.
Ambient Temperature -
Operating: 0 to +40 °C.
Nonoperating: -55 to +75 °C.
Humidity -
Percent: 95% five cycles, (120 hours) 90% +40 °C 4 hours.
Reference: MIL-T-28800D, paragraph 4.5.5.1.2.2.
Tektronix 062-2847-00, ll.
Altitude -
Operating: To 4,500 m (15,000 ft.), maximum operating temp decreases 1 °C per 300 m above 1500 m.
Nonoperating: To 15,000 m (50,000 ft.).
EMC - Meets Class B requirements per VDE 0871-B for radiated and conducted emissions and FCC requirements.
Vibration -
Operating: 15 minutes along each of the 3 major axes, 10 Hz to 55 Hz to 10 Hz in one-minute cycles. Hold for 10 minutes at 55 Hz. Displ. (in pk-pk): 0.015.
Shock -
Operating: 30 g, 1/2 sine, 11-ms duration, 3 shocks per axis along each major axis. Total of 18 shocks.
Other Characteristics
Safety - UL 1244 listed, Certified to CSA C22.2 No. 231-M89
Warranty - 1 year; 3 years optional
Physical Characteristics
Dimensions |
mm |
in. |
---|---|---|
Height |
137 |
5.4 |
Width (with handle) |
380 |
15.0 |
Depth (without front cover) |
440 |
17.3 |
Weight |
kg |
lbs. |
Net |
6.7 |
14.8 |