Kontaktaufnahme
Live-Chat mit Tektronix-Vertretern. Verfügbar von 9 bis 17 Uhr CET Geschäftstage.
Kontaktieren Sie uns telefonisch unter
Verfügbar von 9 bis 17 Uhr CET Geschäftstage.
Download
Laden Sie Handbücher, Datenblätter, Software und vieles mehr herunter:
Feedback
Curve Tracer
371A
Die Produkte dieses Datenblatts gehören nicht mehr zum Angebot von Tektronix.
Tektronix Encore nach aufgearbeiteten Testgeräten durchsuchen.
Garantie- und Supportstatus dieser Produkte überprüfen.
Features & Benefits
- HighPower Measurements of Semiconductor Devices
- Up to 3,000 Volts
- Up to 400 Amps
- Up to 3,000 Watts
- Waveform Comparison
- Envelope Display
- Waveform Averaging
- Kelvin Sense Measurements
- Fully Programmable
- 3.5Inch MS DOSCompatible Disk Storage to Quickly Save and Recall Setups
Applications
- Manual or Automated High Resolution DC
- Parametric Characterization of Semiconductors
- Incoming Inspection
- Manufacturing Test
- Process Monitoring and Quality Control
- Data Sheet Generation
- Component Matching
- Failure Analysis
- Engineering
371A High Power Programmable Curve Tracer
The 371A, the industry leader in high power curve tracers, is used for testing a wide variety of power semiconductors. It performs DC parametric characterization of thyristors, SCRs and power MOSFETs.
The high voltage collector mode permits testing the Off-characteristics of a device up to 3,000 volts. The pulsed high current collector mode provides output current pulses greater than 400 amps peak for testing On-characteristics. It also permits high power testing up to 3,000 watts.
In the R&D lab, the 371A is used for characterization of new designs of power devices, extraction of SPICE parameters, failure analysis and data sheet generation.
In manufacturing test, the 371A is used to verify device quality and process monitoring.
The 371A is used for incoming inspection to verify power device performance and perform failure analysis.
Interactive Programmable Control
Interactive control of all measurements can be done from the full featured front panel or over the GPIB. Operating parameters can be adjusted and stored and recalled using several storage methods including the 371A nonvolatile memory, the built-in MS DOScompatible floppy disk or to an external controller.
Automated Cursor Measurements
The 371A provides three cursor measurement modes. The Dot cursor provides direct screen readout of voltage, current, gm or DC beta at any point. The Window cursor can be positioned between two curves to measure small signal beta or gm, and can also be used for visual go/no-go tests. The Function Line cursor provides screen readout of a slope or intercept value.
Sweep Measurement Mode
In the sweep measurement mode, the 371A will automatically construct a family of curves while stimulating the device with low duty-cycle pulses. With this capability, power curves can be displayed without excessive heating of the device.
Test Fixturing
A test fixture is a standard accessory that provides safe device enclosure to ensure operator protection during measurements. The test fixture accommodates standard A1001 through A1005 adapters with Kelvin sensing, 3-Pin adapters without Kelvin sensing and the A1023 and A1024 surface mount adapters.
Characteristics
Collector Supply
Modes -
High Current: 250 µs pulses with max. peak of 30 volts
High Voltage: full rectified sine with max. peak of 3000 volts (positive and negative polarities for both modes)
|
Range |
Peak Current |
Peak Power |
---|---|---|---|
High Current |
30 V |
400 A |
3 kW |
(pulsed) |
30 V |
40 A |
300 W |
High Voltage |
3 kV |
40 mA ±20% |
30 W |
|
3 kV |
4 mA ±20% |
3 W |
|
300 V |
4 mA ±20% |
300 mW |
|
300 V |
0.4 mA ±20% |
30 mW |
Measurement Characteristics
Collector Current
Maximum Power Setting |
Vertical Range (in 1-2-5 increments) |
---|---|
3 kW |
1 A/div to 50 A/div |
300 W |
500 mA/div to 5 A/div |
30 W |
100 µA/div to 5 mA/div |
3 W, 300 mW |
10 µA/div to 500 µA/div |
30 mW |
1 µA/div to 50 µA/div |
Accuracy - Within 0.1 vertical division.
Test Fixture - Designed to allow easy connection to a variety of devices. It has a safety enclosure to assure operator protection. Special patch cords are provided for connecting large devices.
Step Generator
Modes -
Current (stair step): DC, 500 µs ±10% of pulse
Voltage (stair step): DC
Step Range -
Current: 1 mA/step to 2 A/step pulsed (high current mode). 1 µA/step to 2 mA/step DC (high voltage mode).
Voltage: 200 mV/step to 5 V/step
Offset: ±10X step amplitude (5X in 5 V/step and 2 V/step settings)
Safety - UL1244, CSA231, HD401 S1