DPO3000/MSO3000: What is the most efficient way to work with post-processed displays such as a math waveform on 5M record length acquisitions?

Вопрос :

DPO3000/MSO3000: What is the most efficient way to work with post-processed displays such as a math waveform on 5M record length acquisitions?

Ответ :

Operations such as math, bus and search that process 5M waveforms will work best if acquisitions are stopped before enabling the operation. The "not yet computed" message will be displayed until the operation is completed.